Development of a compact wearable pneumatic drive unit for a ventricular assist device.

Akihiko Homma1, Yoshiyuki Taenaka, Eisuke Tatsumi, Eiki Akagawa, Hwansung Lee, Tomohiro Nishinaka, Yoshiaki Takewa, Toshihide Mizuno, Tomonori Tsukiya, Yukihide Kakuta, Nobumasa Katagiri, Isao Shimosaki, Shigeru Hamada, Hiroshi Mukaibayashi, Wataru Iwaoka.   

Abstract

The purpose of this study was to develop a compact wearable pneumatic drive unit for a ventricular assist device (VAD). This newly developed drive unit, 20 x 8.5 x 20 cm in size and weighing approximately 1.8 kg, consists of a brushless DC motor, noncircular gears, a crankshaft, a cylinder-piston, and air pressure regulation valves. The driving air pressure is generated by the reciprocating motion of the piston and is controlled by the air pressure regulation valves. The systolic ratio is determined by the noncircular gears, and so is fixed for a given configuration. As a result of an overflow-type mock circulation test, a drive unit with a 44% systolic ratio connected to a Toyobo VAD blood pump with a 70-ml stroke volume achieved a pump output of more than 7 l/min at 100 bpm against a 120 mmHg afterload. Long-term animal tests were also performed using drive units with systolic ratios of 45% and 53% in two Holstein calves weighing 62 kg and 74 kg; the tests were terminated on days 30 and 39, respectively, without any malfunction. The mean aortic pressure, bypass flow, and power consumption for the first calf were maintained at 90 x 13 mmHg, 3.9 x 0.9 l/min, and 12 x 1 W, and those for the second calf were maintained at 88 x 13 mmHg, 5.0 x 0.5 l/min, and 16 x 2 W, respectively. These results indicate that the newly developed drive unit may be used as a wearable pneumatic drive unit for the Toyobo VAD blood pump.

Entities:  

Mesh:

Year:  2008        PMID: 19184282     DOI: 10.1007/s10047-008-0433-4

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  22 in total

Review 1.  Mechanical circulatory support devices (MCSD) in Japan: current status and future directions.

Authors:  Setsuo Takatani; Hikaru Matsuda; Akihisa Hanatani; Chisato Nojiri; Kenji Yamazaki; Tadashi Motomura; Katsuhiro Ohuchi; Tohru Sakamoto; Takashi Yamane
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

2.  A servomatic pneumatic driver system for left ventricular assist devices.

Authors:  T Takahama; F Kanai; K Onishi; M Hiraishi; Z Yamazaki; Y Naruse; A Furuse; T Togawa; T Yoshitake
Journal:  ASAIO Trans       Date:  1991 Jul-Sep

3.  Development of a compact portable driver for a pneumatic ventricular assist device.

Authors:  Tomohiro Nishinaka; Yoshiyuki Taenaka; Eisuke Tatsumi; Hiroyuki Ohnishi; Akihiko Homma; Kyoko Shioya; Toshihide Mizuno; Tomonori Tsukiya; Sadahiko Mushika; Yasuhiro Hashiguchi; Akira Suzuki; Soichiro Kitamura
Journal:  J Artif Organs       Date:  2007-12-20       Impact factor: 1.731

4.  A linear motor and compact cylinder-piston driver for left ventricular bypass.

Authors:  K X Qian
Journal:  J Biomed Eng       Date:  1990-01

5.  An electromagnetic pneumatic blood pump driver.

Authors:  R L Whalen; R N Briskman
Journal:  ASAIO Trans       Date:  1988 Jul-Sep

6.  Completely integrated wearable TAH-drive unit.

Authors:  H P Heimes; F Klasen
Journal:  Int J Artif Organs       Date:  1982-05       Impact factor: 1.595

7.  Multi-institutional studies of the National Cardiovascular Center Ventricular Assist System: use in 92 patients.

Authors:  H Takano; Y Taenaka; H Noda; M Kinoshita; A Yagura; E Tatsumi; H Sekii; E Sasaki; M Umezu; T Nakatani
Journal:  ASAIO Trans       Date:  1989 Jul-Sep

8.  Ventricular assist systems: experience in Japan with Toyobo pump and Zeon pump.

Authors:  H Takano; T Nakatani
Journal:  Ann Thorac Surg       Date:  1996-01       Impact factor: 4.330

9.  The Thoratec mobile computer: initial in-hospital and outpatient experience.

Authors:  Michael A Sobieski; Terri A George; Mark S Slaughter
Journal:  ASAIO J       Date:  2004 Jul-Aug       Impact factor: 2.872

10.  Home discharge experience with the Thoratec TLC-II portable driver.

Authors:  Mark S Slaughter; Michael A Sobieski; Michele Martin; Muhyaldeen Dia; Marc A Silver
Journal:  ASAIO J       Date:  2007 Mar-Apr       Impact factor: 2.872

View more
  3 in total

Review 1.  Journal of Artificial Organs 2008: the year in review.

Authors:  Y Sawa; E Tatsumi; A Funakubo; T Horiuchi; K Iwasaki; A Kishida; T Masuzawa; K Matsuda; M Nishimura; T Nishimura; Y Tomizawa; T Yamaoka; H Watanabe
Journal:  J Artif Organs       Date:  2009-03-29       Impact factor: 1.731

2.  Development of a flow rate monitoring method for the wearable ventricular assist device driver.

Authors:  Kentaro Ohnuma; Akihiko Homma; Hirohito Sumikura; Tomonori Tsukiya; Yoshiaki Takewa; Toshihide Mizuno; Hiroshi Mukaibayashi; Koichi Kojima; Kazuo Katano; Yoshiyuki Taenaka; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2014-12-13       Impact factor: 1.731

3.  Biventricular support using a centrifugal pump in a 6 year old with fulminant myocarditis.

Authors:  Hiromu Kehara; Tamaki Takano; Takamitsu Terasaki; Kenji Okada
Journal:  J Artif Organs       Date:  2016-12-01       Impact factor: 1.731

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.