The azygos defibrillator lead for elevated defibrillation thresholds: implant technique, lead stability, and patient series.

Jonas A Cooper1, Matthew P Latacha, Gabriel E Soto, Richard G Garmany, Marye J Gleva, Jane Chen, Mitchell N Faddis, Timothy W Smith.   

Abstract

BACKGROUND: Conventional insertion of implantable cardioverter-defibrillator (ICD) includes an evaluation of the defibrillation threshold (DFT). Implanting an ancillary defibrillation lead in the azygos vein has been introduced as a therapeutic option in patients with "high" DFT. This study reports the efficacy and stability of azygos defibrillation coils implanted for elevated DFTs.
METHODS: This is a retrospective review of seven consecutive patients with right and left pectoral, single- and dual-chamber, and biventricular ICDs and elevated DFTs, in whom an azygos defibrillation coil was introduced.
RESULTS: Addition of an azygos defibrillator lead achieved a satisfactory safety margin during single energy defibrillation efficacy testing in four out of seven patients, with success at maximum device output in two patients. No satisfactory safety margin was achieved in the remaining patient, despite the further addition of a subcutaneous defibrillation coil. No change in lead position was observed over a mean radiographic follow-up of 8 months. No complications were noted during a mean follow-up of 14 months, including no deaths, and no ICD shocks.
CONCLUSION: Implanting a defibrillation coil into the azygos vein is feasible and safe. In a majority of patients with failed defibrillation efficacy testing, adding an azygos coil achieves success on repeat testing. Therefore, this technique is one option for lowering the defibrillation threshold in patients who fail DFT testing of their ICD.

Entities:  

Mesh:

Year:  2008        PMID: 18950297     DOI: 10.1111/j.1540-8159.2008.01203.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  12 in total

1.  Axillary sub-pectoral pulse generator pocket for lowering defibrillation threshold.

Authors:  Ernest W Lau
Journal:  J Interv Card Electrophysiol       Date:  2012-02-23       Impact factor: 1.900

2.  Case report of an anomalous single azygos venous coil insertion to reduce the defibrillation threshold in a patient with a right-sided deltopectoral ICD implant.

Authors:  Darragh Patrick Moran; Usman Bhutta; Ibrahim Yearoo; Edward Keelan; James O'Neill; Joseph Galvin
Journal:  Heart Asia       Date:  2013-03-05

3.  A novel approach to eliminate intraventricular lead placement in patients with congenital heart disease.

Authors:  Nikhil Gupta; Jeremy P Moore; Kevin Shannon
Journal:  J Interv Card Electrophysiol       Date:  2012-05-04       Impact factor: 1.900

4.  Additional coronary sinus defibrillation lead with a right pectoral ICD and high DFT : a case report.

Authors:  M Strauss; T Kleemann; U Weisse; F-U Sack; R Zahn
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2011-04-10

5.  Azygos vein lead implantation for high defibrillation thresholds in implantable cardioverter defibrillator placement.

Authors:  Naga Va Kommuri; Sri Lakshmi S Kollepara; E Saulitis; Ma Siddiqui
Journal:  Indian Pacing Electrophysiol J       Date:  2010-01-07

6.  Orthostatic increase in defibrillation threshold leading to defibrillation failure and prolonged cardiac arrest in a sitting position: Lessons from a patient's near-fatal experience.

Authors:  Xiaoke Liu; Samuel Asirvatham; G Glenn Kabell
Journal:  HeartRhythm Case Rep       Date:  2016-03-28

7.  Transvenous lead implantation via the accessory hemiazygos vein in superior vena cava occlusion.

Authors:  Peter H Waddingham; Adam S C Dennis; Anu Balan; Anthony W C Chow
Journal:  Heart Rhythm O2       Date:  2021-12-17

8.  Cases of Azygous Coil Extraction.

Authors:  James R Sampognaro; Robert K Lewis; Eric Black-Maier; Sean D Pokorney; Donald D Hegland; Jonathan P Piccini
Journal:  Heart Rhythm O2       Date:  2021-12-28

9.  Electroanatomical mapping- and CT scan image integration-guided pacing lead implantation: A case series and review of the recent literature.

Authors:  Jens Kristensen; Mads Brix Kronborg; Christian Gerdes; Jens Cosedis Nielsen
Journal:  Heart Rhythm O2       Date:  2020-10-28

10.  Coronary vein defibrillator coil placement in patients with high defibrillation thresholds.

Authors:  Moisés Rodríguez-Mañero; Bahij Kreidieh; Sergio H Ibarra-Cortez; Paulino Álvarez; Paul Schurmann; Amish S Dave; Miguel Valderrábano
Journal:  J Arrhythm       Date:  2018-12-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.