James R Sampognaro1, Robert K Lewis1, Eric Black-Maier1, Sean D Pokorney1, Donald D Hegland1, Jonathan P Piccini1.
Abstract
BACKGROUND: Procedural and clinical outcomes of patients undergoing extraction or removal of azygous coils are not well characterized.Entities:
Keywords: Azygous coil; Cardiac implanted electronic devices; Implantable cardioverter-defibrillator; Lead extraction; Lead management
Year: 2021 PMID: 35243437 PMCID: PMC8859796 DOI: 10.1016/j.hroo.2021.12.004
Source DB: PubMed Journal: Heart Rhythm O2 ISSN: 2666-5018
Patient characteristics
| Patient | 1 | 2 |
|---|---|---|
| Age at extraction (years) | 73 | 83 |
| Sex | M | M |
| BMI (kg/m2) | 26.72 | 26.31 |
| Diabetes mellitus | Yes | Yes |
| eGFR (mL/min/1.73 m2) | 52 | 39 |
| HTN | Yes | Yes |
| CAD | Yes | Yes |
| PAD | No | No |
| Atrial fibrillation | Yes | Yes |
| Prior sternotomy | No | Yes |
| HF | Yes | Yes |
| EF (%) | <15 | 20 |
| Etiology of CM | Nonischemic | Ischemic |
| Type of CIED | Single-chamber ICD | CRT-D |
| Device indication | 1° prevention | 2° prevention |
| Implant side | Left | Left |
| Pacemaker dependent | No | No |
| Extraction indication | MSSA sepsis | Localized pocket infection |
1° = primary; 2° = secondary; BMI = body mass index; CAD = coronary artery disease; CIED = cardiac implantable electronic device; CM = cardiomyopathy; CRT-D = cardiac resynchronization therapy defibrillator; EF = ejection fraction; eGFR = estimated glomerular filtration rate; HF = heart failure; HTN = hypertension; ICD = implantable cardioverter-defibrillator; MSSA = methicillin-susceptible Staphylococcus aureus; PAD = peripheral arterial disease.
Procedural details
| Patient | Time, generator implant to removal (years) | Lead location | Lead model | Time from lead implant to removal (years) | Removal technique by lead | Extraction outcome | Procedural complications |
|---|---|---|---|---|---|---|---|
| 1 | 1.3 | RV | 6947 | 6.58 | 14F laser (16 s) | Complete procedural success | None |
| Azygos | 6937A | 1.33 | 14F laser (6 s) | ||||
| 2 | 6 | Azygos | 6937A | 6.01 | 14F laser (5 s) | Complete procedural success | Patient death 26 days after procedure |
| RV | 6935 | 6.01 | 14F laser (8 s) | ||||
| CS | 4196 | 6.01 | 14F laser (6 s) | ||||
| RV (aband.) | 0062 | 12.63 | 16F laser (1 s), 13F TightRail (Philips Healthcare, Andover, MA) (5 clicks) |
Aband. = abandoned; CIED = cardiac implantable electronic device; CS = coronary sinus; RA = right atrium; RV = right ventricle.
Figure 1Azygous coils on radiography (x-ray) and computed tomography (CT) images. Pt = patient.
Figure 2Construction of the 6937A defibrillator coil. The lead has a single conductor with inner silicone and outer polyurethane insulation. The outer diameter is 2.3 mm or 6.9 French.