Rates and severity of perforation from implantable cardioverter-defibrillator leads: a 4-year study.

Mintu Turakhia1, Megha Prasad, Jeffrey Olgin, Nitish Badhwar, Zian H Tseng, Randall Lee, Gregory M Marcus, Byron K Lee.   

Abstract

INTRODUCTION: Although recent case reports and one small single-year observational study report a substantially increased rate of perforation with the St. Jude Riata series defibrillator lead, these results have not been externally validated. METHODS AND
RESULTS: From 2004 to 2007, 593 implantable cardioverter-defibrillator (ICD) implants were performed by six faculty and 13 fellows at four University of California, San Francisco, CA, USA and affiliate hospitals. An electronic medical records system was systematically searched to identify clinically significant cases of ICD lead dislodgment or perforation. Of 307 (56%) St. Jude leads (all Riata series 6.3- and 7.3-French leads), 188 (29%) Medtronic leads (including 99 Sprint Fidelis 6.6 French leads), and 98 (15%) Guidant/Boston Scientific leads, there were three perforations in 593 cases (0.51%). One perforation occurred with a Medtronic Sprint Fidelis 6949 lead (0.53%), and two with a St. Jude Riata 1581 lead (0.65%). There were no statistically significant differences in perforation or dislodgement rates between manufacturers or lead models (p = NS for all). In both cases of perforation with the St. Jude Riata leads, the lead tip perforated through the pericardium into the pleural space.
CONCLUSIONS: In our 4-year series of ICD implants, perforation and dislodgement rates were low, similar across all lead makes and models, and well below published and accepted complication rates. Our findings contradict previously-reported higher rates of perforation with the Riata lead. Registry and product performance reports should also classify complications by severity and outcome to provide a more complete assessment of product safety.

Entities:  

Mesh:

Year:  2008        PMID: 18855128     DOI: 10.1007/s10840-008-9308-0

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  12 in total

1.  Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator.

Authors:  Stephan B Danik; Moussa Mansour; Jagmeet Singh; Vivek Y Reddy; Patrick T Ellinor; David Milan; E Kevin Heist; Andre d'Avila; Jeremy N Ruskin; Theofanie Mela
Journal:  Heart Rhythm       Date:  2007-01-07       Impact factor: 6.343

2.  Lead perforation with a small body diameter implantable defibrillator lead.

Authors:  Ranjit Suri; Seth Keller
Journal:  Heart Rhythm       Date:  2007-07-14       Impact factor: 6.343

3.  Increased incidence of subacute perforation noted with one manufacturer of an implantable cardioverter-defibrillator lead.

Authors:  Harry G Mond
Journal:  Heart Rhythm       Date:  2007-06-21       Impact factor: 6.343

4.  Concerns about the Riata ST (St. Jude Medical) ICD lead.

Authors:  Stephen C Vlay
Journal:  Pacing Clin Electrophysiol       Date:  2008-01       Impact factor: 1.976

5.  Delayed defibrillator lead perforation: an increasing phenomenon.

Authors:  Ruby Satpathy; Tom Hee; Dennis Esterbrooks; Syed Mohiuddin
Journal:  Pacing Clin Electrophysiol       Date:  2008-01       Impact factor: 1.976

6.  Lead perforation: incidence in registries.

Authors:  Mark D Carlson; Roger A Freedman; Paul A Levine
Journal:  Pacing Clin Electrophysiol       Date:  2008-01       Impact factor: 1.976

7.  Perforation of the right ventricle by transvenous defibrillator leads: prevention and treatment.

Authors:  J E Molina
Journal:  Pacing Clin Electrophysiol       Date:  1996-03       Impact factor: 1.976

8.  Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years.

Authors:  Thomas Kleemann; Torsten Becker; Klaus Doenges; Margit Vater; Jochen Senges; Steffen Schneider; Werner Saggau; Udo Weisse; Karlheinz Seidl
Journal:  Circulation       Date:  2007-04-30       Impact factor: 29.690

9.  Right ventricular perforation with an ICD defibrillation lead managed by surgical revision and epicardial leads--case reports.

Authors:  Lubomír Krivan; Milan Kozák; Jitka Vlasínová; Milan Sepsi
Journal:  Pacing Clin Electrophysiol       Date:  2008-01       Impact factor: 1.976

10.  Asymptomatic anterior perforation of an ICD lead into subcutaneous tissues.

Authors:  John D Fisher; Marshal Fox; Soo G Kim; Daniel Goldstein; Linda B Haramati
Journal:  Pacing Clin Electrophysiol       Date:  2008-01       Impact factor: 1.976

View more
  4 in total

1.  Late perforation of Durata ICD lead--a case report.

Authors:  Bharati Das; Sachin Yalagudri; S Maiya; Sunita Abhraham; J Kannan; Ravi Kishore
Journal:  Indian Heart J       Date:  2014 Jul-Aug

2.  Delayed perforation of the right ventricular wall by a single standard-caliber implantable cardioverter-defibrillator lead detected by multidetector computed tomography.

Authors:  An Yoshimori; Atsushi Kobori; Nasu Michihiro; Yutaka Furukawa
Journal:  Korean Circ J       Date:  2011-11-29       Impact factor: 3.243

3.  Missed diagnosis: Dual right ventricular lead perforation resulting in chest pain.

Authors:  Daniel W Nelson; Caroline Vloka; Jule Wetherbee
Journal:  HeartRhythm Case Rep       Date:  2016-10-03

4.  Incidence of lead dislodgement, malfunction and perforation during the first year following device implantation.

Authors:  A Ghani; P P H M Delnoy; A R Ramdat Misier; J J J Smit; A Adiyaman; J P Ottervanger; A Elvan
Journal:  Neth Heart J       Date:  2014-06       Impact factor: 2.380

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.