The effect of varying LED intensity on pulse oximeter accuracy.

J P de Kock1, K J Reynolds, L Tarassenko, J T Moyle.   

Abstract

Most pulse oximeters automatically alter the intensity of their light-emitting diodes (LEDs) according to the absorption of the finger, toe or earlobe to which they are attached. This paper investigates the effect of changing LED intensity on pulse oximeter accuracy. Our results show that the peak wavelength of a red LED typically increases by 8 nm as its intensity is increased ten-fold. To determine whether this shift introduces a significant error, a simple theoretical model based on the Beer-Lambert law is used. The model predicts that a 10:1 change in LED intensity results in a 2.5% error at 50% arterial oxygen saturation (SpO2). At high saturations (SpO2 greater than or equal to 85%) the model predicts little loss of accuracy and thus any effect due to changes in LED intensity will be apparent only at low saturations.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1941984     DOI: 10.3109/03091909109016208

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  5 in total

1.  Evaluation of a new pulse oximeter testing device.

Authors:  J A Fisher; T Martire; G A Volgyesi
Journal:  Can J Anaesth       Date:  1996-02       Impact factor: 5.063

2.  Evaluation of a pulse oximeter sensor tester.

Authors:  Shuba Dugani; Iljaz Hodzovic; Seema Sindhakar; Aida Nadra; Clare Dunstan; Anthony Richard Wilkes; John Mecklenburgh
Journal:  J Clin Monit Comput       Date:  2011-05-18       Impact factor: 2.502

3.  The effect of dyshemoglobins on pulse oximetry: Part I, Theoretical approach and Part II, Experimental results using an in vitro test system.

Authors:  K J Reynolds; E Palayiwa; J T Moyle; M K Sykes; C E Hahn
Journal:  J Clin Monit       Date:  1993-04

Review 4.  Skin Pigmentation Influence on Pulse Oximetry Accuracy: A Systematic Review and Bibliometric Analysis.

Authors:  Ana M Cabanas; Macarena Fuentes-Guajardo; Katina Latorre; Dayneri León; Pilar Martín-Escudero
Journal:  Sensors (Basel)       Date:  2022-04-29       Impact factor: 3.847

5.  Noninvasive Monitoring of Blood Glucose Using Color-Coded Photoplethysmographic Images of the Illuminated Fingertip Within the Visible and Near-Infrared Range: Opportunities and Questions.

Authors:  Thorsten Vahlsing; Sven Delbeck; Steffen Leonhardt; H Michael Heise
Journal:  J Diabetes Sci Technol       Date:  2018-09-15
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.