Inappropriate shocks delivered by implantable cardiac defibrillators during oversensing of activity of diaphagmatic muscle.

D Babuty1, L Fauchier, P Cosnay.   

Abstract

Two cases are reported (both men, one 72 and one 54 years old) of inappropriate shocks delivered by an implantable cardiac defibrillator (ICD) device, which oversensed the myopotentials induced by deep breathing and Valsalva manoeuvre. No damage to leads was associated with the oversensing of myopotentials. The mechanism of the inappropriate shocks was determined using real time electrograms. Modification of the duration of ventricular detection and decrease in sensitivity made it possible to avoid the oversensing of myopotentials and to deliver ICD treatment.

Entities:  

Mesh:

Year:  1999        PMID: 10220554      PMCID: PMC1728919          DOI: 10.1136/hrt.81.1.94

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  5 in total

1.  Inappropriate arrhythmia detection in implantable defibrillator therapy due to oversensing of diaphragmatic myopotentials.

Authors:  B Schulte; J Sperzel; J Carlsson; M Dürsch; A Erdogan; H F Pitschner; J Neuzner
Journal:  J Interv Card Electrophysiol       Date:  2001-12       Impact factor: 1.900

2.  Prevalence of sensing abnormalities in dual chamber implantable cardioverter defibrillators.

Authors:  Mohammad Saeed; Anna Jin; Gregory Pontone; Steve Higgins; Michael Gold; David Harari; Steven Nunley; Mark S Link; Munther K Homoud; N A Mark Estes; Paul J Wang
Journal:  Ann Noninvasive Electrocardiol       Date:  2003-07       Impact factor: 1.468

3.  Inappropriate shock delivery and biventricular pacing cardiac defibrillators.

Authors:  Brant C Liu; Rollo P Villareal; Ramesh Hariharan; Abdi Rasekh; Ali Massumi
Journal:  Tex Heart Inst J       Date:  2003

4.  A fortuitous syncope. The pitfalls of integrated bipolar defibrillator leads.

Authors:  Tushar V Salukhe; Ian Wright; Matthew Wright; Prapa Kanagaratnam; Mark D O'Neill
Journal:  Indian Pacing Electrophysiol J       Date:  2008-11-01

5.  Inappropriate Shock delivered by Implantable Cardioverter Defibrillator--Cardiac Resynchronization Therapy (ICD-CRT) due to myopotential oversensing.

Authors:  Hamid Barakpour; Zahra Emkanjoo; Abolfath Alizadeh; Mohammad Ali Sadr-Ameli
Journal:  Indian Pacing Electrophysiol J       Date:  2009-01-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.