A rat head holder for simultaneous scanning of two rats in small animal PET scanners: design, construction, feasibility testing and kinetic validation.

Tee Ean Cheng1, Karmen K Yoder, Marc D Normandin, Shannon L Risacher, Alexander K Converse, Joseph A Hampel, Michael A Miller, Evan D Morris.   

Abstract

UNLABELLED: To reduce imaging costs, we designed a head holder for scanning two rats simultaneously in small animal PET scanners. Our goals were (i) to maintain high sensitivity and (ii) to minimize repositioning error between scans.
METHODS: A semi-stereotaxic dual rat head holder was designed and constructed for dual rat scanning in our IndyPET-II scanner and the commercial microPET P4. It was also used for single rat scanning in a small-bore, high-resolution animal scanner ("ISAP"). Positional repeatability was validated via multiple [11C]Raclopride scans of a single rat on different days. Accuracy of repositioning was determined by visual comparison of images, and by metrics derived through image alignment. Kinetic validation was assessed via analysis of [18F]Fluorodeoxyglucose ([18F]FDG) dynamic PET studies of six rats. Each rat was scanned twice: once individually, with brain positioned at the center of field of view (CFOV), and once with a partner, with brain away from CFOV. Both rats were injected with FDG during each dual rat session. Patlak uptake constants (Ki) were calculated from whole brain images. Effects of attenuation and scatter correction on single versus dual scan images were explored.
RESULTS: Image comparison and alignment metrics indicated excellent repositioning of rats. Scaled time-activity-curves from single and dual rat scans were indistinguishable. Average single and dual scan Ki values differed by only 6.3+/-7.5%.
CONCLUSION: Dual rat scanning in a semi-stereotaxic holder is practical for economical small animal scanning and does not compromise kinetic accuracy of [18F]FDG dynamic scan data.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18824025      PMCID: PMC3725621          DOI: 10.1016/j.jneumeth.2008.08.031

Source DB:  PubMed          Journal:  J Neurosci Methods        ISSN: 0165-0270            Impact factor:   2.390


  8 in total

1.  Performance evaluation of the microPET P4: a PET system dedicated to animal imaging.

Authors:  C Tai; A Chatziioannou; S Siegel; J Young; D Newport; R N Goble; R E Nutt; S R Cherry
Journal:  Phys Med Biol       Date:  2001-07       Impact factor: 3.609

2.  Evaluation of a stereotactic frame for repositioning of the rat brain in serial positron emission tomography imaging studies.

Authors:  D J Rubins; A K Meadors; S Yee; W P Melega; S R Cherry
Journal:  J Neurosci Methods       Date:  2001-05-30       Impact factor: 2.390

Review 3.  Small animal PET.

Authors:  Ralph Myers; Susan Hume
Journal:  Eur Neuropsychopharmacol       Date:  2002-12       Impact factor: 4.600

Review 4.  Dedicated small animal scanners: a new tool for drug development?

Authors:  S P Hume; R Myers
Journal:  Curr Pharm Des       Date:  2002       Impact factor: 3.116

5.  A head holder for magnetic resonance imaging that allows the stereotaxic alignment of spontaneously occurring intracranial mouse tumors.

Authors:  Tsuyoshi Tada; Mike Wendland; Nathan Watson; Nagato Kuriyama; Hiroko Kuriyama; Tim Roberts; Michael Burns; William Weiss; Mark A Israel
Journal:  J Neurosci Methods       Date:  2002-04-30       Impact factor: 2.390

6.  Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data.

Authors:  C S Patlak; R G Blasberg; J D Fenstermacher
Journal:  J Cereb Blood Flow Metab       Date:  1983-03       Impact factor: 6.200

7.  Analysis of emission tomographic scan data: limitations imposed by resolution and background.

Authors:  R M Kessler; J R Ellis; M Eden
Journal:  J Comput Assist Tomogr       Date:  1984-06       Impact factor: 1.826

8.  Consistent and reproducible slice selection in rodent brain using a novel stereotaxic device for MRI.

Authors:  S T Fricke; R Vink; C Chiodo; I Cernak; L Ileva; A I Faden
Journal:  J Neurosci Methods       Date:  2004-06-15       Impact factor: 2.390

  8 in total
  8 in total

1.  Imaging of alcohol-induced dopamine release in rats:preliminary findings with [(11) C]raclopride PET.

Authors:  Jenna M Sullivan; Shannon L Risacher; Marc D Normandin; Karmen K Yoder; Janice C Froehlich; Evan D Morris
Journal:  Synapse       Date:  2011-03-28       Impact factor: 2.562

2.  Optimized Methodology for Reference Region and Image-Derived Input Function Kinetic Modeling in Preclinical PET.

Authors:  Jason Bini; Christine R Lattin; Takuya Toyonaga; Sjoerd J Finnema; Richard Carson
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-06-11

3.  Impact of a multiple mice holder on quantitation of high-throughput MicroPET imaging with and without Ct attenuation correction.

Authors:  Frezghi Habte; Gang Ren; Timothy C Doyle; Hongguang Liu; Zhen Cheng; David S Paik
Journal:  Mol Imaging Biol       Date:  2013-10       Impact factor: 3.488

4.  Fully automated synthesis and initial PET evaluation of [11C]PBR28.

Authors:  Min Wang; Karmen K Yoder; Mingzhang Gao; Bruce H Mock; Xiao-Ming Xu; Andrew J Saykin; Gary D Hutchins; Qi-Huang Zheng
Journal:  Bioorg Med Chem Lett       Date:  2009-08-15       Impact factor: 2.823

5.  High throughput static and dynamic small animal imaging using clinical PET/CT: potential preclinical applications.

Authors:  Nicolas Aide; Cédric Desmonts; Jean-Mathieu Beauregard; Thomas Beyer; Kathryn Kinross; Peter Roselt; Oliver Neels; Denis Agostini; Stéphane Bardet; Gérard Bouvard; Rodney J Hicks
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-01-27       Impact factor: 9.236

6.  High-Throughput PET/CT Imaging Using a Multiple-Mouse Imaging System.

Authors:  Hannah E Greenwood; Zoltan Nyitrai; Gabor Mocsai; Sandor Hobor; Timothy H Witney
Journal:  J Nucl Med       Date:  2019-09-13       Impact factor: 10.057

7.  Simultaneous functional MRI of two awake marmosets.

Authors:  Kyle M Gilbert; Justine C Cléry; Joseph S Gati; Yuki Hori; Kevin D Johnston; Alexander Mashkovtsev; Janahan Selvanayagam; Peter Zeman; Ravi S Menon; David J Schaeffer; Stefan Everling
Journal:  Nat Commun       Date:  2021-11-16       Impact factor: 14.919

8.  High-resolution dynamic imaging and quantitative analysis of lung cancer xenografts in nude mice using clinical PET/CT.

Authors:  Ying Yi Wang; Kai Wang; Zuo Yu Xu; Yan Song; Chu Nan Wang; Chong Qing Zhang; Xi Lin Sun; Bao Zhong Shen
Journal:  Oncotarget       Date:  2017-04-20
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.