Implanted cardiac defibrillator care in radiation oncology patient population.

Daphna Y Gelblum1, Howard Amols.   

Abstract

PURPOSE: To review the experience of a large cancer center with radiotherapy (RT) patients bearing implantable cardiac defibrillators (ICDs) to propose some preliminary care guidelines as we learn more about the devices and their interaction with the therapeutic radiation environment. METHODS AND MATERIALS: We collected data on patients with implanted ICDs treated with RT during a 2.5-year period at any of the five Memorial Sloan-Kettering clinical campuses. Information regarding the model, location, and dose detected from the device, as well as the treatment fields, fraction size, and treatment energy was collected. During this time, a new management policy for these patients had been implemented requiring treatment with low-energy beams (6 MV) and close surveillance of the patients in partnership with their electrophysiologist, as they received RT.
RESULTS: During the study period, 33 patients were treated with an ICD in place. One patient experienced a default of the device to its initial factory setting that was detected by the patient hearing an auditory signal from the device. This patient had initially been treated with a 15-MV beam. After this episode, his treatment was replanned to be completed with 6-MV photons, and he experienced no further events.
CONCLUSION: Patients with ICDs and other implanted computer-controlled devices will be encountered more frequently in the RT department, and proper management is important. We present a policy for the safe treatment of these patients in the radiation oncology environment.

Entities:  

Mesh:

Year:  2008        PMID: 18977096     DOI: 10.1016/j.ijrobp.2008.06.1903

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  16 in total

1.  Radiotherapy in patients with cardiac implantable electronic device either a permanent pacemaker or an implantable cardioverter defibrillator.

Authors:  Yasemin Benderli Cihan
Journal:  Radiol Med       Date:  2017-03-09       Impact factor: 3.469

2.  Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients.

Authors:  Vincenzo L Malavasi; Giuseppina De Marco; Jacopo F Imberti; Filippo Placentino; Marco Vitolo; Ercole Mazzeo; Gianfranco Cicoria; Edoardo Casali; Vincenzo Turco; Frank Lohr; Giuseppe Boriani
Journal:  Intern Emerg Med       Date:  2019-12-02       Impact factor: 3.397

Review 3.  [Radiation therapy in patients with cardiac pacemakers or implantable cardioverter defibrillators. Interdisciplinary safety recommendations].

Authors:  M Dorenkamp; C Stromberger; C von Heymann; W Haverkamp; P Wust; M Roser
Journal:  Strahlenther Onkol       Date:  2012-11-11       Impact factor: 3.621

4.  Effect of X-ray dose rates higher than 8 Gy/min on the functioning of cardiac implantable electronic devices.

Authors:  Kazuhiko Nakamura; Takahiro Aoyama; Naoki Kaneda; Masashi Otsuji; Yoshitaka Minami; Ami Sakuragi; Masaru Nakamura
Journal:  J Radiat Res       Date:  2020-05-22       Impact factor: 2.724

5.  A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices.

Authors:  Saman Nazarian; Rozann Hansford; Ariel Roguin; Dorith Goldsher; Menekhem M Zviman; Albert C Lardo; Brian S Caffo; Kevin D Frick; Michael A Kraut; Ihab R Kamel; Hugh Calkins; Ronald D Berger; David A Bluemke; Henry R Halperin
Journal:  Ann Intern Med       Date:  2011-10-04       Impact factor: 25.391

6.  DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.

Authors:  Benjamin Gauter-Fleckenstein; Carsten W Israel; Marc Dorenkamp; Jürgen Dunst; Mattias Roser; Rainer Schimpf; Volker Steil; Jörg Schäfer; Ulrike Höller; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2015-03-05       Impact factor: 3.621

7.  A clinical example of extreme dose exposure for an implanted cardioverter-defibrillator : Beyond the DEGRO guidelines.

Authors:  Yoana Hristova; Janett Köhn; Stefanie Preuß; Claus Rödel; Panagiotis Balermpas
Journal:  Strahlenther Onkol       Date:  2017-05-31       Impact factor: 3.621

8.  High dose radiotherapy to automated implantable cardioverter-defibrillator: a case report and review of the literature.

Authors:  Inaya Ahmed; Wei Zou; Salma K Jabbour
Journal:  Case Rep Oncol Med       Date:  2014-09-07

9.  A case of radiotherapy for an advanced bronchial carcinoma patient with implanted cardiac rhythm machines as well as heart assist device.

Authors:  Sergiu Scobioala; Iris Ernst; Christos Moustakis; Uwe Haverkamp; Sven Martens; Hans Theodor Eich
Journal:  Radiat Oncol       Date:  2015-04-07       Impact factor: 3.481

10.  Management of radiation oncology patients with a pacemaker or ICD: a new comprehensive practical guideline in The Netherlands. Dutch Society of Radiotherapy and Oncology (NVRO).

Authors:  Coen W Hurkmans; Joost L Knegjens; Bing S Oei; Ad J J Maas; G J Uiterwaal; Arnoud J van der Borden; Marleen M J Ploegmakers; Lieselot van Erven
Journal:  Radiat Oncol       Date:  2012-11-24       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.