Comparison of two highly automated DNA extraction systems for quantifying Epstein-Barr virus in whole blood.

C Mengelle1, F Legrand-Abravanel, J-M Mansuy, C Barthe, I Da Silva, J Izopet.   

Abstract

BACKGROUND: Optimal automated molecular methods are needed to monitor Epstein-Barr virus (EBV) infections in transplant recipients.
OBJECTIVES: To compare the extraction of EBV DNA from whole blood using the COBAS Ampliprep and the MagNA Pure instruments (Roche) for quantifying EBV DNA by real-time PCR. STUDY
DESIGN: EBV DNA content was determined on clinical samples extracted by both systems.
RESULTS: The detection limit was 2.16log(10)copies/mL using the COBAS Ampliprep extraction system. Specificity was 100% and we saw no cross-contamination. Extraction was linear from 2.60 to 5.60log(10)copies/mL. The intra-assay variation was 1.91% for 3.60, 2% for 4.60 and 4.51% for 5.60log(10)copies/mL; inter-assay variation was 4.88%. Sixty-six samples were tested: 26 were positive and 28 were negative by both methods. One sample was MagNA Pure positive/COBAS Ampliprep negative (virus load 3.15log(10)copies/mL) and 10 samples were MagNA Pure negative/COBAS Ampliprep positive (virus loads from 1.59 to 3.51log(10)copies/mL) (P<0.0001). Both methods gave similar quantitative results (average difference 0.07log(10)copies/mL) which were well correlated (r=0.73, P<0.001).
CONCLUSIONS: The COBAS Ampliprep extraction system is comparable to the MagNA Pure and offers a high reliability for extracting EBV DNA from whole blood.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18819838     DOI: 10.1016/j.jcv.2008.08.005

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


  5 in total

1.  Comparison of commercial extraction systems and PCR assays for quantification of Epstein-Barr virus DNA load in whole blood.

Authors:  Raphaële Germi; Julien Lupo; Touyana Semenova; Sylvie Larrat; Nelly Magnat; Laurence Grossi; Jean-Marie Seigneurin; Patrice Morand
Journal:  J Clin Microbiol       Date:  2012-01-11       Impact factor: 5.948

2.  Comparative evaluation of a commercially available automated system for extraction of viral DNA from whole blood: application to monitoring of epstein-barr virus and cytomegalovirus load.

Authors:  Sylvie Pillet; Thomas Bourlet; Bruno Pozzetto
Journal:  J Clin Microbiol       Date:  2009-08-26       Impact factor: 5.948

3.  Comparison of QIAsymphony automated and QIAamp manual DNA extraction systems for measuring Epstein-Barr virus DNA load in whole blood using real-time PCR.

Authors:  Stella Laus; Lawrence A Kingsley; Michael Green; Robert M Wadowsky
Journal:  J Mol Diagn       Date:  2011-09-01       Impact factor: 5.568

Review 4.  Using Epstein-Barr viral load assays to diagnose, monitor, and prevent posttransplant lymphoproliferative disorder.

Authors:  Margaret L Gulley; Weihua Tang
Journal:  Clin Microbiol Rev       Date:  2010-04       Impact factor: 26.132

5.  Prospective evaluation of a new automated nucleic acid extraction system using routine clinical respiratory specimens.

Authors:  C Mengelle; J-M Mansuy; K Sandres-Sauné; C Barthe; J Boineau; J Izopet
Journal:  J Med Virol       Date:  2012-06       Impact factor: 2.327

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.