AIMS: The size of stent cells is an important design parameter influencing the resulting wall coverage and strut positioning of several bifurcation stenting techniques. Therefore, we report accurate geometrical information on stent cell size of five different stents with the key purpose to assist with stent selection for bifurcation lesions. METHODS AND
RESULTS: We developed an innovative technique to assess cell size with high accuracy based on micro-CT images. In addition, the size of the ostium was calculated for several idealised bifurcations with a cylindrical main and side branch. The cell circumferences of the investigated stents vary substantially, with values between 9.5 and 19.8 mm (or an equivalent maximal cell diameter between 3.0 and 6.3 mm). Comparison of these cell circumferences with the calculated ostium circumferences shows which stents should preferably be used for which bifurcations, as these stents have cells that can be sufficiently enlarged by dilating through the side.
CONCLUSIONS: The cell sizes of the investigated stents differ considerably. This should be taken into account when applying a particular bifurcation stenting technique. The risk on limited main or side branch patency due to obstructing stent struts can be minimised by using the reported data for stent selection.

Entities:  

Mesh:

Year:  2008        PMID: 19110814     DOI: 10.4244/eijv4i3a68

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  8 in total

1.  Comparison of plaque prolapse in consecutive patients treated with Xience V and Taxus Liberte stents.

Authors:  Zhu Jun Shen; Salvatore Brugaletta; Hector M Garcia-Garcia; Jurgen Ligthart; Josep Gomez-Lara; Roberto Diletti; Giovanna Sarno; Karen Witberg; Patrick W Serruys
Journal:  Int J Cardiovasc Imaging       Date:  2010-12-21       Impact factor: 2.357

2.  New strategies in the treatment of coronary bifurcations.

Authors:  I Iakovou; N Foin; A Andreou; N Viceconte; C Di Mario
Journal:  Herz       Date:  2011-05       Impact factor: 1.443

3.  A comparison of tissue prolapse with optical coherence tomography and intravascular ultrasound after drug-eluting stent implantation.

Authors:  Jihyun Sohn; Seung-Ho Hur; In-Cheol Kim; Yun-Kyeong Cho; Hyung-Seob Park; Hyuck-Jun Yoon; Chang-Wook Nam; Yoon-Nyun Kim; Kwon-Bae Kim
Journal:  Int J Cardiovasc Imaging       Date:  2014-10-02       Impact factor: 2.357

Review 4.  Stenting of coronary bifurcation lesions: a literature and technical review.

Authors:  Joo Myung Lee; Kyung Woo Park; Bon-Kwon Koo; Hyo-Soo Kim
Journal:  Curr Cardiol Rep       Date:  2015-06       Impact factor: 2.931

5.  Expansion of the Multi-Link Frontier™ coronary bifurcation stent: micro-computed tomographic assessment in human autopsy and porcine heart samples.

Authors:  Stefan Kralev; Benjamin Haag; Jens Spannenberger; Siegfried Lang; Marc A Brockmann; Soenke Bartling; Alexander Marx; Karl-Konstantin Haase; Martin Borggrefe; Tim Süselbeck
Journal:  PLoS One       Date:  2011-07-21       Impact factor: 3.240

6.  Mechanical and physio-biological properties of peptide-coated stent for re-endothelialization.

Authors:  In-Ho Bae; Myung Ho Jeong; Dae Sung Park; Kyung Seob Lim; Jae Won Shim; Mun Ki Kim; Jun-Kyu Park
Journal:  Biomater Res       Date:  2020-01-23

Review 7.  Impact of coronary bifurcation angle on the pathogenesis of atherosclerosis and clinical outcome of coronary bifurcation intervention-A scoping review.

Authors:  Yoshinobu Murasato; Kyohei Meno; Takahiro Mori; Katsuhiko Tanenaka
Journal:  PLoS One       Date:  2022-08-17       Impact factor: 3.752

8.  Novel Polymer-Free Everolimus-Eluting Stent Fabricated using Femtosecond Laser Improves Re-endothelialization and Anti-inflammation.

Authors:  In-Ho Bae; Myung Ho Jeong; Kyung Seob Lim; Dae Sung Park; Jae Won Shim; Jun-Kyu Park; Kwang Hwan Oh; Mi Rim Jin; Doo Sun Sim
Journal:  Sci Rep       Date:  2018-05-09       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.