Literature DB >> 19263204

Initial clinical experience of remote magnetic navigation system for catheter mapping and ablation of supraventricular tachycardias.

Dongjie Xu1, Bin Yang, Qijun Shan, Jiangang Zou, Minglong Chen, Chun Chen, Xiaofeng Hou, Fengxiang Zhang, Wen-Qi Li, Kejiang Cao, Hung-Fat Tse.   

Abstract

BACKGROUND: A remote magnetic navigation system (MNS) has been developed for mapping and catheter ablation of cardiac arrhythmias. The present study evaluates the safety and feasibility of this system to perform radiofrequency (RF) ablation in patients with supraventricular tachycardias (SVT).
METHODS: A total of 32 patients (22 female; mean age 44 +/- 16 years) with documented SVT underwent mapping and ablation using Helios II (a 4-mm-tip magnetic catheter), under the guidance of the MNS (Niobe II, Stereotaxis, Inc.).
RESULTS: Catheter ablation procedure with MNS was successful in 30/32 (94%) patients including all patients (27/27, 100%) with atrioventricular nodal reentrant tachycardia (AVNRT) and three of five patients (60%) with atrioventricular reentrant tachycardia (AVRT) without any complication. The procedural successful rate in patients with AVNRT was significantly higher than those in patients with AVRT (P < 0.001). Overall, the medium number of RF application using the MNS was 2 (mean 2.7 +/- 1.6, range 1 to 7), and the medium numbers of RF for AVNRT and AVRT were 2 and 3, respectively. There was no significant difference in the mean procedural time between patients with AVNRT and AVRT (126.3 +/- 38.6 vs. 138.0 +/- 40.3 min, P = 0.54). However, the mean fluoroscopy time was significantly shorter in patients with AVNRT than those with AVRT (5.7 +/- 3.0 vs. 16.5 +/- 2.5 min, P < 0.001). Among those patients with AVNRT, the mean procedural time (139.3 +/- 45.0 vs. 112.3 +/- 24.9 min, P = 0.07) and fluoroscopic time (3.2 +/- 1.0 vs. 8.0 +/- 2.2 min, P < 0.001) were shorter for the later 13 patients than the first 14 patients, suggesting a learning curve in using the MNS for RF ablation.
CONCLUSIONS: The Niobe MNS is a new technique that can allow safe and effective remote-controlled navigation and minimize the need for fluoroscopic guidance for ablation catheter of AVNRT. However, further improvement is required to achieve a higher successful rate for treatment of AVRT.

Entities:  

Mesh:

Year:  2009        PMID: 19263204     DOI: 10.1007/s10840-008-9356-5

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  10 in total

1.  Novel, magnetically guided catheter for endocardial mapping and radiofrequency catheter ablation.

Authors:  Mitchell N Faddis; Walter Blume; Jennifer Finney; Andrew Hall; John Rauch; Jon Sell; Kyongtae Ty Bae; Michael Talcott; Bruce Lindsay
Journal:  Circulation       Date:  2002-12-03       Impact factor: 29.690

2.  Perspectives on the development of a magnetic navigation system for remote control of electrophysiology catheters.

Authors:  Bruce D Lindsay
Journal:  Europace       Date:  2006-03-17       Impact factor: 5.214

3.  Remote-controlled catheter ablation of accessory pathways: results from the magnetic laboratory.

Authors:  Julian Kyoung-Ryul Chun; Sabine Ernst; Shibu Matthews; Boris Schmidt; Dietmar Bansch; Sigrid Boczor; Amaar Ujeyl; Matthias Antz; Feifan Ouyang; Karl-Heinz Kuck
Journal:  Eur Heart J       Date:  2007-01-11       Impact factor: 29.983

Review 4.  Catheter ablation of paroxysmal supraventricular tachycardia.

Authors:  Hiroshi Nakagawa; Warren M Jackman
Journal:  Circulation       Date:  2007-11-20       Impact factor: 29.690

5.  Radiofrequency ablation of atrioventricular nodal reentrant tachycardia using a novel magnetic guidance system compared with a conventional approach.

Authors:  Roger Kerzner; José Mauricio Sánchez; Judy L Osborn; Jane Chen; Mitchell N Faddis; Marye J Gleva; Bruce D Lindsay; Timothy W Smith
Journal:  Heart Rhythm       Date:  2006-03       Impact factor: 6.343

6.  Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis.

Authors:  Efstathios P Efstathopoulos; Demosthenes G Katritsis; Sofia Kottou; Nectarios Kalivas; Efthalia Tzanalaridou; Eleftherios Giazitzoglou; Socrates Korovesis; Keith Faulkner
Journal:  Europace       Date:  2006-05-11       Impact factor: 5.214

7.  Robotic magnetic navigation for atrial fibrillation ablation.

Authors:  Carlo Pappone; Gabriele Vicedomini; Francesco Manguso; Filippo Gugliotta; Patrizio Mazzone; Simone Gulletta; Nicoleta Sora; Simone Sala; Alessandra Marzi; Giuseppe Augello; Laura Livolsi; Andreina Santagostino; Vincenzo Santinelli
Journal:  J Am Coll Cardiol       Date:  2006-03-15       Impact factor: 24.094

8.  Magnetic guidance system for cardiac electrophysiology: a prospective trial of safety and efficacy in humans.

Authors:  Mitchell N Faddis; Jane Chen; Judy Osborn; Michael Talcott; Michael E Cain; Bruce D Lindsay
Journal:  J Am Coll Cardiol       Date:  2003-12-03       Impact factor: 24.094

9.  Initial experience with remote catheter ablation using a novel magnetic navigation system: magnetic remote catheter ablation.

Authors:  Sabine Ernst; Feifan Ouyang; Christian Linder; Klaus Hertting; Fabian Stahl; Julian Chun; Hitoshi Hachiya; Dietmar Bänsch; Matthias Antz; Karl-Heinz Kuck
Journal:  Circulation       Date:  2004-03-15       Impact factor: 29.690

10.  Magnetic navigation in left-sided AV reentrant tachycardias: preliminary results of a retrograde approach.

Authors:  Andrew S Thornton; Maximo Rivero-Ayerza; Paul Knops; Luc J Jordaens
Journal:  J Cardiovasc Electrophysiol       Date:  2007-03-06
  10 in total
  3 in total

1.  Forces on cardiac implantable electronic devices during remote magnetic navigation.

Authors:  C Jilek; C Lennerz; B Stracke; H Badran; V Semmler; T Reents; S Ammar; S Fichtner; B Haller; G Hessling; I Deisenhofer; C Kolb
Journal:  Clin Res Cardiol       Date:  2012-09-29       Impact factor: 5.460

2.  Long-term outcomes of remote magnetic navigation for ablation of supraventricular tachycardias.

Authors:  Sung-Hwan Kim; Yong-Seog Oh; Dong-Hwi Kim; Ik Jun Choi; Tae-Seok Kim; Woo-Seung Shin; Ji-Hoon Kim; Sung-Won Jang; Man Young Lee; Tai-Ho Rho
Journal:  J Interv Card Electrophysiol       Date:  2015-03-18       Impact factor: 1.900

Review 3.  Safety and efficacy of the remote magnetic navigation for ablation of ventricular tachycardias--a systematic review.

Authors:  Ferdi Akca; Ibrahim Önsesveren; Luc Jordaens; Tamas Szili-Torok
Journal:  J Interv Card Electrophysiol       Date:  2011-12-20       Impact factor: 1.900