Literature DB >> 19119073

Reassessment of radiation risks from electrophysiology procedures compared to coronary angiography.

Ian R Smith1, John T Rivers, John Hayes, Wayne Stafford, Catrina Codd.   

Abstract

BACKGROUND: Electrophysiology (EP) procedures have been reported to carry a significantly greater radiation risk than that of coronary angiography (CA). This is largely due to numerous reports linking severe deterministic radiation effects to long procedure and fluoroscopy times (FTs). This study documents low radiation doses achieved by strategies involving operator training and education as well as equipment and technique optimisation to reduce radiation risks.
METHODS: Records relating to 732 diagnostic EP and 1744 therapeutic EP procedures performed between January 2002 and December 2007 were analysed. Data from 1458 diagnostic only CA procedures performed in 2006 was used for comparison. For each procedure type, FT, number of digital frames acquired and estimated effective dose (E) were compared.
RESULTS: Although the FT for CA procedures is significantly less than for therapeutic EP procedures (FT for diagnostic EP being similar), EP procedures generally are associated with lower E, the exception being procedures for atrial fibrillation (AF).
CONCLUSION: Through the application of a comprehensive exposure minimisation strategy, the radiation risk to patients undergoing diagnostic and, therapeutic EP procedures (except AF ablation procedures) is significantly less than that faced by patients undergoing CA. E, however, is heavily dependent on procedure type and as such care must be taken in undertaking generalised comparisons for audit and benchmarking purposes.

Entities:  

Mesh:

Year:  2008        PMID: 19119073     DOI: 10.1016/j.hlc.2008.10.006

Source DB:  PubMed          Journal:  Heart Lung Circ        ISSN: 1443-9506            Impact factor:   2.975


  4 in total

1.  Relation between projection positions and X-ray radiation doses during placement of the coronary sinus mapping electrode.

Authors:  Gang Lin; Jian-Fei Huang; Hui-He Lu; Lin-Sheng Shi; Jue Wang; Shu-Qing Zhang
Journal:  J Interv Card Electrophysiol       Date:  2012-03-15       Impact factor: 1.900

Review 2.  "Near-zero" fluoroscopic exposure in supraventricular arrhythmia ablation using the EnSite NavX™ mapping system: personal experience and review of the literature.

Authors:  Michela Casella; Gemma Pelargonio; Antonio Dello Russo; Stefania Riva; Stefano Bartoletti; Pasquale Santangeli; Antonio Scarà; Tommaso Sanna; Riccardo Proietti; Luigi Di Biase; G Joseph Gallinghouse; Maria Lucia Narducci; Luigi Sisto; Fulvio Bellocci; Andrea Natale; Claudio Tondo
Journal:  J Interv Card Electrophysiol       Date:  2011-03-02       Impact factor: 1.900

3.  Reduction of radiation exposure in catheter ablation of atrial fibrillation: Lesson learned.

Authors:  Roberto De Ponti
Journal:  World J Cardiol       Date:  2015-08-26

4.  Catheter Ablation of Right-Sided Accessory Pathways in Adults Using the Three-Dimensional Mapping System: A Randomized Comparison to the Conventional Approach.

Authors:  Yuedong Ma; Jia Qiu; Yang Yang; Anli Tang
Journal:  PLoS One       Date:  2015-06-17       Impact factor: 3.240

  4 in total