Literature DB >> 19268852

Detection of stent fractures: a comparison of 64-slice CT, conventional cine-angiography, and intravascular ultrasonography.

John H Pang1, David Kim, Nirat Beohar, Sheridan N Meyers, Donald Lloyd-Jones, Vahid Yaghmai.   

Abstract

RATIONAL AND
OBJECTIVES: We evaluated the ability of 64-slice computed tomography (CT), conventional cine-angiography, and intravascular ultrasound (IVUS) to detect stent fractures under ideal conditions. Coronary stent fracture has been implicated as one of the mechanisms of stent thrombosis and, perhaps, in-stent restenosis. However, the preferred imaging modality in detecting fractures in coronary stents has not been well established.
MATERIALS AND METHODS: Four different types of commonly used coronary stents (Cypher, Taxus, Vision, Hepacoat) each with three strut fractures (Cypher, 5; Taxus, 5; Vision, 4; Hepacoat, 5) were nominally deployed in polyurethane tubes and imaged with 64-slice CT, conventional cine-angiography, and IVUS. For each stent type, an unfractured control stent was also imaged.
RESULTS: Overall accuracy (84.1% vs. 73.9%), sensitivity (80.7 vs. 77.2%), and specificity (100% vs. 58.3%) for stent fracture detection was higher with 64 multislice CT compared to conventional cine-angiography. Stent fractures were not accurately detected by IVUS. Fracture detection by multislice CT was best when the stents were imaged at 45 degrees to the z-axis.
CONCLUSIONS: Under ideal in vitro conditions, CT has a high accuracy when used to evaluate coronary stent fractures. The overall accuracy, sensitivity, and specificity of detecting stent fractures are lower by conventional cine-angiography. Stent fractures were not detected using IVUS.

Entities:  

Mesh:

Year:  2009        PMID: 19268852     DOI: 10.1016/j.acra.2008.10.010

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  10 in total

1.  Prevalence and prognosis of coronary stent gap detected by multi-detector CT: a follow-up study.

Authors:  Xinghua Zhang; Li Yang; Haiyue Ju; Fan Zhang; Jian Wu; Bin He; Yundai Chen
Journal:  Eur Radiol       Date:  2012-04-27       Impact factor: 5.315

Review 2.  Are we aware of stent fracture?

Authors:  R D Acar; M Bulut; M Akcakoyun
Journal:  Herz       Date:  2013-10-25       Impact factor: 1.443

3.  Coronary in-stent restenosis: predisposing clinical and stent-related factors, diagnostic performance and analyses of inaccuracies in 320-row computed tomography angiography.

Authors:  Yung-Liang Wan; Pei-Kwei Tsay; Chun-Chi Chen; Yu-Hsiang Juan; Yu-Chieh Huang; Wen-Hui Chan; Ming-Shien Wen; I-Chang Hsieh
Journal:  Int J Cardiovasc Imaging       Date:  2016-04-22       Impact factor: 2.357

4.  Impact of sirolimus-eluting stent fractures without early cardiac events on long-term clinical outcomes: a multislice computed tomography study.

Authors:  Tsuyoshi Ito; Masashi Kimura; Mariko Ehara; Mitsuyasu Terashima; Kenya Nasu; Yoshihisa Kinoshita; Maoto Habara; Etsuo Tsuchikane; Takahiko Suzuki
Journal:  Eur Radiol       Date:  2014-02-25       Impact factor: 5.315

5.  Recurrent episode of very late stent thrombosis in a patient who underwent percutaneous coronary intervention with sirolimus-eluting stent: Same or different mechanism?

Authors:  Soon Yong Suh; Woong Chol Kang; Taehoon Ahn
Journal:  J Cardiol Cases       Date:  2011-12-06

6.  CT Imaging of Coronary Stents: Past, Present, and Future.

Authors:  Andreas H Mahnken
Journal:  ISRN Cardiol       Date:  2012-09-11

Review 7.  Intravascular imaging in coronary stent restenosis: Prevention, characterization, and management.

Authors:  Amr Abouelnour; Tommaso Gori
Journal:  Front Cardiovasc Med       Date:  2022-08-09

8.  Incidence of coronary drug-eluting stent fracture: A systematic review and meta-analysis.

Authors:  Yang Chen; Dandan Li; Yanhui Liao; Xiongda Yao; Yuehua Ruan; Kai Zou; Hanhui Liao; Jingwen Ding; Hao Qin; Zuozhong Yu; Yuanbin Zhao; Longlong Hu; Renqiang Yang
Journal:  Front Cardiovasc Med       Date:  2022-08-23

9.  Stent fracture: how frequently is it recognized?

Authors:  Mohammed Khalil Mohsen; Awad Alqahtani; Jassim Al Suwaidi
Journal:  Heart Views       Date:  2013-04

Review 10.  Coronary stent fracture: a recently appreciated phenomenon with clinical relevance.

Authors:  Madjid Chinikar; Parham Sadeghipour
Journal:  Curr Cardiol Rev       Date:  2014-11
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.