Absorbed and effective doses from cone beam volumetric imaging for implant planning.

T Okano1, Y Harata, Y Sugihara, R Sakaino, R Tsuchida, K Iwai, K Seki, K Araki.   

Abstract

OBJECTIVES: Volumetric CT using a cone beam has been developed by several manufacturers for dentomaxillofacial imaging. The purpose of this study was to measure doses for implant planning with cone beam volumetric imaging (CBVI) in comparison with conventional multidetector CT (MDCT).
METHODS: The two CBVI systems used were a 3D Accuitomo (J. Morita), including an image-intensifier type (II) and a flat-panel type (FPD), and a CB MercuRay (Hitachi). The 3D Accuitomo operated at 80 kV, 5 mA and 18 s. The CB MercuRay operated at 120 kV, 15 mA, 9.8 s. The MDCT used was a HiSpeed QX/i (GE), operated at 120 kV, 100 mA and 0.7 s, and its scan length was 77 mm for both jaws. Measurement of the absorbed tissue and organ doses was performed with an Alderson phantom, embedding the radiophotoluminescence glass dosemeter into the organs/tissues. The values obtained were converted into the absorbed dose. The effective dose as defined by the International Commission on Radiological Protection was then calculated.
RESULTS: The absorbed doses of the 3D Accuitomo of the organs in the primary beam ranged from 1-5 mGy, and were several to ten times lower than other doses. The effective dose of the 3D Accuitomo ranged from 18 muSv to 66 muSv, and was an order of magnitude smaller than the others. In conclusion, these results show that the dose in the 3D Accuitomo is lower than the CB MercuRay and much less than MDCT.

Mesh:

Substances:

Year:  2009        PMID: 19176649     DOI: 10.1259/dmfr/14769929

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  35 in total

1.  Radiographic evaluation on prevalence of Stafne bone defect: a study from two centres in Turkey.

Authors:  Y Sisman; O Miloglu; A E Sekerci; A B Yilmaz; O Demirtas; T T Tokmak
Journal:  Dentomaxillofac Radiol       Date:  2011-11-10       Impact factor: 2.419

2.  Effect of electric potential and current on mandibular linear measurements in cone beam CT.

Authors:  S Panmekiate; W Apinhasmit; A Petersson
Journal:  Dentomaxillofac Radiol       Date:  2012-04-12       Impact factor: 2.419

3.  Estimation of paediatric organ and effective doses from dental cone beam CT using anthropomorphic phantoms.

Authors:  C Theodorakou; A Walker; K Horner; R Pauwels; R Bogaerts; R Jacobs
Journal:  Br J Radiol       Date:  2012-02       Impact factor: 3.039

4.  Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields.

Authors:  Anna Kelaranta; Marja Ekholm; Paula Toroi; Mika Kortesniemi
Journal:  Dentomaxillofac Radiol       Date:  2015-08-27       Impact factor: 2.419

5.  Accuracy of anatomical landmark identification using different CBCT- and MSCT-based 3D images: an in vitro study.

Authors:  Jürgen Medelnik; Klaus Hertrich; Stefanie Steinhäuser-Andresen; Ursula Hirschfelder; Elisabeth Hofmann
Journal:  J Orofac Orthop       Date:  2011-09-08       Impact factor: 1.938

Review 6.  Effective dose of cone beam CT (CBCT) of the facial skeleton: a systematic review.

Authors:  A Al-Okshi; C Lindh; H Salé; M Gunnarsson; M Rohlin
Journal:  Br J Radiol       Date:  2015-01       Impact factor: 3.039

7.  Using cone-beam CT as a low-dose 3D imaging technique for the extremities: initial experience in 50 subjects.

Authors:  Ambrose J Huang; Connie Y Chang; Bijoy J Thomas; Peter J MacMahon; William E Palmer
Journal:  Skeletal Radiol       Date:  2015-02-05       Impact factor: 2.199

8.  Assessment of non-invasive chronic fungal rhinosinusitis by cone beam CT: comparison with multidetector CT findings.

Authors:  Tomohiko Yamauchi; Akiko Tani; Shuji Yokoyama; Hiroshi Ogawa
Journal:  Fukushima J Med Sci       Date:  2017-07-26

Review 9.  Clinical guidelines for dental cone-beam computed tomography.

Authors:  Takafumi Hayashi; Yoshinori Arai; Toru Chikui; Sachiko Hayashi-Sakai; Kazuya Honda; Hiroko Indo; Taisuke Kawai; Kaoru Kobayashi; Shumei Murakami; Masako Nagasawa; Munetaka Naitoh; Eiji Nakayama; Yutaka Nikkuni; Hideyoshi Nishiyama; Noriaki Shoji; Shigeaki Suenaga; Ray Tanaka
Journal:  Oral Radiol       Date:  2018-01-11       Impact factor: 1.852

10.  Use of cone beam computed tomography in endodontics.

Authors:  William C Scarfe; Martin D Levin; David Gane; Allan G Farman
Journal:  Int J Dent       Date:  2010-03-31