Literature DB >> 19304919

Pediatric 99mTc-DMSA SPECT performed by using iterative reconstruction with isotropic resolution recovery: improved image quality and reduced radiopharmaceutical activity.

Niall Sheehy1, Tracy A Tetrault, David Zurakowski, A Hans Vija, Frederic H Fahey, S Ted Treves.   

Abstract

PURPOSE: To compare two methods of reconstructing technetium 99m (99mTc) dimercaptosuccinic acid (DMSA) renal single photon emission computed tomographic (SPECT) data--ordered subset expectation maximization with three-dimensional resolution recovery (OSEM-3D) and filtered back projection (FBP)--in children in terms of improving image quality and reducing the radiopharmaceutical activity and radiation dose.
MATERIALS AND METHODS: The institutional review board approved this retrospective study and waived the requirement for informed patient consent. Fifty sequential pediatric patient 99mTc-DMSA SPECT studies of 98 kidneys were retrospectively analyzed by using a dual-detector gamma camera. FBP reconstruction with data from both detectors was compared with OSEM-3D reconstruction with half the gamma photon counts (ie, data from only one detector). Two nuclear medicine physicians blinded to the patients' medical histories and reconstruction techniques evaluated the studies. Scores for image quality, renal size, and relative function were compared by using paired t tests. Total scores for renal cortical defects were compared by using the Wilcoxon signed rank test. The kappa coefficient was calculated as an indicator of the concordance between the OSEM-3D and FBP reconstruction methods.
RESULTS: Image quality was significantly enhanced with OSEM-3D (P < .001, paired t test). Cortical defects were identified better on OSEM-3D images than on FBP images. Of the 98 kidney SPECT studies analyzed, 19 showed identical cortical defects and 75 showed none at both OSEM-3D and FBP. In four kidneys, OSEM-3D depicted cortical defects that were not seen with FBP. No significant difference in relative renal function between the two methods was observed (P = .973).
CONCLUSION: Compared with FBP, OSEM-3D yielded superior image quality in the evaluation of 99mTc-DMSA renal SPECT data, with the potential for markedly reduced radiation doses and/or shorter scanning times for patients.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19304919     DOI: 10.1148/radiol.2512081440

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  7 in total

1.  Use of (99m)Tc-MDP SPECT for assessment of mandibular growth: development of normal values.

Authors:  Frederic H Fahey; Zachary R Abramson; Bonnie L Padwa; Robert E Zimmerman; David Zurakowski; Mark Nissenbaum; Leonard B Kaban; S Ted Treves
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-12-22       Impact factor: 9.236

Review 2.  Dose Estimation in Pediatric Nuclear Medicine.

Authors:  Frederic H Fahey; Alison B Goodkind; Donika Plyku; Kitiwat Khamwan; Shannon E O'Reilly; Xinhua Cao; Eric C Frey; Ye Li; Wesley E Bolch; George Sgouros; S Ted Treves
Journal:  Semin Nucl Med       Date:  2016-11-09       Impact factor: 4.446

3.  Filtering in SPECT Image Reconstruction.

Authors:  Maria Lyra; Agapi Ploussi
Journal:  Int J Biomed Imaging       Date:  2011-06-23

4.  A Rare Case of a Pediatric Patient Affected by Crossed Fused Renal Ectopia: The usefulness of Dual Tracer99mTc-Diethylenetriamine Pentaacetic Acid/99mTc-Dimercaptosuccinic Acid Scintigraphy.

Authors:  Luca Filippi; Oreste Bagni; Alessia Marcellino; Riccardo Lubrano
Journal:  Indian J Nucl Med       Date:  2020-10-21

Review 5.  Current Status of Radiopharmaceutical Therapy.

Authors:  Sara St James; Bryan Bednarz; Stanley Benedict; Jeffrey C Buchsbaum; Yuni Dewaraja; Eric Frey; Robert Hobbs; Joseph Grudzinski; Emilie Roncali; George Sgouros; Jacek Capala; Ying Xiao
Journal:  Int J Radiat Oncol Biol Phys       Date:  2020-08-14       Impact factor: 7.038

6.  Development of a radiopharmaceutical dose calculator for pediatric patients undergoing diagnostic nuclear medicine studies.

Authors:  Anil Kumar Pandey; Sanjay Kumar Sharma; Punit Sharma; Priyanka Gupta; Rakesh Kumar
Journal:  Indian J Nucl Med       Date:  2013-04

7.  Targeted design of a recombinant tracer for SPECT renal imaging.

Authors:  Pan Liu; Steven E Johnson; Xinfang Xie; Li Gao; Chad R Haney; Ming Zhao; Jing Jin
Journal:  Theranostics       Date:  2021-08-27       Impact factor: 11.556

  7 in total