Abstract
Infrared spectroscopy is one of the most widely used techniques for measurement of conversion degree in dental composites. However, to obtain good quality spectra and quantitative analysis from spectral data, appropriate expertise and knowledge of the technique are mandatory. This paper presents important details to use infrared spectroscopy for determination of the conversion degree.Entities:
Mesh:
Substances:
Year: 2008 PMID: 19089207 PMCID: PMC4327635 DOI: 10.1590/s1678-77572008000200012
Source DB: PubMed Journal: J Appl Oral Sci ISSN: 1678-7757 Impact factor: 2.698
FIGURE 1Background spectrum in NIR (a) and MIR(b) spectral regions. The horizontal bars indicate where the absorption bands used to determine the conversion degree are localized (a.u.= arbitrary units)
FIGURE 2Deformed absorption spectrum of dental composite sample (bottom) and the background (bg) spectrum (top) showing the presence of water absorptions (a.u.= arbitrary units)
FIGURE 3TPH absorption bands in the MIR region before (a) and after (b) polymerization
FIGURE 4TPH absorption bands in the NIR region before (a) and after (b) polymerization
Information about the composites used in this study
| Material | Manufacturer | Monomers |
|---|---|---|
| TPH Spectrum | Dentsply Ind. Com. Ltda., Petrópolis, RJ, Brazil | Urethane modified Bis-GMA |
| AC Acron | GC Dental Ind. Corp., Tokyo, Japan | Methyl methacrylate (95%) Ethyleneglicol dimethacrylate (5%) |
| Classic | Clássico Artigos Odontológios Ltda, SP, Brazil | Methyl methacrylate Topanol and ethyleneglicol dimethacrylate |
FIGURE 5Spectra with corrected baseline of a diluted (with hexane) monomer of ACRON MC composite
FIGURE 6Standard or calibration curve obtained with a monomer of ACRON MC composite