Literature DB >> 19194524

A Multiclass Model Observer for Multislice-Multiview Images.

H C Gifford1, A Lehovich, M A King.   

Abstract

A human-model observer for tumor detection-localization studies featuring multislice-multiview (or volumetric) image displays has been introduced. This volumetric observer, an extension of multiclass linear observers previously tested with single-slice and multislice displays, produces rating and localization data by integrating perception measurements from the different image views. A channelized NPW (CNPW) version of the observer was evaluated against humans for a background-known-exactly (BKE) detection task involving localization of Tc-99m Neotect lesions in simulated SPECT lung images. An LROC study evaluated two RBI reconstruction strategies that used different combinations of corrections for attenuation, scatter, and distance-dependent system resolution, and coronal, sagittal, and transverse slices were presented to the observers. Model-observer ranking of these strategies did not match that of the humans. Follow-up studies exploring several possible remedies for the model observer, including strategy-specific search regions and an internal-noise mechanism, showed little change. Future work will examine variations from the BKE assumption as a means of reconciling the rankings.

Entities:  

Year:  2006        PMID: 19194524      PMCID: PMC2633942          DOI: 10.1109/NSSMIC.2006.354223

Source DB:  PubMed          Journal:  IEEE Nucl Sci Symp Conf Rec (1997)        ISSN: 1095-7863


  5 in total

1.  A comparison of human and model observers in multislice LROC studies.

Authors:  Howard C Gifford; Michael A King; P Hendrik Pretorius; R Glenn Wells
Journal:  IEEE Trans Med Imaging       Date:  2005-02       Impact factor: 10.048

2.  Block-iterative methods for image reconstruction from projections.

Authors:  C L Byrne
Journal:  IEEE Trans Image Process       Date:  1996       Impact factor: 10.856

3.  A practical method for position-dependent Compton-scatter correction in single photon emission CT.

Authors:  K Ogawa; Y Harata; T Ichihara; A Kubo; S Hashimoto
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

4.  A Monte Carlo program for the simulation of scintillation camera characteristics.

Authors:  M Ljungberg; S E Strand
Journal:  Comput Methods Programs Biomed       Date:  1989-08       Impact factor: 5.428

5.  The utility of a somatostatin-type receptor binding peptide radiopharmaceutical (P829) in the evaluation of solitary pulmonary nodules.

Authors:  J E Blum; H Handmaker; N A Rinne
Journal:  Chest       Date:  1999-01       Impact factor: 9.410

  5 in total
  3 in total

1.  A visual-search model observer for multislice-multiview SPECT images.

Authors:  Howard C Gifford
Journal:  Med Phys       Date:  2013-09       Impact factor: 4.071

2.  Regularization design in penalized maximum-likelihood image reconstruction for lesion detection in 3D PET.

Authors:  Li Yang; Jian Zhou; Andrea Ferrero; Ramsey D Badawi; Jinyi Qi
Journal:  Phys Med Biol       Date:  2013-12-19       Impact factor: 3.609

3.  Factors Influencing Lesion Detection in SPECT Lung Images.

Authors:  H C Gifford; X M Zheng; R Licho; P H Pretorius; P B Schneider; P H Simkin; M A King
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2006
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.