Evaluating the accuracy of an aneroid sphygmomanometer in a clinical trial setting.

Yong Ma1, Marinella Temprosa, Sarah Fowler, Ronald J Prineas, Maria G Montez, Janet Brown-Friday, Mary L Carrion-Petersen, Tracy Whittington.   

Abstract

BACKGROUND: The mercury sphygmomanometer, the "gold" standard for blood pressure measurements, has been gradually phased out in many institutions because of environmental concerns. Our on-going clinical trial compared the use of mercury vs. aneroid sphygmomanometers, before implementing a study-wide transition to the aneroid sphygmomanometer.
METHODS: The Diabetes Prevention Program Outcomes Study (DPPOS) studied the accuracy of the Welch Allyn Tycos 767-Series Mobile aneroid sphygmomanometer from 20 March 2006 to 21 June 2006 at multiple clinic centers. We compared readings from 997 participants in 24 clinic centers using both mercury and aneroid sphygmomanometers.
RESULTS: The study found no statistically significant difference for systolic blood pressure (SBP) (P > 0.05) and a small but significantly (P < 0.0001) lower (0.8 mm Hg) reading for diastolic blood pressure (DBP) using the aneroid sphygmomanometer. Regression analysis of aneroid vs. mercury showed regression lines (Y = 4.8 + 0.96X for SBP, Y = 3.1 + 0.95X for DBP) slightly but statistically significantly different from the line of equality (P < 0.001). Participants' age, sex, race/ethnicity, body mass index, blood pressure, and clinical center together explain about 8-10% of the variation of the difference between readings from the two sphygmomanometers. Based on the above result, on 1 August 2006, DPPOS clinics began the conversion from mercury to aneroid sphygmomanometers.
CONCLUSIONS: The Welch Allyn Tycos 767-Series Mobile Aneroid model 7670-04 tested in this validation study can be used to replace mercury model in clinical trials.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19057514      PMCID: PMC2699679          DOI: 10.1038/ajh.2008.338

Source DB:  PubMed          Journal:  Am J Hypertens        ISSN: 0895-7061            Impact factor:   2.689


  8 in total

1.  Are aneroid sphygmomanometers accurate in hospital and clinic settings?

Authors:  V J Canzanello; P L Jensen; G L Schwartz
Journal:  Arch Intern Med       Date:  2001-03-12

2.  Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience.

Authors:  S A Yarows; K Qian
Journal:  Blood Press Monit       Date:  2001-04       Impact factor: 1.444

3.  Aneroid sphygmomanometers. An assessment of accuracy at a university hospital and clinics.

Authors:  R H Bailey; V L Knaus; J H Bauer
Journal:  Arch Intern Med       Date:  1991-07

Review 4.  Expressions of accuracy in the evaluation of biomedical instrumentation.

Authors:  N S Latman; R Lanier
Journal:  Biomed Instrum Technol       Date:  1998 May-Jun

5.  Hypertension, insulin, and proinsulin in participants with impaired glucose tolerance.

Authors: 
Journal:  Hypertension       Date:  2002-11       Impact factor: 10.190

6.  The Diabetes Prevention Program. Design and methods for a clinical trial in the prevention of type 2 diabetes.

Authors: 
Journal:  Diabetes Care       Date:  1999-04       Impact factor: 19.112

7.  Reduction in the incidence of type 2 diabetes with lifestyle intervention or metformin.

Authors:  William C Knowler; Elizabeth Barrett-Connor; Sarah E Fowler; Richard F Hamman; John M Lachin; Elizabeth A Walker; David M Nathan
Journal:  N Engl J Med       Date:  2002-02-07       Impact factor: 91.245

8.  Applying environmental product design to biomedical products research.

Authors:  J Messelbeck; L Sutherland
Journal:  Environ Health Perspect       Date:  2000-12       Impact factor: 9.031

  8 in total
  16 in total

1.  Blood pressure measurement guidelines for physical therapists.

Authors:  Ethel M Frese; Ann Fick; H Steven Sadowsky
Journal:  Cardiopulm Phys Ther J       Date:  2011-06

2.  Long-term effects of the Diabetes Prevention Program interventions on cardiovascular risk factors: a report from the DPP Outcomes Study.

Authors:  T J Orchard; M Temprosa; E Barrett-Connor; S E Fowler; R B Goldberg; K J Mather; S M Marcovina; M Montez; R E Ratner; C D Saudek; H Sherif; K E Watson
Journal:  Diabet Med       Date:  2013-01       Impact factor: 4.359

3.  Comparison of mercury and aneroid blood pressure measurements in youth.

Authors:  Amy S Shah; Lawrence M Dolan; Ralph B D'Agostino; Debra Standiford; Cralen Davis; Lisa Testaverde; Catherine Pihoker; Stephen R Daniels; Elaine M Urbina
Journal:  Pediatrics       Date:  2012-04-16       Impact factor: 7.124

4.  Baseline check of blood pressure readings of an automated device in severe pre-eclampsia and healthy normotensive pregnancy.

Authors:  Nnabuike C Ngene; Jagidesa Moodley
Journal:  Pregnancy Hypertens       Date:  2018-03-01       Impact factor: 2.899

5.  Chronic kidney disease and prevalent atrial fibrillation: the Chronic Renal Insufficiency Cohort (CRIC).

Authors:  Elsayed Z Soliman; Ronald J Prineas; Alan S Go; Dawei Xie; James P Lash; Mahboob Rahman; Akinlolu Ojo; Val L Teal; Nancy G Jensvold; Nancy L Robinson; Daniel L Dries; Lydia Bazzano; Emile R Mohler; Jackson T Wright; Harold I Feldman
Journal:  Am Heart J       Date:  2010-06       Impact factor: 4.749

Review 6.  Principles and techniques of blood pressure measurement.

Authors:  Gbenga Ogedegbe; Thomas Pickering
Journal:  Cardiol Clin       Date:  2010-11       Impact factor: 2.213

7.  Optimal Blood Pressure Thresholds for Minimal Coronary Artery Disease Risk in Type 1 Diabetes.

Authors:  Jingchuan Guo; Maria M Brooks; Matthew F Muldoon; Ashely I Naimi; Trevor J Orchard; Tina Costacou
Journal:  Diabetes Care       Date:  2019-06-27       Impact factor: 19.112

8.  Clinic, Home, and Kiosk Blood Pressure Measurements for Diagnosing Hypertension: a Randomized Diagnostic Study.

Authors:  Beverly B Green; Melissa L Anderson; Andrea J Cook; Kelly Ehrlich; Yoshio N Hall; Clarissa Hsu; Dwayne Joseph; Predrag Klasnja; Karen L Margolis; Jennifer B McClure; Sean A Munson; Mathew J Thompson
Journal:  J Gen Intern Med       Date:  2022-03-03       Impact factor: 6.473

9.  Aortic pulse pressure is associated with carotid IMT in chronic kidney disease: report from Chronic Renal Insufficiency Cohort.

Authors:  Stephanie S DeLoach; Lawrence J Appel; Jing Chen; Marshall M Joffe; Crystal A Gadegbeku; Emile R Mohler; Afshin Parsa; Kalyani Perumal; Mohammed A Rafey; Susan P Steigerwalt; Valerie Teal; Raymond R Townsend; Sylvia E Rosas
Journal:  Am J Hypertens       Date:  2009-09-24       Impact factor: 2.689

10.  Infant-feeding patterns and cardiovascular risk factors in young adulthood: data from five cohorts in low- and middle-income countries.

Authors:  Caroline Hd Fall; Judith B Borja; Clive Osmond; Linda Richter; Santosh K Bhargava; Reynaldo Martorell; Aryeh D Stein; Fernando C Barros; Cesar G Victora
Journal:  Int J Epidemiol       Date:  2010-09-17       Impact factor: 7.196

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.