Evaluation of an immunoassay (Abbott-IMX Analyzer) allowing routine determination of sirolimus: comparison with LC-MS method.

A S Bargnoux1, A Bonardet, G Chong, V Garrigue, S Deleuze, A M Dupuy, G Mourad, J P Cristol.   

Abstract

BACKGROUND: The use of the immunosuppressive agent sirolimus is increasing in renal transplantation but its monitoring often requires high-performance liquid chromatography (HPLC) with ultra-violet (UV) or tandem mass spectrometric (MS-MS) detection. The aim of this study was to compare a new microparticle enzyme immunoassay (MEIA, Microparticle Enzyme Immunoassay) on IMx Abbott Analyser with a liquid chromatography-mass spectometry (LC-MS) method.
METHOD: The accuracy of immunoassay analytical performance including within run and between run imprecision and linearity was tested. For comparison studies, sirolimus level was then determined with the two methods on 98 samples from 52 transplant patients.
RESULTS: Total intra-assay and inter-assay variation coefficients were below 10% at the three levels tested, and the coefficient of linearity was r = 0.99. The values obtained were highly correlated with the LC-MS method (MEIA = 1.02LC-MS + 0.91; r(2) = 0.87). As a result, the immunoassay showed good performance, and clinical sample measurements were not affected by the method. The MEIA may be a useful alternative for routine monitoring of sirolimus.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16980087     DOI: 10.1016/j.transproceed.2006.06.127

Source DB:  PubMed          Journal:  Transplant Proc        ISSN: 0041-1345            Impact factor:   1.066


  4 in total

1.  The effects of lyophilization on the physico-chemical stability of sirolimus liposomes.

Authors:  Saeed Ghanbarzadeh; Hadi Valizadeh; Parvin Zakeri-Milani
Journal:  Adv Pharm Bull       Date:  2013-02-07

2.  Protein biochip array technology to monitor rituximab in rheumatoid arthritis.

Authors:  S Fabre; C Guisset; L Tatem; N Dossat; A M Dupuy; J D Cohen; J P Cristol; J P Daures; C Jorgensen
Journal:  Clin Exp Immunol       Date:  2009-03       Impact factor: 4.330

3.  Protein biochip array technology for cytokine profiling predicts etanercept responsiveness in rheumatoid arthritis.

Authors:  S Fabre; A M Dupuy; N Dossat; C Guisset; J D Cohen; J P Cristol; J P Daures; C Jorgensen
Journal:  Clin Exp Immunol       Date:  2008-06-18       Impact factor: 4.330

4.  Determination of blood sirolimus concentrations in liver and kidney transplant recipients using the Innofluor fluorescence polarization immunoassay: comparison with the microparticle enzyme immunoassay and high-performance liquid chromatography-ultraviolet method.

Authors:  Lorena Bouzas; Jesús Hermida; J Carlos Tutor
Journal:  Ups J Med Sci       Date:  2009       Impact factor: 2.384

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.