Rapid determination of serum testosterone by liquid chromatography-isotope dilution tandem mass spectrometry and a split sample comparison with three automated immunoassays.

Yu Chen1, Mehrdad Yazdanpanah, Barry R Hoffman, Eleftherios P Diamandis, Pui-Yuen Wong.   

Abstract

OBJECTIVES: To develop a rapid convenient-to-implement high performance liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) method for determination of serum testosterone concentration in routine clinical laboratories.
METHODS: Following extraction by organic solvents, an Agilent 1200 Series HPLC system coupled to an API 5000 mass spectrometer equipped with an atmospheric pressure chemical ionization ion source was used to separate, detect and quantify serum testosterone. Ion-transitions of m/z 289.2-->109.1 and 294.2-->113.2 were used to monitor testosterone and testosterone-2,2,4,6,6-d(5), respectively.
RESULTS: Functional sensitivity was 0.056 nmol/L (CV 20%). Within-run and total imprecision were 4.6% and 5.2% at 1.3 nmol/L, 2.4% and 4.3% at 11.0 nmol/L, and 1.9% and 1.9% at 23.4 nmol/L respectively. The LC-MS/MS method agreed closely with three automated immunoassays when the concentration of testosterone exceeded 3 nmol/L. However, the immunoassays showed a positive bias at concentrations below 3 nmol/L.
CONCLUSION: This method provides a rapid, simple, highly selective and sensitive procedure that can be easily used for determination of serum testosterone in routine clinical laboratories. It measures serum testosterone precisely and accurately at concentrations found in children and adults of both genders.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19109939     DOI: 10.1016/j.clinbiochem.2008.11.009

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  8 in total

1.  A multicenter evaluation of immunoassays for follicle-stimulating hormone, luteinizing hormone and testosterone: concordance, imprecision and reference values.

Authors:  A Radicioni; A Lenzi; M Spaziani; A Anzuini; G Ruga; G Papi; M Raimondo; C Foresta
Journal:  J Endocrinol Invest       Date:  2013-10       Impact factor: 4.256

2.  Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays.

Authors:  Yu Chen; Lois Kinney; Andrea Božović; Hilary Smith; Heather Tarr; Eleftherios P Diamandis; Adrien LeBlanc
Journal:  Clin Biochem       Date:  2012-06-13       Impact factor: 3.281

3.  Comparison of four clinically validated testosterone LC-MS/MS assays: Harmonization is an attainable goal.

Authors:  Deborah French; Julia Drees; Judith A Stone; Daniel T Holmes; J Grace van der Gugten
Journal:  Clin Mass Spectrom       Date:  2018-12-01

4.  Revisiting hyper- and hypo-androgenism by tandem mass spectrometry.

Authors:  Flaminia Fanelli; Alessandra Gambineri; Marco Mezzullo; Valentina Vicennati; Carla Pelusi; Renato Pasquali; Uberto Pagotto
Journal:  Rev Endocr Metab Disord       Date:  2013-06       Impact factor: 6.514

5.  Evaluation of new Beckman Coulter 25(OH) Vitamin D assay and potential improvement of clinical interpretation.

Authors:  Özlem Ç Madenci; Asuman Orçun; Zeynep Yildiz; Rana Sirmali; Nazan Tunçbilek; Nihal Yücel; Yusuf Temel; Aycan Bölük
Journal:  Biochem Med (Zagreb)       Date:  2017-06-15       Impact factor: 2.313

6.  Discordance between testosterone measurement methods in castrated prostate cancer patients.

Authors:  Mélanie Rouleau; Francis Lemire; Michel Déry; Benoît Thériault; Gabriel Dubois; Yves Fradet; Paul Toren; Chantal Guillemette; Louis Lacombe; Laurence Klotz; Fred Saad; Dominique Guérette; Frédéric Pouliot
Journal:  Endocr Connect       Date:  2019-02       Impact factor: 3.335

7.  New Tb3+-simvastatin optical biosensor for sensitive determination of folic acid, progesterone, testosterone and vitamin D3 in biological fluids.

Authors:  Mohamed S Attia; Amal M Ahmed; Tarek A Amin; Ahmed O Youssef; Mohammed A Amin; Ekram H Mohamed; Safwat A Mahmoud; Mona N Abou-Omar
Journal:  RSC Adv       Date:  2021-10-06       Impact factor: 4.036

8.  Analytical bias of automated immunoassays for six serum steroid hormones assessed by LC-MS/MS.

Authors:  Željko Debeljak; Ivana Marković; Jasna Pavela; Iva Lukić; Dario Mandić; Sanja Mandić; Vesna Horvat; Vatroslav Šerić
Journal:  Biochem Med (Zagreb)       Date:  2020-08-05       Impact factor: 2.313

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.