Limits of preservation of samples for urine strip tests and particle counting.

Timo Kouri1, Outi Malminiemi, Joris Penders, Virpi Pelkonen, Lotta Vuotari, Joris Delanghe.   

Abstract

BACKGROUND: Preservation of urine samples is important for centralised laboratory services with automated instruments.
METHODS: A multicentre evaluation was carried out to assess preservative tubes from BD Diagnostics-Preanalytical Systems and from Greiner Bio-One for test strip reading (documented at the level of remission values), for particle counting by flow cytometers (UF-100) and for visual microscopy. Failures were expressed as percentages of originally positive samples beyond a two-fold change (+100% or -50%) from the original values.
RESULTS: The preservative-containing BD Plus C&S plastic, BD Plus UAP and Greiner Stabilur tubes succeeded in preservation of test strip results for 6-24 h (exceptions were glucose and nitrite tests). Greiner boric acid tube showed false negative results in leukocyte, protein and ketone strip tests immediately after adding the preservative. Urine red blood cell counts (with Sysmex UF-100) were preserved for 5 h in BD Plus C&S plastic and Greiner Stabilur tubes (Greiner tubes having clearly larger preservative-related background). Bacteria or white blood cell counting succeeded in BD Plus C&S plastic tubes for 5 or 24 h, respectively, but up to 72 h in Greiner Stabilur tubes. In visual microscopy, the Greiner Stabilur tube was slightly better than the BD Plus C&S plastic tube.
CONCLUSIONS: Urine specimens can be transported at +20 degrees C on the day of collection if preserved properly. Longer delays need careful planning with current preservatives. Flow cytometry with UF-100 is sensitive to non-dissolved preservative remnants.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18839472     DOI: 10.1515/cclm.2008.122

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  8 in total

Review 1.  Gold Nanoparticles for In Vitro Diagnostics.

Authors:  Wen Zhou; Xia Gao; Dingbin Liu; Xiaoyuan Chen
Journal:  Chem Rev       Date:  2015-06-26       Impact factor: 60.622

2.  Biospecimen reporting for improved study quality (BRISQ).

Authors:  Helen M Moore; Andrea B Kelly; Scott D Jewell; Lisa M McShane; Douglas P Clark; Renata Greenspan; Daniel F Hayes; Pierre Hainaut; Paula Kim; Elizabeth Mansfield; Olga Potapova; Peter Riegman; Yaffa Rubinstein; Edward Seijo; Stella Somiari; Peter Watson; Heinz-Ulrich Weier; Claire Zhu; Jim Vaught
Journal:  J Proteome Res       Date:  2011-06-21       Impact factor: 4.466

3.  Biospecimen Reporting for Improved Study Quality.

Authors:  Helen M Moore; Andrea Kelly; Scott D Jewell; Lisa M McShane; Douglas P Clark; Renata Greenspan; Pierre Hainaut; Daniel F Hayes; Paula Kim; Elizabeth Mansfield; Olga Potapova; Peter Riegman; Yaffa Rubinstein; Edward Seijo; Stella Somiari; Peter Watson; Heinz-Ulrich Weier; Claire Zhu; Jim Vaught
Journal:  Biopreserv Biobank       Date:  2011-04       Impact factor: 2.300

4.  Analytical performance, agreement and user-friendliness of six point-of-care testing urine analysers for urinary tract infection in general practice.

Authors:  Marjolein J C Schot; Sanne van Delft; Antoinette M J Kooijman-Buiting; Niek J de Wit; Rogier M Hopstaken
Journal:  BMJ Open       Date:  2015-05-18       Impact factor: 2.692

5.  Use of Sysmex UF-5000 flow cytometry in rapid diagnosis of urinary tract infection and the importance of validating carryover rates against bacterial count cut-off.

Authors:  Kjersti Haugum; Maria Schei Haugan; Jannicke Skage; Mariann Tetik; Aleksandra Jakovljev; Hans-Johnny Schjelderup Nilsen; Jan Egil Afset
Journal:  J Med Microbiol       Date:  2021-12       Impact factor: 2.472

Review 6.  Preanalytical requirements of urinalysis.

Authors:  Joris Delanghe; Marijn Speeckaert
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

7.  The effectiveness of BD Vacutainer® Plus Urinalysis Preservative Tubes in preservation of urine for chemical strip analysis and particle counting.

Authors:  Merve Kaymak Ekşioğlu; Özlem Çakır Madenci; Nihal Yücel; Abdullah Elçi; Bülent Turhan; Gani Orhan; Asuman Orçun
Journal:  Biochem Med (Zagreb)       Date:  2016       Impact factor: 2.313

8.  Prospective, observational study comparing automated and visual point-of-care urinalysis in general practice.

Authors:  Sanne van Delft; Annelijn Goedhart; Mark Spigt; Bart van Pinxteren; Niek de Wit; Rogier Hopstaken
Journal:  BMJ Open       Date:  2016-08-08       Impact factor: 2.692

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.