Yuichi Kasai1, Tadashi Inaba, Koji Akeda, Atsumasa Uchida.
Abstract
BACKGROUND: There have been reports of serious complications associated with pedicle screw fixation, including nerve root injuries caused by accidental screw insertion. We have developed a new system of lumbar spinal instrumentation that we call Tadpole system. The purposes of this report were to show the results of a biomechanical study and the short-term outcome of a clinical study, as well as to determine the usefulness of this system.Entities:
Year: 2008 PMID: 18786272 PMCID: PMC2565584 DOI: 10.1186/1749-799X-3-41
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Figure 1Tadpole system®; hook-and-rod system of spinal instrumentation.
Figure 2Tadpole system® positioned using a unilateral approach.
Figure 3X-rays of a case of two-level fusion
Figure 4X-rays of a case of single-level fusion
Figure 5Operative techniques for Tadpole system® a; Guide holes for hooks are made on the interspinous ligament using a starter awl
b; The hook is held with a hook holder, and it is inserted into the interspinous ligament
c; The 2 hooks are connected in the center to a rod, and the rod is connected to the cranial and caudal hooks.
d; The hooks are tightened temporarily while compression is maintained between the cranial and caudal hooks.
e; The set of 2 hooks for the stabilization of a spinous process is tightened temporarily, pinching each spinous process
f; The screws of all hooks are tightened using a screwdriver
Mean range of motion of each model on the biomechanical study
| (mean± SD) | |||
| ROM of forward – backward flextion | ROM of right – left side bending | ROM of right – left side rotaton | |
| Normal control model | 7.2° ± 4.8° | 7.8° ± 3.8° | 13.8° ± 7.1° |
| Injured model | 18.6° ± 6.2° | 22.1° ± 8.8° | 24.8° ± 10.4° |
| PSF model | 4.3° ± 2.9° | 2.5° ± 1.3° | 3.4° ± 2.6° |
| TS model | 3.4° ± 2.2° | 5.1° ± 2.5° | 7.9° ± 5.2° |
Normal control model; non-treatment model
Injured model; model resected the facet joint between L3 and L4
PSF model; model generated from the injured model with pedicle screw fixation
TS model; model created from the injured model with the Tadpole system®