Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry.

B C Ferreira1, M C Lopes, M Capela.   

Abstract

Gafchromic EBT (EBT) films are becoming increasingly popular due to their advantageous properties. When flatbed colour scanners are used for film dosimetry, a good quality control of the scanning device is a crucial step for accurate results. The proposal of this work was to fully assess the performance of the scanner Epson Expression 10000XL in order to quantify all parameters and needed corrections to minimize dose uncertainties. A standard step tablet, with 32 steps and optical densities from 0.06 to 3.8, was used to check the scanner linearity. The scanner warming-up effect and reproducibility were evaluated by performing 30 consecutive scans plus 20 scans in 15 min intervals. The scanning colour modes: 24 and 48 bits and scanning resolutions from 50 to 300 dpi were tested. A Wiener filter with different pixels regions was applied with the purpose of reducing the film noise. All scans were made in transmission mode with a constant film orientation. The red colour channel was posteriorly extracted from the images to maximize readout sensitivity. Two EBT films were irradiated, perpendicularly and parallel to beam incidence, with a 6 MV photon beam with doses that ranged from 0.2 to 3 Gy. A polynomial expression was used to convert optical density into dose. Dose uncertainty was quantified applying error propagation analysis. A correction for the non-uniform response of the scanner was determined using five films irradiated with a uniform dose. The scanner response was linear until an optical density of approximately 1 which corresponds to doses higher than those of clinical interest for EBT films. The scanner signal stabilized after seven readings. Scanner reproducibility around 0.2% was obtained either with the scanner warm or cold. However, reproducibility was significantly reduced when comparing images digitized with the scanner at different temperatures. Neither the colour depth mode, the scanning resolution, the multiscan option nor the Wiener filter had a significant effect on the shape of the calibration curve. However, a reduction in dose uncertainty was possible by selecting appropriate reading parameters. These are a 48 bit colour depth, a 75 dpi resolution and repeating the scan four times. Finally, the two dimensional Wiener filter applied to a 3 x 3 pixel region to the red component of the image reduced the experimental scan uncertainty to about 0.5% for doses higher than 0.5 Gy. Total scan uncertainty was less than 2% for a perpendicular calibration and reduced to less than 1% for a parallel calibration. A dose over-estimation of around 5% for clinical doses may be made if the image acquired is not corrected for the non-uniform response of the scanner. A protocol to read EBT films using the Epson Expression 10000XL scanner was established for IMRT verification. The contribution for the overall uncertainty in film dosimetry coming from the scanning process was estimated to be around 0.5% for doses higher than 0.5 Gy when reading parameters are optimized. Total scan uncertainty achieved is about 2% when using a perpendicular calibration. It can further be reduced if a parallel calibration is used.

Entities:  

Mesh:

Year:  2009        PMID: 19168937     DOI: 10.1088/0031-9155/54/4/017

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  23 in total

1.  Implementation of radiochromic film dosimetry protocol for volumetric dose assessments to various organs during diagnostic CT procedures.

Authors:  Samuel Brady; Terry Yoshizumi; Greta Toncheva; Donald Frus
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

2.  Evaluation of the absorbed dose to the breast using radiochromic film in a dedicated CT mammotomography system employing a quasi-monochromatic x-ray beam.

Authors:  Dominic J Crotty; Samuel L Brady; D'Vone C Jackson; Greta I Toncheva; Colin E Anderson; Terry T Yoshizumi; Martin P Tornai
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

3.  Air-electron stream interactions during magnetic resonance IGRT : Skin irradiation outside the treatment field during accelerated partial breast irradiation.

Authors:  Jong Min Park; Kyung Hwan Shin; Jung-In Kim; So-Yeon Park; Seung Hyuck Jeon; Noorie Choi; Jin Ho Kim; Hong-Gyun Wu
Journal:  Strahlenther Onkol       Date:  2017-09-15       Impact factor: 3.621

4.  Isodose curve mappings measured while undergoing rotation for quality assurance testing of a 137Cs irradiator.

Authors:  Samuel L Brady; Terry T Yoshizumi; Colin Anderson-Evans; Giao Nguyen
Journal:  Health Phys       Date:  2012-02       Impact factor: 1.316

5.  A feasibility study using radiochromic films for fast neutron 2D passive dosimetry.

Authors:  Samuel L Brady; Rathnayaka Gunasingha; Terry T Yoshizumi; Calvin R Howell; Alexander S Crowell; Brent Fallin; Anton P Tonchev; Mark W Dewhirst
Journal:  Phys Med Biol       Date:  2010-08-06       Impact factor: 3.609

6.  SBRT/SRS patient-specific QA using GAFchromicTM EBT3 and FilmQATM Pro software.

Authors:  Giuseppe Stella; Nina Cavalli; Elisa Bonanno; Lucia Zirone; Giuseppina Rita Borzì; Martina Pace; Andrea Girlando; Anna M Gueli; Carmelo Marino
Journal:  J Radiosurg SBRT       Date:  2022

7.  Application of a color scanner for (60)Co high dose rate brachytherapy dosimetry with EBT radiochromic film.

Authors:  Mahdi Ghorbani; Mohammad Taghi Bahreyni Toossi; Ali Asghar Mowlavi; Shahram Bayani Roodi; Ali Soleimani Meigooni
Journal:  Radiol Oncol       Date:  2012-11-09       Impact factor: 2.991

8.  Clinical commissioning and use of the Novalis Tx linear accelerator for SRS and SBRT.

Authors:  Jinkoo Kim; Ning Wen; Jian-Yue Jin; Nicole Walls; Sangroh Kim; Haisen Li; Lei Ren; Yimei Huang; Anthony Doemer; Kathleen Faber; Tina Kunkel; Ahssan Balawi; Kimberly Garbarino; Kenneth Levin; Samir Patel; Munther Ajlouni; Brett Miller; Teamor Nurushev; Calvin Huntzinger; Raymond Schulz; Indrin J Chetty; Benjamin Movsas; Samuel Ryu
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

9.  The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner.

Authors:  G S Sim; J H D Wong; K H Ng
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

10.  Probability Distribution of Pixel Intensities of EBT3 Films and its Application in the Correction of Uncertainty Budget.

Authors:  Rahul Kumar Chaudhary; Munir Pathan; Rajesh Kumar; S D Sharma; B K Sapra
Journal:  J Med Phys       Date:  2021-05-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.