[Comparison of high-resolution liquid chromatography versus microparticle enzyme immunoassay for the measurement of sirolimus levels in renal transplantation].

R Asensi Díez1, M P Quesada Sanz, E Valverde Molina, E de Ramón Garrido, J Gómez Maldonado, I Muñoz Castillo.   

Abstract

OBJECTIVE: To compare sirolimus levels measured in whole blood using two analytical techniques: high-resolution liquid chromatography and microparticle enzyme immunoassay, and to evaluate whether hemoglobin, hematocrit, and time from kidney transplantation influence results obtained using the immune-enzymatic technique.
METHOD: A retrospective, observational study in which all transplanted patients with at least one measurement of sirolimus levels using high-resolution liquid chromatography or microparticle enzyme immunoassay from October 2004 to May 2005 were consecutively included. For statistical comparisons simple linear regression, ANCOVA, intra-class correlation coefficient, and the method of agreement limits were all used.
RESULTS: Ninety-one patients were assessed for a total of 307 measurements (median: 2, inter-quartile range: 1-4, range: 1-15) of sirolimus levels. The straight-line equation using the linear regression analysis was as follows: MEIA = 0.70 (95% CI: 0.39-1.01) + 1.14 (95% CI: 1.10-1.17) x HPLC/UV. The intra-class correlation coefficient between both measurements was 0.955 (95% CI 0.944-0.964). Mean overestimation using enzyme immunoassay was 24.8% +/- 19.4%. Difference in means between both measurements was 1.9 +/- 1.3 ng/mL. Agreement limits were established between -0.8 ng/mL (95% CI: -1.05; -0.55) and +4.6 ng/mL (95% CI: 4.35; 4.85). Factors such as post-transplant time, hemoglobin, and hematocrit did not influence overestimates obtained using enzyme immunoassays. These results were not influenced by non-independence in measurements.
CONCLUSIONS: Despite enzyme immunoassay overestimates in establishing sirolimus levels in whole blood, its correlation with chromatography is acceptable. Added to its benefits versus chromatographic techniques, this renders enzyme immunoassay a good alternative for the measurement of sirolimus levels in whole blood.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16999560     DOI: 10.1016/s1130-6343(06)73964-2

Source DB:  PubMed          Journal:  Farm Hosp        ISSN: 1130-6343


  2 in total

1.  Development and validation of an HPLC method for the analysis of sirolimus in drug products.

Authors:  Ziba Islambulchilar; Saeed Ghanbarzadeh; Shahram Emami; Hadi Valizadeh; Parvin Zakeri-Milani
Journal:  Adv Pharm Bull       Date:  2012-05-20

2.  Determination of blood sirolimus concentrations in liver and kidney transplant recipients using the Innofluor fluorescence polarization immunoassay: comparison with the microparticle enzyme immunoassay and high-performance liquid chromatography-ultraviolet method.

Authors:  Lorena Bouzas; Jesús Hermida; J Carlos Tutor
Journal:  Ups J Med Sci       Date:  2009       Impact factor: 2.384