Experience from long-term monitoring of RAKR ratios in (192)Ir brachytherapy.

Asa Carlsson Tedgren1, Emil Bengtsson, Håkan Hedtjärn, Asa Johansson, Leif Karlsson, Inger-Lena Lamm, Marie Lundell, Younes Mejaddem, Per Munck Af Rosenschöld, Josef Nilsson, Elinore Wieslander, Jeanette Wolke.   

Abstract

BACKGROUND: Ratios of values of brachytherapy source strengths, as measured by hospitals and vendors, comprise constant differences as, e.g., systematic errors in ion chamber calibration factors and measurement setup. Such ratios therefore have the potential to reveal the systematic changes in routines or calibration services at either the hospital or the vendor laboratory, which could otherwise be hidden by the uncertainty in the source strength values.
METHODS: The RAKR of each new source in 13 afterloading units at five hospitals were measured by well-type ion chambers and compared to values for the same source stated on vendor certificates.
RESULTS: Differences from unity in the ratios of RAKR values determined by hospitals and vendors are most often small and stable around their mean values to within +/- 1.5%. Larger deviations are rare but occur. A decreasing ratio, seen at two hospitals for the same source, was useful in detecting an erroneous pressure gauge at the vendor's site.
CONCLUSIONS: Establishing a mean ratio of RAKR values, as measured at the hospital and supplied on the vendor certificate, and monitoring this as a function of time are an easy way for the early detection of problems with equipment or routines at either the hospital or the vendor site.

Mesh:

Substances:

Year:  2008        PMID: 18774192     DOI: 10.1016/j.radonc.2008.07.024

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  2 in total

1.  The Design of an Audit Test for 60Co Brachytherapy Treatment Planning System.

Authors:  Mohammad Javad Tahmasebi Birgani; Hadis Khorshidsavar; Ali Bagheri; Amir Danyaei; Neda Abdalvand
Journal:  J Med Signals Sens       Date:  2022-05-12

2.  Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources.

Authors:  S Vandana; S D Sharma
Journal:  J Med Phys       Date:  2010-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.