Implantable CRT device diagnostics identify patients with increased risk for heart failure hospitalization.

Giovanni B Perego1, Maurizio Landolina, Giuseppe Vergara, Maurizio Lunati, Gabriele Zanotto, Alessia Pappone, Gabriele Lonardi, Giancarlo Speca, Saverio Iacopino, Annamaria Varbaro, Shantanu Sarkar, Doug A Hettrick, Alessandra Denaro.   

Abstract

PURPOSE: To determine the association between device-determined diagnostic indices, including intrathoracic impedance, and heart failure (HF) hospitalization.
METHODS: Clinical and device diagnostic data of 558 HF patients indicated for CRT-D therapy (In Sync Sentry, Medtronic Inc.) were prospectively collected from 34 centers. Device-recorded intrathoracic impedance fluid index threshold crossing event (TCE), mean activity counts, tachyarrhythmia events, night heart rate (NHR) and heart rate variability (HRV) were compared within patients with vs. without documented HF hospitalization.
RESULTS: Mean follow-up was 326 +/- 216 days. Patients hospitalized for HF had significantly higher rates of TCE, a higher percentage of days with the thoracic impedance fluid index above the programmed threshold, a higher percentage of days with low activity, with low HRV or with high NHR. Multivariate analysis showed that TCE resulted in a 36% increased probability of HF hospitalization. Both TCE duration and patient activity were also significantly associated with hospitalization. Kaplan Meier analysis indicated that patients with more TCE events were significantly more likely to be hospitalized (log rank test, p = 0.005).
CONCLUSIONS: Decreased intrathoracic impedance, low patient activity and low HRV were all independently associated with increased risk for HF hospitalization in HF patients treated with resynchronization therapy. Device-derived diagnostic data may provide valuable and reliable indices for the prognostic stratification of HF patients.

Entities:  

Mesh:

Year:  2008        PMID: 18810621     DOI: 10.1007/s10840-008-9303-5

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  12 in total

1.  Critical analysis of dual-chamber implantable cardioverter-defibrillator arrhythmia detection : results and technical considerations.

Authors:  B L Wilkoff; V Kühlkamp; K Volosin; K Ellenbogen; B Waldecker; S Kacet; J M Gillberg; C M DeSouza
Journal:  Circulation       Date:  2001-01-23       Impact factor: 29.690

2.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction.

Authors:  Arthur J Moss; Wojciech Zareba; W Jackson Hall; Helmut Klein; David J Wilber; David S Cannom; James P Daubert; Steven L Higgins; Mary W Brown; Mark L Andrews
Journal:  N Engl J Med       Date:  2002-03-19       Impact factor: 91.245

3.  Daily shock impedance measured by implantable cardioverter defibrillator is useful in the management of congestive heart failure.

Authors:  Kohei Matsushita; Toshiyuki Ishikawa; Shinichi Sumita; Tsukasa Kobayashi; Hideyuki Ogawa; Noriko Inoue; Katsumi Matsumoto; Minoru Taima; Ichirou Nakazawa; Teruyasu Sugano; Tomoaki Ishigami; Kazuaki Uchino; Satoshi Umemura
Journal:  Circ J       Date:  2006-11       Impact factor: 2.993

4.  Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure.

Authors:  Gust H Bardy; Kerry L Lee; Daniel B Mark; Jeanne E Poole; Douglas L Packer; Robin Boineau; Michael Domanski; Charles Troutman; Jill Anderson; George Johnson; Steven E McNulty; Nancy Clapp-Channing; Linda D Davidson-Ray; Elizabeth S Fraulo; Daniel P Fishbein; Richard M Luceri; John H Ip
Journal:  N Engl J Med       Date:  2005-01-20       Impact factor: 91.245

5.  Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure.

Authors:  Dirk Vollmann; Herbert Nägele; Patrick Schauerte; Uwe Wiegand; Christian Butter; Gabriele Zanotto; Aurelio Quesada; Axel Guthmann; Michael R S Hill; Barbara Lamp
Journal:  Eur Heart J       Date:  2007-02-19       Impact factor: 29.983

6.  Physiological approach to monitor patients in congestive heart failure: application of a new implantable device-based system to monitor daily life activity and ventilation.

Authors:  Eric Page; Serge Cazeau; Philippe Ritter; Daniel Galley; Cyrille Casset
Journal:  Europace       Date:  2007-05-03       Impact factor: 5.214

7.  Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization.

Authors:  Cheuk-Man Yu; Li Wang; Elaine Chau; Raymond Hon-Wah Chan; Shun-Ling Kong; Man-Oi Tang; Jill Christensen; Robert W Stadler; Chu-Pak Lau
Journal:  Circulation       Date:  2005-08-01       Impact factor: 29.690

8.  Continuous autonomic assessment in patients with symptomatic heart failure: prognostic value of heart rate variability measured by an implanted cardiac resynchronization device.

Authors:  Philip B Adamson; Andrew L Smith; William T Abraham; Karen J Kleckner; Robert W Stadler; Alex Shih; Melissa M Rhodes
Journal:  Circulation       Date:  2004-08-16       Impact factor: 29.690

9.  Intrathoracic impedance monitoring to detect chronic heart failure deterioration: relationship to changes in NT-proBNP.

Authors:  Lars Lüthje; Dirk Vollmann; Till Drescher; Peter Schott; Dieter Zenker; Gerd Hasenfuss; Christina Unterberg
Journal:  Eur J Heart Fail       Date:  2007-04-25       Impact factor: 15.534

Review 10.  Atrial fibrillation and congestive heart failure: risk factors, mechanisms, and treatment.

Authors:  E Kevin Heist; Jeremy N Ruskin
Journal:  Prog Cardiovasc Dis       Date:  2006 Jan-Feb       Impact factor: 8.194

View more
  12 in total

Review 1.  How to improve outcomes: should we put more emphasis on programming and medical care and less on patient selection?

Authors:  Laszlo Buga
Journal:  Heart Fail Rev       Date:  2012-11       Impact factor: 4.214

Review 2.  Measuring impedance in congestive heart failure: current options and clinical applications.

Authors:  W H Wilson Tang; Wilson Tong
Journal:  Am Heart J       Date:  2008-12-16       Impact factor: 4.749

3.  Economic impact of remote monitoring on ordinary follow-up of implantable cardioverter defibrillators as compared with conventional in-hospital visits. A single-center prospective and randomized study.

Authors:  Leonardo Calò; Alessio Gargaro; Ermenegildo De Ruvo; Gabriele Palozzi; Luigi Sciarra; Marco Rebecchi; Fabrizio Guarracini; Alessandro Fagagnini; Enrico Piroli; Ernesto Lioy; Antonio Chirico
Journal:  J Interv Card Electrophysiol       Date:  2013-03-21       Impact factor: 1.900

4.  Threshold crossing of device-based intrathoracic impedance trends identifies relatively increased mortality risk.

Authors:  Wai Hong Wilson Tang; Eduardo N Warman; James W Johnson; Roy S Small; James Thomas Heywood
Journal:  Eur Heart J       Date:  2012-06-07       Impact factor: 29.983

5.  Potential value of automated daily screening of cardiac resynchronization therapy defibrillator diagnostics for prediction of major cardiovascular events: results from Home-CARE (Home Monitoring in Cardiac Resynchronization Therapy) study.

Authors:  Stefan Sack; Christian Michael Wende; Herbert Nägele; Amos Katz; Wolfgang Rudolf Bauer; Craig Scott Barr; Klaus Malinowski; Harald Schwacke; Francisco Leyva; Jochen Proff; Sergey Berdyshev; Vincent Paul
Journal:  Eur J Heart Fail       Date:  2011-09       Impact factor: 15.534

6.  Stratifying patients at the risk of heart failure hospitalization using existing device diagnostic thresholds.

Authors:  Vinod Sharma; Lisa D Rathman; Roy S Small; David J Whellan; Jodi Koehler; Eduardo Warman; William T Abraham
Journal:  Heart Lung       Date:  2014-12-24       Impact factor: 2.210

7.  Incremental Value of Implantable Cardiac Device Diagnostic Variables Over Clinical Parameters to Predict Mortality in Patients With Mild to Moderate Heart Failure.

Authors:  Jaimie Manlucu; Vinod Sharma; Jodi Koehler; Eduardo N Warman; George A Wells; Lorne J Gula; Raymond Yee; Anthony S Tang
Journal:  J Am Heart Assoc       Date:  2019-07-11       Impact factor: 5.501

Review 8.  Accelerometer-assessed physical behavior and the association with clinical outcomes in implantable cardioverter-defibrillator recipients: A systematic review.

Authors:  Maarten Z H Kolk; Diana M Frodi; Tariq O Andersen; Joss Langford; Soeren Z Diederichsen; Jesper H Svendsen; Hanno L Tan; Reinoud E Knops; Fleur V Y Tjong
Journal:  Cardiovasc Digit Health J       Date:  2021-11-24

9.  Design of the evolution of management strategies of heart failure patients with implantable defibrillators (EVOLVO) study to assess the ability of remote monitoring to treat and triage patients more effectively.

Authors:  Maurizio Marzegalli; Maurizio Landolina; Maurizio Lunati; Giovanni B Perego; Alessia Pappone; Giuseppe Guenzati; Carlo Campana; Maria Frigerio; Gianfranco Parati; Antonio Curnis; Irene Colangelo; Sergio Valsecchi
Journal:  Trials       Date:  2009-06-18       Impact factor: 2.279

10.  Development and validation of an integrated diagnostic algorithm derived from parameters monitored in implantable devices for identifying patients at risk for heart failure hospitalization in an ambulatory setting.

Authors:  Martin R Cowie; Shantanu Sarkar; Jodi Koehler; David J Whellan; George H Crossley; Wai Hong Wilson Tang; William T Abraham; Vinod Sharma; Massimo Santini
Journal:  Eur Heart J       Date:  2013-03-19       Impact factor: 29.983

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.