A clinical evaluation of rapid automatic noninvasive blood pressure determination with the Ohmeda 2120 "return-to-flow" method.

F E Block1, M V Fletcher, T J Morris, R Dzwonczyk.   

Abstract

Most commercial noninvasive blood pressure devices use the oscillometric method for determination of blood pressure. The Ohmeda 2120 noninvasive blood pressure monitor uses the oscillometric technique, but it also includes a "Sys Stat" mode for rapid determination of systolic blood pressure up to ten times per minute. This Sys Stat mode uses the "return-to-flow" method, in which the pressure on an upper arm cuff decreases in small steps until a finger sensor detects the first pulse of blood. Data from 16 patients monitored with the Ohmeda 2120 unit and with an arterial cannula were analyzed. The line of regression was Sys Stat = 0.85 (arterial) + 9.49. The Ohmeda 2120 monitor tended to underestimate the arterial pressure, particularly at higher pressures. Several factors could have contributed to this underestimation. Even so, the accuracy is sufficient for clinical use under most circumstances. More importantly, in an urgent situation, the Sys Stat mode can provide a reading of systolic blood pressure within seconds.

Entities:  

Mesh:

Year:  1991        PMID: 1890447     DOI: 10.1007/bf01619267

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  5 in total

1.  Comparison of intra-arterial and automated oscillometric blood pressure measurement methods in postoperative hypertensive patients.

Authors:  P G Loubser
Journal:  Med Instrum       Date:  1986 Sep-Oct

2.  Indirect measurement of mean blood pressure in the anesthetized patient.

Authors:  M Yelderman; A K Ream
Journal:  Anesthesiology       Date:  1979-03       Impact factor: 7.892

3.  Statistical evaluation of agreement between two methods for measuring a quantitative variable.

Authors:  J Lee; D Koh; C N Ong
Journal:  Comput Biol Med       Date:  1989       Impact factor: 4.589

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Noninvasive automatic determination of mean arterial pressure.

Authors:  M Ramsey
Journal:  Med Biol Eng Comput       Date:  1979-01       Impact factor: 2.602

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.