Image distortion and spatial resolution of a commercially available cone-beam computed tomography machine.

John W Ballrick1, J Martin Palomo, Edward Ruch, B Douglas Amberman, Mark G Hans.   

Abstract

INTRODUCTION: Our objective was to evaluate images produced by a commercially available cone-beam computed tomography (CBCT) machine (i-CAT model 9140-0035-000C, Imaging Sciences International, Hatfield, Pa) for measurement and spatial resolution (ie, the ability to separate 2 objects in close proximity in the image) for all settings and in all dimensions.
METHODS: A custom phantom containing 0.3 mm diameter chromium metal markers approximately 5 mm apart in 3 planes of space was developed for analyzing distortion and measurement accuracy. This phantom was scanned in the CBCT machine by using all 12 commercially available settings. The distance between the markers was measured 3 times on the 3-dimensional images by using a Digital Imaging and Communications in Medicine (DICOM) viewer and was also measured 3 times directly on the phantom with a fine-tipped digital caliper. A line-pair phantom was used to evaluate spatial resolution. Thirty evaluators analyzed images and assigned a resolution from 0.2 to 1.6 mm according to the separation of the line pairs.
RESULTS: There were no statistically significant differences among the 3-dimensional images for any setting, in any dimension, or in images divided by thirds in terms of measurement accuracy. Comparison of the CBCT measurements to the direct digital caliper measurements showed a statistically significant difference (P <0.01). However, the absolute difference was <0.1 mm and is probably not clinically significant for most applications. The worst spatial resolution found was 0.86 mm. Spatial resolution was lower at faster scan times and larger voxel sizes.
CONCLUSIONS: This CBCT machine has clinically accurate measurements and acceptable resolution.

Entities:  

Mesh:

Year:  2008        PMID: 18929276     DOI: 10.1016/j.ajodo.2007.11.025

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  38 in total

1.  The use of cone beam computed tomography to predetermine root canal lengths in molar teeth: a comparison between two-dimensional and three-dimensional measurements.

Authors:  J P Tchorz; P J Poxleitner; S Stampf; S B M Patzelt; D Rottke; E Hellwig; M J Altenburger
Journal:  Clin Oral Investig       Date:  2013-07-30       Impact factor: 3.573

2.  Accuracy assessment of the axial images obtained from cone beam computed tomography.

Authors:  Francine Kühl Panzarella; J L C Junqueira; L B Oliveira; N S de Araújo; C Costa
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

3.  Comparison of the influence of FOV sizes and different voxel resolutions for the assessment of periodontal defects.

Authors:  M E Kolsuz; N Bagis; K Orhan; H Avsever; K Ö Demiralp
Journal:  Dentomaxillofac Radiol       Date:  2015-04-22       Impact factor: 2.419

4.  Analysis of linear measurements on 3D surface models using CBCT data segmentation obtained by automatic standard pre-set thresholds in two segmentation software programs: an in vitro study.

Authors:  Marcelo Lupion Poleti; Thais Maria Freire Fernandes; Otávio Pagin; Marcela Rodrigues Moretti; Izabel Regina Fischer Rubira-Bullen
Journal:  Clin Oral Investig       Date:  2015-05-13       Impact factor: 3.573

5.  Effect of voxel size on detection of fenestration, dehiscence and furcation defects using cone-beam computed tomography.

Authors:  Masoumeh Eftekhar; Hanieh Kaviani; Nina Rouzmeh; Aitin Torabinia; Alireza Akbarzadeh Baghban
Journal:  Oral Radiol       Date:  2021-01-11       Impact factor: 1.852

Review 6.  Quality assurance phantoms for cone beam computed tomography: a systematic literature review.

Authors:  Marcus V L de Oliveira; Ann Wenzel; Paulo S F Campos; Rubens Spin-Neto
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

Review 7.  Impact of voxel size variation on CBCT-based diagnostic outcome in dentistry: a systematic review.

Authors:  Rubens Spin-Neto; Erik Gotfredsen; Ann Wenzel
Journal:  J Digit Imaging       Date:  2013-08       Impact factor: 4.056

8.  Performance of a photostimulable storage phosphor digital system with or without the sharpen filter and cone beam CT for detecting approximal enamel subsurface demineralization.

Authors:  M D F Belém; C P M Tabchoury; R I Ferreira-Santos; F C Groppo; F Haiter-Neto
Journal:  Dentomaxillofac Radiol       Date:  2013-02-14       Impact factor: 2.419

9.  Analysis of bite marks in foodstuffs by computer tomography (cone beam CT)--3D reconstruction.

Authors:  Jeidson Marques; Jamilly Musse; Catarina Caetano; Francisco Corte-Real; Ana Teresa Corte-Real
Journal:  J Forensic Odontostomatol       Date:  2013-12-01

10.  Buccal cortical bone thickness at miniscrew placement sites in patients with different vertical skeletal patterns.

Authors:  Ilknur Veli; Tancan Uysal; Asli Baysal; Irfan Karadede
Journal:  J Orofac Orthop       Date:  2014-10-26       Impact factor: 1.938

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.