Literature DB >> 19372725

Analysis of fifty-six cochlear implant device failures.

Jan Gosepath1, Karl Lippert, Annerose Keilmann, Wolf J Mann.   

Abstract

OBJECTIVE: Our aim was to present a failure analysis after cochlear implant revision surgery in a large series of children and adults and to assess the outcome and audiologic performance.
METHODS: Fifty-six cochlear implant failures that occurred in 422 devices implanted between 1990 and 2007 at the Department of Otolaryngology, Head and Neck Surgery at the University of Mainz, Germany, were retrospectively analyzed. The causes of failure were reviewed evaluating the individual history, telemetric and intraoperative findings and manufacturer's investigation reports.
RESULTS: We performed 56 surgical revisions in a series of 422 consecutive implants (overall revision rate: 13.27%). The most frequent causes for revision were hard failures (58.9%), most commonly caused by traumatic impact (37.5%), especially in the pediatric population. Soft failures were less frequent (21.4%). Surgical reimplantations, although challenging in some cases, were performed without complications and with an electrode insertion depth comparable to that at the time of the initial implantation in all patients. The average audiologic performance improved by 2.4 dB in pure-tone perception levels after reimplantation.
CONCLUSION: The cochlear implant failure rates vary between children and adults as well as between different implant manufactures. However, cochlear reimplantation is safe with excellent and predictable results in audiologic performance.

Entities:  

Mesh:

Year:  2009        PMID: 19372725     DOI: 10.1159/000212756

Source DB:  PubMed          Journal:  ORL J Otorhinolaryngol Relat Spec        ISSN: 0301-1569            Impact factor:   1.538


  6 in total

1.  The histopathology of revision cochlear implantation.

Authors:  Joonhan Lee; Donald K Eddington; Joseph B Nadol
Journal:  Audiol Neurootol       Date:  2010-12-24       Impact factor: 1.854

2.  Cochlear Implant Reliability: On the Reporting of Rates of Revision Surgery.

Authors:  Graham O'Neill; Neil S Tolley
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2020-04-01

Review 3.  [Magnetic resonance imaging in patients with magnetic hearing implants: overview and procedural management].

Authors:  S Nospes; W Mann; A Keilmann
Journal:  Radiologe       Date:  2013-11       Impact factor: 0.635

4.  A Retrospective Study of Cochlear Re-Implantations - Experience from a Large Centre in India.

Authors:  Abha Kumari; Sunil Goyal; Senthil Vadivu Arumugam; Viswanathan Vasumathi; Neha Chauhan; Mohan Kameswaran
Journal:  Turk Arch Otorhinolaryngol       Date:  2019-03-01

5.  Cochlear re-implant rates in children: 20 years experience in a quaternary paediatric cochlear implant centre.

Authors:  Marilena Trozzi; Harry R F Powell; Shamim Toma; Waseem Ahmed; Christopher G Jephson; Kaukab Rajput; Lesley A Cochrane
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-08-09       Impact factor: 2.503

6.  Audit of Bilateral Simultaneous Cochlear Implantation in Pediatric Population: South Indian Study.

Authors:  Abha Kumari; Sunil Goyal; Neha Chauhan; Thirugnanam Sarankumar; Krushna Chaitanya; Mohan Kameswaran
Journal:  Turk Arch Otorhinolaryngol       Date:  2018-03-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.