Patient dose evaluation by means of DICOM images for a direct radiography system.

O Rampado1, E Garelli, R Zatteri, U Escoffier, R De Lucchi, R Ropolo.   

Abstract

PURPOSE: The purpose of this work was to evaluate the statistical distribution of patient dose for different examinations by using the data stored in a DICOM image archive of a direct digital radiography system.
MATERIALS AND METHODS: An automatic procedure to extract dose data and exposure parameters from the image archives was implemented. During a 4.5-month period, 8,292 images were collected. Exposure parameters and the air kerma area product (P (KA)) stored in the image DICOM header were examined for each image. The entrance surface air kerma (K (a,e)), a quantity comparable to the current diagnostic reference levels (DRL), was estimated considering the P (KA) and the geometric parameters of each examination.
RESULTS: P (KA) and K (a,e ) distributions showed highly variable values. The obtained K (a,e ) values were substantially lower than the DRL. DRL were more than six times the mean value of K (a,e ) distribution for the abdomen anteroposterior (AP) and lumbar spine lateral (LAT) projections, whereas the ratio was equal to 2.7 for posteroanterior (PA) chest examinations.
CONCLUSIONS: The method adopted proved to be effective for characterising the dose of each examination by means of the statistical analysis of the dose quantities over extensive samples. The dose values obtained and the comparison with the DRL showed that this radiographic device allows substantial dose savings compared with estimations made for nondigital or for phosphor-based systems.

Entities:  

Mesh:

Year:  2008        PMID: 18956145     DOI: 10.1007/s11547-008-0339-5

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  10 in total

1.  [Check of compliance with quality criteria for diagnostic radiographic images in quality control].

Authors:  R Ropolo; P Isoardi; P Catuzzo; S D Bianchi; G B Bradac; T Cammarota; O Davini; R De Lucchi; G Gandini
Journal:  Radiol Med       Date:  2001 Jan-Feb       Impact factor: 3.469

2.  Chest radiography: optimization of X-ray spectrum for cesium iodide-amorphous silicon flat-panel detector.

Authors:  James T Dobbins; Ehsan Samei; Harrell G Chotas; Richard J Warp; Alan H Baydush; Carey E Floyd; Carl E Ravin
Journal:  Radiology       Date:  2003-01       Impact factor: 11.105

Review 3.  Dose reduction in skeletal and chest radiography using a large-area flat-panel detector based on amorphous silicon and thallium-doped cesium iodide: technical background, basic image quality parameters, and review of the literature.

Authors:  Markus Völk; Okka W Hamer; Stefan Feuerbach; Michael Strotzer
Journal:  Eur Radiol       Date:  2004-02-17       Impact factor: 5.315

4.  Retrospective patient dose analysis of a digital radiography system in routine clinical use.

Authors:  Astrid Schuncke; Ulrich Neitzel
Journal:  Radiat Prot Dosimetry       Date:  2005       Impact factor: 0.972

5.  Routine chest radiography using a flat-panel detector: image quality at standard detector dose and 33% dose reduction.

Authors:  Michael Strotzer; Markus Völk; Rüdiger Fründ; Okka Hamer; Niels Zorger; Stefan Feuerbach
Journal:  AJR Am J Roentgenol       Date:  2002-01       Impact factor: 3.959

6.  Computing patient doses of X-ray examinations using a patient size- and sex-adjustable phantom.

Authors:  S Rannikko; I Ermakov; J S Lampinen; M Toivonen; K T Karila; A Chervjakov
Journal:  Br J Radiol       Date:  1997-07       Impact factor: 3.039

7.  Comparison of eight different digital chest radiography systems: variation in detection of simulated chest disease.

Authors:  Lucia J M Kroft; Wouter J H Veldkamp; Bart J A Mertens; Mireille V Boot; Jacob Geleijns
Journal:  AJR Am J Roentgenol       Date:  2005-08       Impact factor: 3.959

8.  Radiation dose in digital chest radiography: comparison among three technologies.

Authors:  Federica Saiani; Cristina Ghirardi; Carlo Alberto Rodella; Piero Feroldi; Antonio Chiesa
Journal:  Radiol Med       Date:  2004-04       Impact factor: 3.469

9.  Optimization of dose and image quality for computed radiography and digital radiography.

Authors:  John E Aldrich; Emerenciana Duran; Pat Dunlop; John R Mayo
Journal:  J Digit Imaging       Date:  2006-06       Impact factor: 4.056

10.  Does digital imaging decrease patient dose? A pilot study and review of the literature.

Authors:  V Neofotistou; V Tsapaki; S Kottou; A Schreiner-Karoussou; E Vano
Journal:  Radiat Prot Dosimetry       Date:  2006-02-07       Impact factor: 0.972

  10 in total
  1 in total

1.  A method for calculating the dose length product from CT DICOM images.

Authors:  I A Tsalafoutas; S I Metallidis
Journal:  Br J Radiol       Date:  2010-11-16       Impact factor: 3.039

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.