Literature DB >> 19410948

Intraindividual comparison of aspherical and spherical intraocular lenses of same material and platform.

Shinichiro Ohtani1, Kazunori Miyata, Tomokazu Samejima, Masato Honbou, Tetsuro Oshika.   

Abstract

PURPOSE: There have been few studies which compared aspherical and spherical intraocular lenses (IOLs) of same material and platform in bilateral cataract cases. We performed an intraindividual comparison of ocular aberration and scotopic, mesopic, and photopic contrast sensitivity with aspherical and spherical IOLs, using the same IOL material and platform manufactured by the same company.
DESIGN: Prospective, randomized, controlled study. PARTICIPANTS: Eighty-two eyes of 41 patients undergoing bilateral cataract surgery.
METHODS: One eye of a patient was assigned to acrylic foldable aspherical IOL (Tecnis ZA9003, Advanced Medical Optics), and the contralateral eye was allocated to acrylic foldable spherical IOL (AR40e, Advanced Medical Optics). All patients were examined at 2 days, 1 week, and 1 month postoperatively. MAIN OUTCOME MEASURES: Best-corrected visual acuity (BCVA), contrast sensitivity under scotopic (15 lux), mesopic (70 lux), and photopic (180 lux) conditions, corneal and ocular wavefront aberrations, anterior chamber depth, amount of IOL decentration and tilt, pupil diameter under scotopic (3 lux) and photopic (250 lux) conditions, area of anterior capsule opening, degree of posterior capsule opacification, and all-distance visual acuity.
RESULTS: There was no significant difference between IOLs in BCVA, anterior chamber depth, amount of IOL decentration and tilt, pupil diameter, area of anterior capsule opening, and degree of posterior capsule opacification. In corneal wavefront aberrations, there was no difference in 3rd-, 4th-, and total higher-order root-mean-square (RMS). In ocular wavefront aberration, aspherical IOL showed significantly lower 4th-order (P<0.001) and total higher-order RMS (P<0.001) than spherical IOL, but not in 3rd-order RMS (P = 0.103). Contrast sensitivity under scotopic conditions was significantly better with aspherical IOL than with spherical IOL at 3 (P = 0.0015), 6 (P = 0.0192), and 12 cycles per degree (P = 0.0315). Contrast sensitivity under mesopic and photopic conditions was not significantly different between IOLs. There was no between-group difference in visual acuity at 0.3, 0.5, 0.7, 1.0, or 5.0 meters measured with full distance correction.
CONCLUSIONS: Acrylic foldable aspherical IOL (Tecnis ZA9003) yielded significantly lower ocular wavefront aberration and better contrast sensitivity under scotopic condition without compromising depth of focus. FINANCIAL DISCLOSURE(S): The authors have no proprietary or commercial interest in any materials discussed in this article.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19410948     DOI: 10.1016/j.ophtha.2008.11.022

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  11 in total

1.  Assessment of contrast sensitivity by Spaeth Richman Contrast Sensitivity Test and Pelli Robson Chart Test in patients with varying severity of glaucoma.

Authors:  Sahil Thakur; Parul Ichhpujani; Suresh Kumar; Ravneet Kaur; Sunandan Sood
Journal:  Eye (Lond)       Date:  2018-05-14       Impact factor: 3.775

2.  Changes in corneal aberrations after cataract surgery.

Authors:  Yuki Hidaka; Takefumi Yamaguchi; Megumi Saiki; Murat Dogru; Kazuo Tsubota; Kazuno Negishi
Journal:  Jpn J Ophthalmol       Date:  2016-03-02       Impact factor: 2.447

3.  Clinical study of customized aspherical intraocular lens implants.

Authors:  Lie-Xi Jia; Zhao-Hui Li
Journal:  Int J Ophthalmol       Date:  2014-10-18       Impact factor: 1.779

4.  Improvement of neural contrast sensitivity after long-term adaptation in pseudophakic eyes.

Authors:  Seung Pil Bang; Justin D Aaker; Ramkumar Sabesan; Geunyoung Yoon
Journal:  Biomed Opt Express       Date:  2022-08-03       Impact factor: 3.562

5.  Visual performance of the Tecnis one-piece lens ZCB00.

Authors:  Sherein S Wahba; Riad F Riad; Fathy F Morkos; Abdallah K Hassouna; Maged M Roshdy
Journal:  Clin Ophthalmol       Date:  2011-12-23

6.  A comparative study on visual and optical performance of Akreos AO and Kontur AB IOLs after phacoemulsification cataract surgery.

Authors:  Hamid Gharaee; Masoud Zabihifard; Alireza Eslampour; Samira Hassanzadeh; Masoud Shafiee
Journal:  J Curr Ophthalmol       Date:  2016-03-28

7.  Distribution of corneal spherical aberration in a Tanzanian population.

Authors:  Hiroki Asano; Takahiro Hiraoka; Yusuke Seki; Teppei Shibata; Hiromi Osada; Takanori Saruta; Natsuko Hatsusaka; Fukumi Fujikake; Yoshiaki Tabata; Cellina Mhina; Anna Sanyiwa; Tetsuro Oshika; Hiroshi Sasaki
Journal:  PLoS One       Date:  2019-09-12       Impact factor: 3.240

8.  Aspheric intraocular lenses implantation for cataract patients with extreme myopia.

Authors:  Yanwen Fang; Yi Lu; Aizhu Miao; Yi Luo
Journal:  ISRN Ophthalmol       Date:  2014-03-19

9.  Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses.

Authors:  Michael Lasta; Kata Miháltz; Illés Kovács; Pia Veronika Vécsei-Marlovits
Journal:  J Ophthalmol       Date:  2017-08-16       Impact factor: 1.909

10.  Visual and Refractive Outcomes following Bilateral Implantation of Extended Range of Vision Intraocular Lens with Micromonovision.

Authors:  Sri Ganesh; Sheetal Brar; Archana Pawar; Kirti J Relekar
Journal:  J Ophthalmol       Date:  2018-02-06       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.