Implant stability in relation to osseointegration: an experimental study in the Labrador dog.

Ingemar Abrahamsson, Elena Linder, Niklaus P Lang.   

Abstract

OBJECTIVE: Resonance frequency analysis (RFA) is supposed to determine implant stability. The relation between RFA and the degree of bone-to-implant contact (BIC), however, is unclear. The objective of the present experiment was to evaluate RFA values in relation to osseointegration.
MATERIAL AND METHODS: In 20 Labrador dogs, all mandibular premolars were extracted bilaterally. After 3 months, four transmucosal screw-shaped experimental implants were placed in each mandibular premolar region. The implants (12 mm length, slashed circle 4.1mm, insertion depth 9 mm) were either SLA surface or turned surface implants. The animals were divided into four groups (five dogs in each group) to study healing following implant installation at 2 h, 4 days, 1, 2, 4, 6, 8 and 12 weeks. Two experimental implants of each type were installed in each edentulous premolar region. A plaque control program was initiated 2 weeks after each implant installation. RFA assessments were performed at the time of implant installation, at one to three occasions during the monitoring period and at the termination of the experiment. At the end of the experiment, the dogs were sacrificed and each implant site was dissected and processed for histological analysis. The results of the histological analysis, i.e. marginal bone level, degree of osseointegration (BIC%) and bone density, were compared with the corresponding Implant Stability Quotient (ISQ) values of the RFA assessment.
RESULTS: No correlations between histological parameters of osseointegration and ISQ values could be identified. Marginal bone level changes, differences in BIC% and bone density were not reflected in the RFA at any time-point during the 12-week monitoring period.
CONCLUSION: The value of RFA to predict implant stability over time and to determine at which time-point an implant may be exposed to functional load has to be questioned.

Entities:  

Mesh:

Year:  2009        PMID: 19405177     DOI: 10.1111/j.1600-0501.2008.01646.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  16 in total

1.  Correlation between resonance frequency, insertion torque and bone-implant contact in self-cutting threaded implants.

Authors:  Yahya Açil; Jan Sievers; Aydin Gülses; Mustafa Ayna; Jörg Wiltfang; Hendrik Terheyden
Journal:  Odontology       Date:  2016-08-18       Impact factor: 2.634

2.  Role of single photon emission computerised tomography in evaluating osseointegration of indigenous DRDO implants: An in vivo study.

Authors:  S K Bhandari; Anupam Mondal
Journal:  Med J Armed Forces India       Date:  2012-12-01

3.  3D full-field strain in bone-implant and bone-tooth constructs and their morphological influential factors.

Authors:  Yuxiao Zhou; Chujie Gong; Mehran Hossaini-Zadeh; Jing Du
Journal:  J Mech Behav Biomed Mater       Date:  2020-05-19

4.  Influence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis: An in vitro study.

Authors:  David Chávarri-Prado; Aritza Brizuela-Velasco; Markel Diéguez-Pereira; Esteban Pérez-Pevida; Antonio Jiménez-Garrudo; Iratxe Viteri-Agustín; Alejandro Estrada-Martínez; Oier Montalbán-Vadillo
Journal:  J Clin Exp Dent       Date:  2020-03-01

5.  Effect of bone quality and implant surgical technique on implant stability quotient (ISQ) value.

Authors:  Hong-Gi Yoon; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim; Su-Young Lee
Journal:  J Adv Prosthodont       Date:  2011-03-31       Impact factor: 1.904

6.  Drilling dimension effects in early stages of osseointegration and implant stability in a canine model.

Authors:  Felipe-Eduardo Baires-Campos; Ryo Jimbo; Estevam-Augusto Bonfante; Maiolino-Thomaz Fonseca-Oliveira; Camila Moura; Darceny Zanetta-Barbosa; Paulo-Guilherme Coelho
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2015-07-01

7.  Ag-plasma modification enhances bone apposition around titanium dental implants: an animal study in Labrador dogs.

Authors:  Shichong Qiao; Huiliang Cao; Xu Zhao; Hueiwen Lo; Longfei Zhuang; Yingxin Gu; Junyu Shi; Xuanyong Liu; Hongchang Lai
Journal:  Int J Nanomedicine       Date:  2015-01-14

8.  Monitoring the Changes of Material Properties at Bone-Implant Interface during the Healing Process In Vivo: A Viscoelastic Investigation.

Authors:  Hsiang-Ho Chen; Wei-Yi Lai; Tze-Jian Chee; Ya-Hui Chan; Sheng-Wei Feng
Journal:  Biomed Res Int       Date:  2017-03-08       Impact factor: 3.411

9.  Development of implant/interconnected porous hydroxyapatite complex as new concept graft material.

Authors:  Kazuya Doi; Hiroshi Oue; Koji Morita; Shiho Kajihara; Takayasu Kubo; Katsunori Koretake; Vittoria Perrotti; Giovanna Iezzi; Adriano Piattelli; Yasumasa Akagawa
Journal:  PLoS One       Date:  2012-11-09       Impact factor: 3.240

10.  An ex vivo model in human femoral heads for histopathological study and resonance frequency analysis of dental implant primary stability.

Authors:  Pedro Hernández-Cortés; Alberto Monje; Pablo Galindo-Moreno; Andrés Catena; Inmaculada Ortega-Oller; José Salas-Pérez; Francisco Mesa; Rafael Gómez-Sánchez; Mariano Aguilar; David Aguilar; Francisco O'Valle
Journal:  Biomed Res Int       Date:  2014-06-04       Impact factor: 3.411