Dose performance and image quality: dual source CT versus single source CT in cardiac CT angiography.

Min Wang1, Heng-Tao Qi, Xi-Ming Wang, Tao Wang, Jiu-Hong Chen, Cheng Liu.   

Abstract

OBJECTIVE: To evaluate dose performance and image quality of 64-slice dual source CT (DSCT) in comparison to 64-slice single source CT (SSCT) in cardiac CT angiography (CTA).
METHODS: 100 patients examined by DSCT and 60 patients scanned by SSCT were included in this study. Objective indices such as image noise, contrast-to-noise ratio and signal-to-noise ratio were analyzed. Subjective image quality was assessed by two cardiovascular radiologists in consensus using a four-point scale (1=excellent to 4=not acceptable). Estimation of effective dose was performed on the basis of dose length product (DLP).
RESULTS: At low heart rates (<70 bpm), image quality of SSCT was equivalent to that of DSCT (P>0.05), but, at high heart rates (>70 bpm), DSCT provided robust image quality (P<0.05). The average effective dose of SSCT was 9.3+/-0.9 mSv at low heart rates (<70 bpm) while, the average estimated effective doses of DSCT were 9.1+/-1.3 mSv, 8.3+/-1.1 mSv, 7.9+/-1.1 mSv, 6.9+/-0.7 mSv, and 5.9+/-1.3 mSv, corresponding to heart rates of 50-59 bpm, 60-69 bpm, 70-79 bpm, 80-89 bpm, and 90-100 bpm.
CONCLUSION: For cardiac CTA, both DSCT and SSCT can get good image quality at low heart rates (<70 bpm) with a similar radiation dose, but, at high heart rates (>70 bpm), DSCT is able to provide robust diagnostic image quality at doses far below that of SSCT.

Entities:  

Mesh:

Year:  2008        PMID: 18835119     DOI: 10.1016/j.ejrad.2008.08.010

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  7 in total

1.  Quantification of radiation dose savings in cardiac computed tomography using prospectively triggered mode and ECG pulsing: a phantom study.

Authors:  Lukas Lehmkuhl; Dieter Gosch; H D Nagel; Patrick Stumpp; Thomas Kahn; Matthias Gutberlet
Journal:  Eur Radiol       Date:  2010-04-09       Impact factor: 5.315

2.  High diagnostic accuracy of prospective ECG-gating 64-slice computed tomography coronary angiography for the detection of in-stent restenosis: in-stent restenosis assessment by low-dose MDCT.

Authors:  Daniele Andreini; Gianluca Pontone; Antonio L Bartorelli; Saima Mushtaq; Daniela Trabattoni; Erika Bertella; Sarah Cortinovis; Andrea Annoni; Alberto Formenti; Giovanni Ballerini; Piergiuseppe Agostoni; Cesare Fiorentini; Mauro Pepi
Journal:  Eur Radiol       Date:  2011-02-18       Impact factor: 5.315

3.  Cardiac CT imaging in coronary artery disease: Current status and future directions.

Authors:  Zhonghua Sun
Journal:  Quant Imaging Med Surg       Date:  2012-06

4.  Comparison of radiation dose and image quality: 320-MDCT versus 64-MDCT coronary angiography.

Authors:  Atif Khan; Faisal Khosa; Khurram Nasir; Aya Yassin; Melvin E Clouse
Journal:  AJR Am J Roentgenol       Date:  2011-07       Impact factor: 3.959

5.  Effect of Tube Voltage (100 vs. 120 kVp) on Radiation Dose and Image Quality using Prospective Gating 320 Row Multi-detector Computed Tomography Angiography.

Authors:  Atif N Khan; Faisal Khosa; Waqas Shuaib; Khurram Nasir; Ron Blankstein; Melvin Clouse
Journal:  J Clin Imaging Sci       Date:  2013-12-31

6.  A rare case of crossed pulmonary arteries in an infant - case report.

Authors:  Jin Chen; Yue Feng
Journal:  J Cardiothorac Surg       Date:  2013-04-11       Impact factor: 1.637

7.  Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography.

Authors:  Noortje van der Bijl; Raoul M S Joemai; Bart J A Mertens; Albert de Roos; Wouter J H Veldkamp; Jeroen J Bax; Joanne D Schuijf; Jacob Geleijns; Lucia J M Kroft
Journal:  Int J Cardiovasc Imaging       Date:  2012-09-22       Impact factor: 2.357

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.