Completely integrated wearable TAH-drive unit.

H P Heimes, F Klasen.   

Abstract

A pneumatic, wearable, batter-powered TAH-drive unit is presented. This drive-unit is characterized by a very good overall efficiency, achieved by design optimization on one hand and controlled economized membrane displacement on the other. The implemented control system facilitates an adoption to circulatory needs. The integrated micro-computer controls the whole pumping cycle and calculates the effective blood-side stroke volume by a beat to beat analysis of drive-side parameters. It could be demonstrated in animal tests (calves) under varying loads that the available driving power is sufficient to maintain an adequate circulatory supply for calves up to 135 kg. Further, it could be demonstrated that a sufficient right/left balance was obtained over the whole animal test duration.

Entities:  

Mesh:

Year:  1982        PMID: 7107050

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  1 in total

1.  Development of a compact wearable pneumatic drive unit for a ventricular assist device.

Authors:  Akihiko Homma; Yoshiyuki Taenaka; Eisuke Tatsumi; Eiki Akagawa; Hwansung Lee; Tomohiro Nishinaka; Yoshiaki Takewa; Toshihide Mizuno; Tomonori Tsukiya; Yukihide Kakuta; Nobumasa Katagiri; Isao Shimosaki; Shigeru Hamada; Hiroshi Mukaibayashi; Wataru Iwaoka