Literature DB >> 18976176

Clinical accuracy of 3D fluoroscopy-assisted cervical pedicle screw insertion.

Yasuo Ito1, Yoshihisa Sugimoto, Masao Tomioka, Yasuhiro Hasegawa, Kie Nakago, Yukihisa Yagata.   

Abstract

OBJECT: Cervical pedicle screw (PS) misplacement leads to injury of the spinal cord, nerve root, and vertebral artery. Recently, several investigators reported on the usefulness of a spinal navigation system that improves the accuracy of PS insertion. In this study, the authors assessed the accuracy of cervical pedicle, lateral mass, and odontoid screw insertions placed using a 3D fluoroscopy navigation system, the Iso-C3D unit.
METHODS: In this prospective analysis of the authors' initial 50 cases of 3D fluoroscopy-assisted cervical screw insertion, the authors inserted 176 PSs, 58 lateral mass screws, and 5 odontoid screws into the C1-7 vertebrae. They placed screws using intraoperative acquisition of data by the isocentric C-arm fluoroscope and a computer navigation system. They obtained postoperative fine-cut CT scans in all patients and assessed the accuracy of screw insertion.
RESULTS: A PS (>or= 3.5 mm) could be inserted into 24 (63%) of 38 pedicles at the level of C-3, 18 (53%) of 34 pedicles at C-4, 30 (65%) of 46 at C-5, 33 (80%) of 41 at C-6, and 43 (100%) of 43 at C-7. Of 176 PSs inserted into vertebrae between C-2 and C-7, 171 screws (97.2%) were classified as Grade 1 (no pedicle perforation), and 5 screws (2.8%) were classified as Grade 2 (screw perforation of the cortex by up to 2 mm). Clinically significant screw deviation in the present study was considered Grade 3 (screw perforation of the cortex by > 2 mm), and this occurred in 0% of the placements.
CONCLUSIONS: In this study, the authors were able to correctly insert cervical PSs using the 3D fluoroscopy and navigation system.

Entities:  

Mesh:

Year:  2008        PMID: 18976176     DOI: 10.3171/SPI.2008.9.11.450

Source DB:  PubMed          Journal:  J Neurosurg Spine        ISSN: 1547-5646


  36 in total

1.  The stabilizing potential of anterior, posterior and combined techniques for the reconstruction of a 2-level cervical corpectomy model: biomechanical study and first results of ATPS prototyping.

Authors:  Heiko Koller; Rene Schmidt; Michael Mayer; Wolfgang Hitzl; Juliane Zenner; Stefan Midderhoff; Stefan Middendorf; Nicolaus Graf; Nicolaus Gräf; H Resch; Hans-Joachim Wilke; Hans-Joachim Willke
Journal:  Eur Spine J       Date:  2010-06-30       Impact factor: 3.134

2.  Cervical pedicle screw insertion using a gutter entry point at the transitional area between the lateral mass and lamina.

Authors:  Katsuhiro Tofuku; Hiroaki Koga; Setsuro Komiya
Journal:  Eur Spine J       Date:  2011-08-10       Impact factor: 3.134

3.  3D-based navigation in posterior stabilisations of the cervical and thoracic spine: problems and benefits. Results of 451 screws.

Authors:  J-S Jarvers; S Katscher; A Franck; S Glasmacher; C Schmidt; T Blattert; C Josten
Journal:  Eur J Trauma Emerg Surg       Date:  2011-04-01       Impact factor: 3.693

4.  Image-guided pedicle screw insertion accuracy: a meta-analysis.

Authors:  Nai-Feng Tian; Hua-Zi Xu
Journal:  Int Orthop       Date:  2009-05-08       Impact factor: 3.075

5.  Real-time imaging with the o-arm for skull base applications: a cadaveric feasibility study.

Authors:  Shaan M Raza; Alfred P See; Michael Lim
Journal:  J Neurol Surg B Skull Base       Date:  2012-10

6.  Cervical pedicle screw instrumentation is more reliable with O-arm-based 3D navigation: analysis of cervical pedicle screw placement accuracy with O-arm-based 3D navigation.

Authors:  Sourabh Chachan; Hamid Rahmatullah Bin Abd Razak; Wee Lim Loo; John Carson Allen; Dinesh Shree Kumar
Journal:  Eur Spine J       Date:  2018-04-12       Impact factor: 3.134

7.  Case report: osteoid osteoma of the C2 pedicle: surgical technique using a navigation system.

Authors:  Hideki Nagashima; Takako Nishi; Koji Yamane; Atsushi Tanida
Journal:  Clin Orthop Relat Res       Date:  2009-06-30       Impact factor: 4.176

Review 8.  An endoscopic surgical technique for treating radiculopathy secondary to S1 nerve compression from a pedicle screw: technical note.

Authors:  Ralf Wagner; Albert E Telfeian
Journal:  J Spine Surg       Date:  2018-12

9.  Cervical pedicle screw insertion using the technique with direct exposure of the pedicle by laminoforaminotomy.

Authors:  Dae-Jean Jo; Eun-Min Seo; Ki-Tack Kim; Sung-Min Kim; Sang-Hun Lee
Journal:  J Korean Neurosurg Soc       Date:  2012-11-30

10.  Percutaneous, Navigated Minimally Invasive Posterior Cervical Pedicle Screw Fixation.

Authors:  Domagoj Coric; Vincent J Rossi; John Peloza; Paul K Kim; Tim E Adamson
Journal:  Int J Spine Surg       Date:  2020-10-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.