Toshinori Ohashi1, Kazuma Mawatari, Kae Sato, Manabu Tokeshi, Takehiko Kitamori.
Abstract
We have developed a novel, practical micro-ELISA system for sensitive and rapid allergy diagnosis. The enzymatic reactions occurred under stopped-flow conditions, resulting in both high precision and high sensitivity. A BSA-biotin-avidin linker was introduced for the immobilization of water-soluble allergens on polystyrene microbeads, enabling immobilization of allergens in sufficient density to provide high sensitivity. Evaluation of the system's performance showed a good detection limit (2 ng/mL) for total IgE measurement. In addition, a good correlation with a conventional method (CAP method) was demonstrated using human serum samples from 85 allergy patients. Importantly, sample volumes (5 microL) were 10 times smaller and analysis time (12 min) was >20 times faster than the conventional method. All procedures were automatically regulated with our simple microfluidic system, and all the fluidic, optic and electronic components were integrated for portability. We believe that our system has the potential to become a very powerful tool, particularly for point-of-care diagnosis.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19294312 DOI: 10.1039/b815475a
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799