Literature DB >> 19033084

In vivo kinematic analysis of a high-flexion, posterior-stabilized, mobile-bearing knee prosthesis in deep knee bending motion.

Masashi Tamaki1, Tetsuya Tomita, Tetsu Watanabe, Takaharu Yamazaki, Hideki Yoshikawa, Kazuomi Sugamoto.   

Abstract

The objective of this study was to evaluate the kinematics of a high-flexion, posterior-stabilized, mobile-bearing total knee arthroplasty (TKA) in weight-bearing, deep knee bending motion. Thirteen patients implanted with the Legacy Posterior Stabilized Flex (Zimmer, Warsaw, IN) mobile-bearing TKA were examined during a deep knee bending motion using fluoroscopy. Femorotibial motion was determined using a 2-dimensional to 3-dimensional registration technique, which used computer-assisted design models to reproduce the position of metallic implants from single-view fluoroscopic images. The average flexion range of motion between the metallic implants was 116 degrees . The average rotation of the femoral component was 9.3 degrees external rotation. The mean kinematic pathway was early rollback, lateral pivot with external rotation, and bicondylar rollback. We found that the kinematic pattern of the Legacy Posterior Stabilized Flex mobile-bearing TKA was different than normal knee kinematics.

Entities:  

Mesh:

Year:  2008        PMID: 19033084     DOI: 10.1016/j.arth.2008.06.012

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  15 in total

1.  Less femorotibial rotation and AP translation in deep-dished total knee arthroplasty. An intraoperative kinematic study using navigation.

Authors:  Philippe Massin; Patrick Boyer; Marc Sabourin
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-11-05       Impact factor: 4.342

2.  In vivo three-dimensional kinematics of total elbow arthroplasty using fluoroscopic imaging.

Authors:  Kazuma Futai; Tetsuya Tomita; Takaharu Yamazaki; Tsuyoshi Murase; Hideki Yoshikawa; Kazuomi Sugamoto
Journal:  Int Orthop       Date:  2010-02-23       Impact factor: 3.075

3.  In vivo determination of cam-post engagement in fixed and mobile-bearing TKA.

Authors:  Sumesh M Zingde; Filip Leszko; Adrija Sharma; Mohamed R Mahfouz; Richard D Komistek; Douglas A Dennis
Journal:  Clin Orthop Relat Res       Date:  2014-01       Impact factor: 4.176

4.  In vivo Kinematics of the Knee after a Posterior Cruciate-Substituting Total Knee Arthroplasty: A Comparison between Caucasian and South Korean Patients.

Authors:  Ji-Hoon Bae; Ali Hosseini; Kyung-Wook Nha; Sang-Eun Park; Tsung Yuan Tsai; Young-Min Kwon; Guoan Li
Journal:  Knee Surg Relat Res       Date:  2016-06-01

5.  Correlation between preoperative and postoperative knee kinematics in total knee arthroplasty using cruciate retaining designs.

Authors:  Jong Keun Seon; Ju Kwon Park; Mun Su Jeong; Woo Bin Jung; Kyung Soon Park; Taek Rim Yoon; Eun Kyoo Song
Journal:  Int Orthop       Date:  2010-04-30       Impact factor: 3.075

6.  Geometric variable designs of cam/post mechanisms influence the kinematics of knee implants.

Authors:  Ali Fallahiarezoodar; Mohammed Rafiq Abdul Kadir; Mina Alizadeh; Sangeetha Vasudevaraj Naveen; T Kamarul
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-08-23       Impact factor: 4.342

7.  Comparison of stability and kinematics of the natural knee versus a PS TKA with a 'third condyle'.

Authors:  Stefano Zaffagnini; Simone Bignozzi; Mo Saffarini; Francesca Colle; Bharat Sharma; Plamen Slavov Kinov; Maurilio Marcacci; David Dejour
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-05-06       Impact factor: 4.342

8.  In vivo kinematics comparison of fixed- and mobile-bearing total knee arthroplasty during deep knee bending motion.

Authors:  Xiaojun Shi; Bin Shen; Jing Yang; Pengde Kang; Zongke Zhou; Fuxing Pei
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-12-12       Impact factor: 4.342

9.  In vivo kinematics of the knee during weight bearing high flexion.

Authors:  Wei Qi; Ali Hosseini; Tsung-Yuan Tsai; Jing-Sheng Li; Harry E Rubash; Guoan Li
Journal:  J Biomech       Date:  2013-04-13       Impact factor: 2.712

10.  Can TKA design affect the clinical outcome? Comparison between two guided-motion systems.

Authors:  Raffaele Mugnai; Vitantonio Digennaro; Andrea Ensini; Alberto Leardini; Fabio Catani
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-04-30       Impact factor: 4.342

View more