Abstract
Well-type ionization chamber is often used to measure strength of brachytherapy sources. This study aims to check long term response stability of High Dose Rate (HDR)-1000 Plus well-type ionization chamber in terms of reference air kerma rate (RAKR) of a reference (137)Cs brachytherapy source and recommend an optimum frequency of recalibration. An HDR-1000 Plus well-type ionization chamber, a reference (137)Cs brachytherapy source (CDCSJ5), and a MAX-4000 electrometer were used in this study. The HDR-1000 Plus well-type chamber was calibrated in terms of reference air kerma rate by the Standards Laboratory of the International Atomic Energy Agency (IAEA), Vienna. The response of the chamber was verified at regular intervals over a period of eight years using the reference (137)Cs source. All required correction factors were applied in the calculation of the RAKR of the (137)Cs source. This study reveals that the response of the HDR-1000 Plus well-type chamber was well within +/-0.5% for about three years after calibration/recalibration. However, it shows deviations larger than +/-0.5% after three years of calibration/recalibration and the maximum variation in response of the chamber during an eight year period was 1.71%. The optimum frequency of recalibration of a high dose rate well-type chamber should be three years.Entities:
Keywords: Brachytherapy; response stability; source calibration; well chamber
Year: 2010 PMID: 20589119 PMCID: PMC2884301 DOI: 10.4103/0971-6203.62200
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Physical parameters of HDR-1000 Plus well-type ionization chamber
| Active Volume | 245 cm3 |
| Outer Diameter | 102 mm |
| Height | 156 mm |
| Sensitivity | 1.97×1010 nA/Bq (7.3 nA/Ci) |
| Range | 37×104 Bq to 74×1010 Bq |
| Wall Material | Aluminum |
| Wall thickness | 20 mm |
Figure 1Photograph showing HDR-1000 Plus well-type ionization chamber, electrometer and source positioning jigs
Figure 2Response of HDR-1000 Plus well-type ionization chamber to a reference 137Cs source (CDCSJ5)
Figure 3Temporal response of HDR-1000 Plus well-type ionization chamber to a reference 137Cs source (CDCSJ5) after first calibration
Figure 4Temporal response of HDR-1000 Plus well-type ionization chamber to a reference 137Cs source (CDCSJ5) after first and second calibration