| Literature DB >> 19242874 |
Lorena Bouzas1, Jesús Hermida, J Carlos Tutor.
Abstract
BACKGROUND: Although high-performance liquid chromatography (HPLC) is the method of choice for blood sirolimus determination, the microparticle enzyme immunoassay (MEIA) run on the IMx analyser is widely used in therapeutic monitoring of this immunosuppressant agent. The aim of our study was to evaluate the possible determination of sirolimus using the fluorescence polarization immunoassay (FPIA) commercialized for everolimus quantification.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19242874 PMCID: PMC2852751 DOI: 10.1080/03009730802608254
Source DB: PubMed Journal: Ups J Med Sci ISSN: 0300-9734 Impact factor: 2.384
Between-run imprecision study for sirolimus determination.
| Mean±SD (µg/L) | CV (%) | ||
|---|---|---|---|
| Innofluor® FPIA | |||
| IMx® Sirolimus control L | 30 | 5.41±0.72 | 13.3 |
| IMx® Sirolimus control M | 30 | 10.58±1.02 | 9.7 |
| IMx® Sirolimus control H | 30 | 21.03±2.08 | 9.9 |
| MEIA | |||
| IMx® Sirolimus control L | 100 | 4.80±0.41 | 8.6 |
| IMx® Sirolimus control M | 100 | 10.30±0.69 | 6.7 |
| IMx® Sirolimus control H | 100 | 22.28±1.56 | 7.0 |
| HPLC | |||
| IMx® Sirolimus control L | 22 | 4.96±0.59 | 11.9 |
| IMx® Sirolimus control M | 22 | 11.09±0.81 | 7.3 |
| IMx® Sirolimus control H | 22 | 22.50±1.98 | 8.8 |
Assigned values: control L = 5.0 µg/L; control M = 11.0 µg/L; control H = 22.0 µg/L.
Figure 1.Correlation and regression (A, C, E) and difference plots (B, D, F) for sirolimus concentrations using the MEIA, FPIA, and HPLC in blood samples from liver (○) and kidney (•) transplant recipient patients. The dotted lines correspond to the limits of the acceptance criterion for deviation.
Sirolimus clearance (CL) and sirolimus concentration ratios in liver and kidney transplant recipients.
| Liver transplant ( | Kidney transplant ( | |
|---|---|---|
| Sirolimus CL (L/h) a | 3.70±2.73 (2.61) | 3.46±1.37 (3.19) |
| FPIA/HPLC ratio | 1.21±0.36 (1.18) | 1.19±0.21 (1.19) |
| MEIA/HPLC ratio | 1.14±0.23 (1.14) | 1.18±0.18 (1.15) |
| FPIA/MEIA ratio | 1.03±0.14 (1.06) | 1.03±0.14 (1.02) |
aCalculated from Css determined by HPLC.