Tumor localization using cone-beam CT reduces setup margins in conventionally fractionated radiotherapy for lung tumors.

Anamaria R Yeung1, Jonathan G Li, Wenyin Shi, Heather E Newlin, Alexei Chvetsov, Chihray Liu, Jatinder R Palta, Kenneth Olivier.   

Abstract

PURPOSE: To determine whether setup margins can be reduced using cone-beam computed tomography (CBCT) to localize tumor in conventionally fractionated radiotherapy for lung tumors. METHODS AND MATERIALS: A total of 22 lung cancer patients were treated with curative intent with conventionally fractionated radiotherapy using daily image guidance with CBCT. Of these, 13 lung cancer patients had sufficient CBCT scans for analysis (389 CBCT scans). The patients underwent treatment simulation in the BodyFix immobilization system using four-dimensional CT to account for respiratory motion. Daily alignment was first done according to skin tattoos, followed by CBCT. All 389 CBCT scans were retrospectively registered to the planning CT scans using automated soft-tissue and bony registration; the resulting couch shifts in three dimensions were recorded.
RESULTS: The daily alignment to skin tattoos with no image guidance resulted in systematic (Sigma) and random (sigma) errors of 3.2-5.6 mm and 2.0-3.5 mm, respectively. The margin required to account for the setup error introduced by aligning to skin tattoos with no image guidance was approximately 1-1.6 cm. The difference in the couch shifts obtained from the bone and soft-tissue registration resulted in systematic (Sigma) and random (sigma) errors of 1.5-4.1 mm and 1.8-5.3 mm, respectively. The margin required to account for the setup error introduced using bony anatomy as a surrogate for the target, instead of localizing the target itself, was 0.5-1.4 cm.
CONCLUSION: Using daily CBCT soft-tissue registration to localize the tumor in conventionally fractionated radiotherapy reduced the required setup margin by up to approximately 1.5 cm compared with both no image guidance and image guidance using bony anatomy as a surrogate for the target.

Entities:  

Mesh:

Year:  2009        PMID: 19395197     DOI: 10.1016/j.ijrobp.2008.09.048

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  35 in total

1.  Localization accuracy of the clinical target volume during image-guided radiotherapy of lung cancer.

Authors:  Geoffrey D Hugo; Elisabeth Weiss; Ahmed Badawi; Matthew Orton
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-01-27       Impact factor: 7.038

2.  Localization accuracy from automatic and semi-automatic rigid registration of locally-advanced lung cancer targets during image-guided radiation therapy.

Authors:  Scott P Robertson; Elisabeth Weiss; Geoffrey D Hugo
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

3.  Use of planar kV vs. CBCT in evaluation of setup errors in oesophagus carcinoma radiotherapy.

Authors:  Liliana Martins; Jose Guilherme Couto; Barbara Barbosa
Journal:  Rep Pract Oncol Radiother       Date:  2015-11-14

4.  Estimation of CT cone-beam geometry using a novel method insensitive to phantom fabrication inaccuracy: implications for isocenter localization accuracy.

Authors:  J Chetley Ford; Dandan Zheng; Jeffrey F Williamson
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

Review 5.  Are three doses of stereotactic ablative radiotherapy (SABR) more effective than 30 doses of conventional radiotherapy?

Authors:  Anna O Simeonova; Katharina Fleckenstein; Hansjörg Wertz; Anian Frauenfeld; Judit Boda-Heggemann; Frank Lohr; Frederik Wenz
Journal:  Transl Lung Cancer Res       Date:  2012-03

6.  Optimization of acquisition parameters and accuracy of target motion trajectory for four-dimensional cone-beam computed tomography with a dynamic thorax phantom.

Authors:  Yoshinobu Shimohigashi; Fujio Araki; Masato Maruyama; Yuji Nakaguchi; Kengo Nakato; Nozomu Nagasue; Yudai Kai
Journal:  Radiol Phys Technol       Date:  2014-10-07

7.  Evaluating the four-dimensional cone beam computed tomography with varying gantry rotation speed.

Authors:  S A Yoganathan; K J Maria Das; Shajahan Mohamed Ali; Arpita Agarwal; Surendra P Mishra; Shaleen Kumar
Journal:  Br J Radiol       Date:  2016-02-26       Impact factor: 3.039

8.  Stereotactic ablative body radiotherapy for non-small-cell lung cancer: setup reproducibility with novel arms-down immobilization.

Authors:  Karen Moore; Claire Paterson; Jonathan Hicks; Stephen Harrow; Mark McJury
Journal:  Br J Radiol       Date:  2016-10-06       Impact factor: 3.039

9.  A study of respiration-correlated cone-beam CT scans to correct target positioning errors in radiotherapy of thoracic cancer.

Authors:  J P Santoro; J McNamara; E Yorke; H Pham; A Rimner; K E Rosenzweig; G S Mageras
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

10.  Technical note: improved positioning protocol for patient setup accuracy in conventional radiotherapy for lung cancer.

Authors:  Hongbo Chai; Yuichiro Narita; Masafumi Takagi; Mikiko Kudo; Tomomi Kimura; Keiichi Kattou
Journal:  Radiol Phys Technol       Date:  2019-09-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.