Literature DB >> 18783418

Stability of prototype two-piece zirconia and titanium implants after artificial aging: an in vitro pilot study.

Ralf-Joachim Kohal1, Hans Christian Finke, Gerold Klaus.   

Abstract

BACKGROUND: Zirconia oral implants are a new topic in implant dentistry. So far, no data are available on the biomechanical behavior of two-piece zirconia implants. Therefore, the purpose of this pilot investigation was to test in vitro the fracture strength of two-piece cylindrical zirconia implants after aging in a chewing simulator.
MATERIALS AND METHODS: This laboratory in vitro investigation comprised three different treatment groups. Each group consisted of 16 specimens. In group 1, two-piece zirconia implants were restored with zirconia crowns (zirconia copings veneered with Triceram; Esprident, Ispringen, Germany), and in group 2 zirconia implants received Empress 2 single crowns (Ivoclar Vivadent AG, Schaan, Liechtenstein). The implants, including the abutments, in the two zirconia groups were identical. In group 3, similar titanium implants were reconstructed with porcelain-fused-to-metal crowns. Eight samples of each group were submitted to artificial aging with a long-term load test in the artificial mouth (chewing simulator). Subsequently, all not artificially aged samples and all artificially aged samples that survived the long-term loading of each group were submitted to a fracture strength test in a universal testing machine. For the pairwise comparisons in the different test groups with or without artificial loading and between the different groups at a given artificial loading condition, the Wilcoxon rank-sum test for independent samples was used. The significance level was set at 5%.
RESULTS: One sample of group 1 (veneer fracture), none of group 2, and six samples of group 3 (implant abutment screw fractures) failed while exposed to the artificial mouth. The values for the fracture strength after artificial loading with 1.2 million cycles for group 1 were between 45 and 377 N (mean: 275.7 N), in group 2 between 240 and 314 N (mean: 280.7 N), and in the titanium group between 45 and 582 N (mean: 165.7 N). The fracture strength results without artificial load for group 1 amounted to between 270 and 393 N (mean: 325.1 N), for group 2 between 235 and 321 N (mean: 281.8 N), and between 474 and 765 N (mean: 595.2 N) for the titanium group. The failure mode during the fracture testing in the zirconia implant groups was a fracture of the implant head and a bending/fracture of the abutment screw in the titanium group.
CONCLUSIONS: Within the limits of this pilot investigation, the biomechanical stability of all tested prototype implant groups seems to be - compared with the possibly exerted occlusal forces - borderline for clinical use. A high number of failures occurred already during the artificial loading in the titanium group at the abutment screw level. The zirconia implant groups showed irreparable implant head fractures at relatively low fracture loads. Therefore, the clinical use of the presented prototype implants has to be questioned.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18783418     DOI: 10.1111/j.1708-8208.2008.00116.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  10 in total

1.  Immediate, Non Submerged Root Analog Zirconia Implant in Single Rooted Tooth Replacement: Case Report with 2 years Follow Up.

Authors:  Amod Patankar; Rajesh Kshirsagar; Swapna Patankar; Sudhir Pawar
Journal:  J Maxillofac Oral Surg       Date:  2015-03-27

2.  Osseointegration of a novel injection molded 2-piece ceramic dental implant: a study in minipigs.

Authors:  Geninho Thomé; Rebecca Sandgren; Sergio Bernardes; Larissa Trojan; Nils Warfving; Benjamin Bellón; Benjamin E Pippenger
Journal:  Clin Oral Investig       Date:  2020-09-10       Impact factor: 3.573

3.  Stability of Cantilever Fixed Dental Prostheses on Zirconia Implants.

Authors:  Nadja Rohr; Reto Nüesch; Rebecca Greune; Gino Mainetti; Sabrina Karlin; Lucia K Zaugg; Nicola U Zitzmann
Journal:  Materials (Basel)       Date:  2022-05-19       Impact factor: 3.748

4.  Comparison of peri-implant bone formation around injection-molded and machined surface zirconia implants in rabbit tibiae.

Authors:  Hong-Kyun Kim; Kyung Mi Woo; Won-Jun Shon; Jin-Soo Ahn; Seunghee Cha; Young-Seok Park
Journal:  Dent Mater J       Date:  2015       Impact factor: 2.102

5.  All-Ceramic Single Crown Restauration of Zirconia Oral Implants and Its Influence on Fracture Resistance: An Investigation in the Artificial Mouth.

Authors:  Ralf-Joachim Kohal; Jolanta Bernadette Kilian; Susanne Stampf; Benedikt Christopher Spies
Journal:  Materials (Basel)       Date:  2015-04-01       Impact factor: 3.623

Review 6.  A Critical Review of Dental Implant Materials with an Emphasis on Titanium versus Zirconia.

Authors:  Reham B Osman; Michael V Swain
Journal:  Materials (Basel)       Date:  2015-03-05       Impact factor: 3.623

7.  The biocompatibility and mechanical properties of plasma sprayed zirconia coated abutment.

Authors:  Zhengfei Huang; Zhifeng Wang; Kaifeng Yin; Chuanhua Li; Meihua Guo; Jing Lan
Journal:  J Adv Prosthodont       Date:  2020-06-18       Impact factor: 1.904

Review 8.  Fracture Resistance of Zirconia Oral Implants In Vitro: A Systematic Review and Meta-Analysis.

Authors:  Annalena Bethke; Stefano Pieralli; Ralf-Joachim Kohal; Felix Burkhardt; Manja von Stein-Lausnitz; Kirstin Vach; Benedikt Christopher Spies
Journal:  Materials (Basel)       Date:  2020-01-24       Impact factor: 3.623

9.  Fatigue and Fracture Resistance Testing of Polyether Ether Ketone (PEEK) Implant Abutments in an Ex Vivo Chewing Simulator Model.

Authors:  Babak Saravi; Anselm Flohr; Sebastian B Patzelt; Benedikt C Spies; Derek Hazard; Ralf J Kohal
Journal:  Materials (Basel)       Date:  2022-10-06       Impact factor: 3.748

Review 10.  Zirconia in dental implantology: A review.

Authors:  Abhishek Apratim; Prashanti Eachempati; Kiran Kumar Krishnappa Salian; Vijendra Singh; Saurabh Chhabra; Sanket Shah
Journal:  J Int Soc Prev Community Dent       Date:  2015 May-Jun
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.