Hamid Barakpour1, Zahra Emkanjoo, Abolfath Alizadeh, Mohammad Ali Sadr-Ameli.
Abstract
The clinical efficacy of ICD-CRT therapy depends on accurate sensing of intracardiac signals and sensing algorithms. We report the occurrence of sensing abnormality in a patient with ICD-CRT. In this patient, oversensing of myopotentials during strenuous muscular activity resulted in an inappropriate ICD-CRT discharge. Although modern ICDs are highly effective in detecting and terminating malignant tachyarrhythmias, their detection specificity must be improved. It is possible to find the mechanism of arrhythmia by EGM. Simple device reprogramming make it possible to avoid the oversensing of myopotentials.Entities:
Keywords: Implantable defibrillator; Inappropriate shock; myopotential oversensing
Year: 2009 PMID: 19165363 PMCID: PMC2615066
Source DB: PubMed Journal: Indian Pacing Electrophysiol J ISSN: 0972-6292
Figure 1
Figure 2Intracardiac EGMs showing AV nodal jump followed by the initiation of slow-fast AVNRT.