Comparison of oxygen consumption measurements: indirect calorimetry versus the reversed Fick method.

M N Smithies1, B Royston, K Makita, K Konieczko, J F Nunn.   

Abstract

OBJECTIVE: To compare measurement of oxygen consumption (VO2) by spirometry and the reversed Fick method.
DESIGN: Within-patient comparison using simultaneous measurements by the two methods, one previously calibrated on a metabolic simulator. PATIENTS: Twenty sets of observations on eight patients (57 to 83 yrs) requiring mechanical ventilation in a critical care unit.
INTERVENTIONS: None during or immediately before the measurements.
MEASUREMENTS AND MAIN RESULTS: Duplicate pairs of measurements of VO2 were made with a previously validated spirometric technique and the reversed Fick method (Qt[CaO2 - CVO2]), where Qt is cardiac output, CaO2 is arterial oxygen content, and CVO2 is mixed venous oxygen content. The coefficient of variation of the difference between duplicate measurements by the former technique was only 2.53% compared with 10.4% for the latter. The mean VO2 measurement by the spirometric method was 285.7 +/- 40.7 (SD) mL/min standard temperature and pressure, dry (STPD) and for the reversed Fick method, the mean VO2 measurement was 249.3 +/- 38.5 mL/min STPD. The mean difference was 36.4 +/- 28.5 mL/min STPD (p less than .001).
CONCLUSIONS: The repeatability of the spirometric method was four times better than the reversed Fick method. The latter gave a significantly lower value that probably, in part, reflects the VO2 of the lung, which is included in the spirometric method but not in the reversed Fick measurement.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1935161     DOI: 10.1097/00003246-199111000-00017

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  17 in total

1.  Results of the oxygen Fick method in a closed blood circulation model including "total arteriovenous diffusive shunt of oxygen".

Authors:  Mustafa Ozbek; Ahmet Akay
Journal:  Theory Biosci       Date:  2004-09       Impact factor: 1.919

2.  Continuous measurement of multiple inert and respiratory gas exchange in an anaesthetic breathing system by continuous indirect calorimetry.

Authors:  Christopher Stuart-Andrews; Philip Peyton; Craig Humphries; Gavin Robinson; Brian Lithgow
Journal:  J Clin Monit Comput       Date:  2009-01-24       Impact factor: 2.502

3.  Comparison of oxygen consumption calculated by Fick's principle (using a central venous catheter) and measured by indirect calorimetry.

Authors:  Chiaki Inadomi; Yoshiaki Terao; Kazunori Yamashita; Makoto Fukusaki; Masafumi Takada; Koji Sumikawa
Journal:  J Anesth       Date:  2008-05-25       Impact factor: 2.078

4.  The accuracy and trending ability of cardiac index measured by the fourth-generation FloTrac/Vigileo system™ and the Fick method in cardiac surgery patients.

Authors:  Takuma Maeda; Eisuke Hamaguchi; Naoko Kubo; Akira Shimokawa; Hiroko Kanazawa; Yoshihiko Ohnishi
Journal:  J Clin Monit Comput       Date:  2018-11-07       Impact factor: 2.502

5.  Continuous measurement of oxygen consumption using the reversed fick method.

Authors:  Michihiko Fukui; Maho Imoto; Nobuaki Shime; Tetsuo Hatanaka; Hideaki Tojo
Journal:  J Anesth       Date:  1997-03       Impact factor: 2.078

6.  Reversed Fick principle versus indirect calorimetry: do systematic differences between methods represent intrapulmonary oxygen consumption?

Authors:  A Weyland; W Weyland; M Sydow
Journal:  Intensive Care Med       Date:  1994-07       Impact factor: 17.440

7.  Validation of diagnostic tests.

Authors:  B Bermejo; J R Masclans; M Olona; F J de Latorre
Journal:  Intensive Care Med       Date:  1993       Impact factor: 17.440

8.  Oxygen consumption after cardiopulmonary bypass--different measuring methods yield different results.

Authors:  H M Oudemans-van Straaten; G J Scheffer
Journal:  Intensive Care Med       Date:  1994-07       Impact factor: 17.440

9.  Oxygen consumption after cardiopulmonary bypass--implications of different measuring methods.

Authors:  H M Oudemans-van Straaten; G J Scheffer; L Eysman; C R Wildevuur
Journal:  Intensive Care Med       Date:  1993       Impact factor: 17.440

10.  Effects of sevoflurane before cardiopulmonary bypass on cerebral oxygen balance and early postoperative cognitive dysfunction.

Authors:  Ji-Hong Xu; Tie-Zheng Zhang; Xiao-Feng Peng; Chun-Ji Jin; Jin Zhou; Yi-Nan Zhang
Journal:  Neurol Sci       Date:  2013-03-23       Impact factor: 3.307

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.