Carlos Nelson Elias1, Flavio Queiroz Henriques.
Abstract
STATEMENT OF PROBLEM: In clinical practice, loss of adhesion between the silicone-based denture liner and the denture base resin is always an undesirable event that might cause loss of material softness, water sorption, bacterial colonization and functional failure of the prosthesis.Entities:
Year: 2007 PMID: 19089094 PMCID: PMC4327206 DOI: 10.1590/s1678-77572007000100005
Source DB: PubMed Journal: J Appl Oral Sci ISSN: 1678-7757 Impact factor: 2.698
Silicone-based resilient liners and denture base resin tested in this study
| Material | Composition and cure type | Batch No. | Manufacturer |
|---|---|---|---|
| Mucopren Soft | Polyvinylsiloxane | 000515 | Kettenbach GmbH & Co, |
| Mollosil Plus | Polydimethylsiloxane. | 001005 | Detax GmbH, Ettlingen, Germany |
| Dentusil | Polyvinylsiloxane | 0004-204 | Bosworth, Skokie, Il |
| QC-20 | Heat-polymerized acrylic resin | 65279 powder/63335 liquid | Dentsply Int Inc., York, Pa |
FIGURE 1Aluminum bars and spacer in the mold used for fabrication of tensile strength specimens
FIGURE 2Aluminum bars and spacer in the mold used for fabrication of shear strength specimens
FIGURE 3Tensile strength specimen in the universal testing machine
FIGURE 4Shear strength specimen in the universal testing machine
Tensile bond strength (MPa) means (±standard deviation) for the water-soaked and thermocycled groups (N=60)
| Water-soaking | Thermocycling | |||||
|---|---|---|---|---|---|---|
| Min | Max | Means (±SD) | Min | Max | Means (±SD) | |
| Mucopren Soft | 1.84 | 3.06 | 2.83 ± 0.48 | 1.03 | 2.05 | 1.63 ± 0.48 |
| Mollosil Plus | 0.78 | 1.46 | 1.04 ± 0.26 | 0.57 | 0.80 | 0.72 ± 0.10 |
| Dentusil | 0.49 | 1.90 | 1.14 ± 0.51 | 0.49 | 1.83 | 0.97 ± 0.51 |
n = 10 specimens per experimental condition.
Shear bond strength (MPa) means (±standard deviation) for the water-soaked and thermocycled groups (N=60)
| Water-soaking | Thermocycling | |||||
|---|---|---|---|---|---|---|
| Min | Max | Means (±SD) | Min | Max | Means (±SD) | |
| Mucopren Soft | 2.35 | 3.41 | 2.84 ± 0.53 | 1.39 | 2.64 | 2.13 ± 0.46 |
| Mollosil Plus | 0.85 | 1.27 | 1.04 ± 0.04 | 0.45 | 1.90 | 0.81 ± 0.42 |
| Dentusil | 0.59 | 0.96 | 0.81 ± 0.02 | 0.61 | 1.03 | 0.78 ± 0.04 |
n = 10 specimens per experimental condition.
Failure mode frequency (%) after debonding
| Water-soaking | Thermocycling | ||||||
|---|---|---|---|---|---|---|---|
| Bond strength test | Adhesive | Cohesive | Mixed | Adhesive | Cohesive | Mixed | |
| Mucopren Soft | Tensile | 20 | 60 | 20 | 40 | - | 60 |
| Shear | 80 | - | 20 | 40 | - | 60 | |
| Mollosil Plus | Tensile | 60 | 20 | 20 | 100 | - | - |
| Shear | 20 | 60 | 20 | 60 | - | 40 | |
| Dentusil | Tensile | 60 | - | 40 | 40 | - | 60 |
| Shear | 20 | 60 | 20 | 20 | 40 | 40 | |
FIGURE 5Specimen during tensile bond strength testing