MRI dynamic range and its compatibility with signal transmission media.

Refaat E Gabr1, Michael Schär, Arthur D Edelstein, Dara L Kraitchman, Paul A Bottomley, William A Edelstein.   

Abstract

As the number of MRI phased array coil elements grows, interactions among cables connecting them to the system receiver become increasingly problematic. Fiber optic or wireless links would reduce electromagnetic interference, but their dynamic range (DR) is generally less than that of coaxial cables. Raw MRI signals, however, have a large DR because of the high signal amplitude near the center of k-space. Here, we study DR in MRI in order to determine the compatibility of MRI multicoil imaging with non-coaxial cable signal transmission. Since raw signal data are routinely discarded, we have developed an improved method for estimating the DR of MRI signals from conventional magnitude images. Our results indicate that the DR of typical surface coil signals at 3T for human subjects is less than 88 dB, even for three-dimensional acquisition protocols. Cardiac and spine coil arrays had a maximum DR of less than 75 dB and head coil arrays less than 88 dB. The DR derived from magnitude images is in good agreement with that measured from raw data. The results suggest that current analog fiber optic links, with a spurious-free DR of 60-70 dB at 500 kHz bandwidth, are not by themselves adequate for transmitting MRI data from volume or array coils with DR approximately 90 dB. However, combining analog links with signal compression might make non-coaxial cable signal transmission viable.

Entities:  

Mesh:

Year:  2009        PMID: 19251444      PMCID: PMC2873084          DOI: 10.1016/j.jmr.2009.01.037

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  24 in total

1.  A comparison of two methods for measuring the signal to noise ratio on MR images.

Authors:  M J Firbank; A Coulthard; R M Harrison; E D Williams
Journal:  Phys Med Biol       Date:  1999-12       Impact factor: 3.609

2.  64-channel array coil for single echo acquisition magnetic resonance imaging.

Authors:  Mary Preston McDougall; Steven M Wright
Journal:  Magn Reson Med       Date:  2005-08       Impact factor: 4.668

3.  A direct modulated optical link for MRI RF receive coil interconnection.

Authors:  Jing Yuan; Juan Wei; G X Shen
Journal:  J Magn Reson       Date:  2007-08-29       Impact factor: 2.229

4.  128-channel body MRI with a flexible high-density receiver-coil array.

Authors:  Christopher J Hardy; Randy O Giaquinto; Joseph E Piel; Kenneth W Rohling; Luca Marinelli; Daniel J Blezek; Eric W Fiveland; Robert D Darrow; Thomas K F Foo
Journal:  J Magn Reson Imaging       Date:  2008-11       Impact factor: 4.813

5.  Maximum-likelihood estimation of Rician distribution parameters.

Authors:  J Sijbers; A J den Dekker; P Scheunders; D Van Dyck
Journal:  IEEE Trans Med Imaging       Date:  1998-06       Impact factor: 10.048

6.  On the Rician distribution of noisy MRI data.

Authors:  A H Andersen
Journal:  Magn Reson Med       Date:  1996-08       Impact factor: 4.668

7.  Estimation of the noise in magnitude MR images.

Authors:  J Sijbers; A J den Dekker; J Van Audekerke; M Verhoye; D Van Dyck
Journal:  Magn Reson Imaging       Date:  1998       Impact factor: 2.546

8.  Measuring signal-to-noise ratios in MR imaging.

Authors:  L Kaufman; D M Kramer; L E Crooks; D A Ortendahl
Journal:  Radiology       Date:  1989-10       Impact factor: 11.105

9.  Dynamic range compression in MRI by means of a nonlinear gradient pulse.

Authors:  V J Wedeen; Y S Chao; J L Ackerman
Journal:  Magn Reson Med       Date:  1988-03       Impact factor: 4.668

10.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

View more
  3 in total

1.  MRI: time is dose--and money and versatility.

Authors:  William A Edelstein; Mahadevappa Mahesh; John A Carrino
Journal:  J Am Coll Radiol       Date:  2010-08       Impact factor: 5.532

2.  Identification and mitigation of interference sources present in SSB-based wireless MRI receiver arrays.

Authors:  Matthew J Riffe; Michael D Twieg; Natalia Gudino; Colin J Blumenthal; Jeremiah A Heilman; Mark A Griswold
Journal:  Magn Reson Med       Date:  2013-02-14       Impact factor: 4.668

3.  Real-time device tracking under MRI using an acousto-optic active marker.

Authors:  Yusuf S Yaras; Dursun Korel Yildirim; Daniel A Herzka; Toby Rogers; Adrienne E Campbell-Washburn; Robert J Lederman; F Levent Degertekin; Ozgur Kocaturk
Journal:  Magn Reson Med       Date:  2020-12-21       Impact factor: 3.737

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.