Literature DB >> 19269985

Selective-site pacing in paediatric patients: a new application of the Select Secure system.

Francesco Cantù1, Paolo De Filippo, Fulvio Gabbarini, Adele Borghi, Roberta Brambilla, Paolo Ferrero, Jennifer Comisso, Tiziana Marotta, Alessandro De Luca, Antonello Gavazzi.   

Abstract

AIMS: The aim of this study was to evaluate the feasibility and reliability of selective-site pacing by means of a new lead system in a paediatric population. This lead system is composed of a 4.1 Fr, active-fixation lead and a steerable catheter that allows easy positioning in selective sites. METHODS AND
RESULTS: Thirty young patients (mean age 9.0 +/- 4.5 years, range 2-16 years) received a single- (10) or dual- (20) chamber pacemaker. The 3830 lead was implanted successfully in the targeted chambers in all patients. The selective RV sites of pacing in 26 of the patients were: 18 mid-septum, 5 outflow tract, 1 low-septum, and 2 LEVO-RV-Apex. In all patients, an intracardiac loop was left in order to avoid stretching of the lead with growth. Mean follow-up duration was 11 +/- 10 months. Atrial sensing and pacing thresholds were 3.2 +/- 1.7 mV and 0.8 +/- 0.6 V at 0.5 ms at implantation and 3.4 +/- 2.1 mV and 0.6 +/- 0.3 V at 0.5 ms at follow-up. Ventricular sensing and pacing thresholds were 12.1 +/- 4.9 and 0.7 +/- 0.4 V at 0.5 ms on implantation and 12.7 +/- 6.1 mV and 0.8 +/- 0.5 V at 0.5 ms at follow-up (P = NS). No adverse events were reported.
CONCLUSION: Select Secure is a promising system for selective-site pacing in children.

Entities:  

Mesh:

Year:  2009        PMID: 19269985     DOI: 10.1093/europace/eup058

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  3 in total

Review 1.  Tricuspid valve incompetence following implantation of ventricular leads.

Authors:  Giselle A Baquero; Jerry Luck; Gerald V Naccarelli; Mario D Gonzalez; Javier E Banchs
Journal:  Curr Heart Fail Rep       Date:  2015-04

Review 2.  Pacing device therapy in infants and children: a review.

Authors:  Daiji Takeuchi; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-10-27       Impact factor: 1.731

3.  Right ventricular lead implantation facilitated by a guiding sheath in a patient with severe chamber dilatation with tricuspid regurgitation.

Authors:  Phang Boon Lim; David C Lefroy
Journal:  Indian Pacing Electrophysiol J       Date:  2011-10-02