Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator.

Stephan B Danik1, Moussa Mansour, Jagmeet Singh, Vivek Y Reddy, Patrick T Ellinor, David Milan, E Kevin Heist, Andre d'Avila, Jeremy N Ruskin, Theofanie Mela.   

Abstract

BACKGROUND: The rapid evolution of implantable cardioverter-defibrillator (ICD) leads has resulted in thinner active fixation leads. While these advances have made the leads more versatile, new configurations may be associated with unforeseen complications.
OBJECTIVE: The purpose of this study was to determine the incidence of perforation and dislodgement of defibrillator leads in a single center in the year 2005.
METHODS: All patients who underwent percutaneous ICD implantation at the Massachusetts General Hospital using an endocardial right ventricular lead were included in this study. The specific leads analyzed were the Riata (1580/1581 and 1590/1591, St. Jude Medical, St Paul, Minnesota, USA;) and Sprint Fidelis (6949-65, Medtronic, Minneapolis, Minnesota, USA.). Information was collected retrospectively.
RESULTS: A total of 130 Riata leads and 111 Sprint Fidelis leads were implanted at the Massachusetts General Hospital during this time period. A total of five lead perforations occurred in patients implanted with the Riata lead as compared with none with the Sprint Fidelis lead (3.8% vs. 0%, respectively; P <.05). Two of the five patients with perforation required pericardiocentesis for tamponade. Clinical symptoms of perforation developed 1-10 days after implant. Moreover, there were five additional lead revisions in the Riata group, which were likely due to dislodgement and/or microperforation, as compared with none in the Sprint Fidelis group (7.7% vs. 0%, respectively; P <.005).
CONCLUSIONS: In 2005, at one institution, there were significantly more cardiac perforations and lead revisions with the Riata lead as compared with the Sprint Fidelis right ventricular defibrillator lead. Further data are required to determine whether certain lead characteristics are responsible for this observation.

Entities:  

Mesh:

Year:  2007        PMID: 17399630     DOI: 10.1016/j.hrthm.2006.12.044

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  15 in total

Review 1.  Strategic choices to reduce implantable cardioverter-defibrillator-related morbidity.

Authors:  Oussama Wazni; Bruce L Wilkoff
Journal:  Nat Rev Cardiol       Date:  2010-04-20       Impact factor: 32.419

2.  Rates and severity of perforation from implantable cardioverter-defibrillator leads: a 4-year study.

Authors:  Mintu Turakhia; Megha Prasad; Jeffrey Olgin; Nitish Badhwar; Zian H Tseng; Randall Lee; Gregory M Marcus; Byron K Lee
Journal:  J Interv Card Electrophysiol       Date:  2008-10-15       Impact factor: 1.900

3.  Riata lead up to date.

Authors:  Nobuhiro Nishii
Journal:  J Cardiol Cases       Date:  2017-02-01

Review 4.  Subcutaneous chronic implantable defibrillation systems in humans.

Authors:  Riccardo Cappato; Warren M Smith; Margaret A Hood; Ian G Crozier; Luc Jordaens; Stefan G Spitzer; Andrey V Ardashev; Lucas Boersma; Pierpaolo Lupo; Andrew A Grace; Gust H Bardy
Journal:  J Interv Card Electrophysiol       Date:  2012-03-17       Impact factor: 1.900

5.  Late perforation of Durata ICD lead--a case report.

Authors:  Bharati Das; Sachin Yalagudri; S Maiya; Sunita Abhraham; J Kannan; Ravi Kishore
Journal:  Indian Heart J       Date:  2014 Jul-Aug

6.  Simulation of mechanical environment in active lead fixation: effect of fixation helix size.

Authors:  Xuefeng Zhao; Jonathan F Wenk; Mike Burger; Yi Liu; Mithilesh K Das; William Combs; Liang Ge; Julius M Guccione; Ghassan S Kassab
Journal:  J Biomech Eng       Date:  2011-06       Impact factor: 2.097

7.  Delayed perforation of the right ventricular wall by a single standard-caliber implantable cardioverter-defibrillator lead detected by multidetector computed tomography.

Authors:  An Yoshimori; Atsushi Kobori; Nasu Michihiro; Yutaka Furukawa
Journal:  Korean Circ J       Date:  2011-11-29       Impact factor: 3.243

8.  Life-threatening perforation of a defibrillation lead.

Authors:  R Hermanides; J Ottervanger; A Elvan; A Ramdat Misier
Journal:  Neth Heart J       Date:  2009-03       Impact factor: 2.380

9.  Subacute perforation of the St. Jude Riata implantable cardioverter-defibrillator lead: a report of two pediatric cases.

Authors:  T Ben Morrison; Michael J Ackerman; Robert F Rea
Journal:  Pediatr Cardiol       Date:  2009-04-14       Impact factor: 1.655

10.  Introduction of new industry standards for cardiac implantable electronic devices: balancing benefits and unexpected risks.

Authors:  Christian Sticherling; Haran Burri
Journal:  Europace       Date:  2012-03-01       Impact factor: 5.214