MR temperature monitoring applying the proton resonance frequency method in liver and kidney at 0.2 and 1.5 T: segment-specific attainable precision and breathing influence.

Hansjörg Rempp1, Petros Martirosian, Andreas Boss, Stephan Clasen, Antje Kickhefel, Markus Kraiger, Christina Schraml, Claus Claussen, Philippe Pereira, Fritz Schick.   

Abstract

OBJECT: The objective of this study was to evaluate breathing influence on precision in temperature determination by using the proton resonance frequency (PRF) shift method depending on the location in abdominal organs at 0.2 and 1.5 T.
MATERIALS AND METHODS: Phase images were acquired with gradient echo sequences in a total of 12 volunteers at 1.5 and 0.2 T. Different examination protocols were performed (each 8 measurements with (1) in-/expiration, (2) free breathing, (3) under breathhold, (4) with breathing belt triggering, and (5) with navigator triggering (integrated in MR signal acquisition). Regions of interest were placed on liver and kidneys, and the resulting phase differences between the measurements were transformed into corresponding temperature differences.
RESULTS: Precision significantly varied depending on the liver segment or location in the kidney. Gating techniques were found better than breathhold techniques and clearly better than non-gated examinations. The most precise approach reached a standard deviation of 2.0 degrees C under continuous breathing when navigator gating was used at 1.5 T.
CONCLUSION: PRF temperature measurement is feasible even for moving organs in the abdomen at 0.2 and 1.5 T. The location of the target region and the required precision of the measurements should direct the choice of examination mode.

Mesh:

Substances:

Year:  2008        PMID: 18763002     DOI: 10.1007/s10334-008-0139-x

Source DB:  PubMed          Journal:  MAGMA        ISSN: 0968-5243            Impact factor:   2.310


  33 in total

1.  Temperature quantification using the proton frequency shift technique: In vitro and in vivo validation in an open 0.5 tesla interventional MR scanner during RF ablation.

Authors:  R M Botnar; P Steiner; B Dubno; P Erhart; G K von Schulthess; J F Debatin
Journal:  J Magn Reson Imaging       Date:  2001-03       Impact factor: 4.813

2.  Sonographically observed echogenic response during intraoperative radiofrequency ablation of cirrhotic livers: pathologic correlation.

Authors:  John R Leyendecker; Gerald D Dodd; Glenn A Halff; Victor A McCoy; Dacia H Napier; Linda G Hubbard; Kedar N Chintapalli; Shailendra Chopra; W Kenneth Washburn; Robert M Esterl; Francisco G Cigarroa; Ruth E Kohlmeier; Francis E Sharkey
Journal:  AJR Am J Roentgenol       Date:  2002-05       Impact factor: 3.959

3.  Temperature measurement using echo-shifted FLASH at low field for interventional MRI.

Authors:  Y C Chung; J L Duerk; A Shankaranarayanan; M Hampke; E M Merkle; J S Lewin
Journal:  J Magn Reson Imaging       Date:  1999-01       Impact factor: 4.813

4.  Magnetic susceptibility effects on the accuracy of MR temperature monitoring by the proton resonance frequency method.

Authors:  Andreas Boss; Hansjörg Graf; Bernd Müller-Bierl; Stephan Clasen; Diethard Schmidt; Philippe L Pereira; Fritz Schick
Journal:  J Magn Reson Imaging       Date:  2005-12       Impact factor: 4.813

5.  Correction of proton resonance frequency shift temperature maps for magnetic field disturbances caused by breathing.

Authors:  Andriy V Shmatukha; Chris J G Bakker
Journal:  Phys Med Biol       Date:  2006-09-04       Impact factor: 3.609

6.  Noninvasive MRI thermometry with the proton resonance frequency method: study of susceptibility effects.

Authors:  J De Poorter
Journal:  Magn Reson Med       Date:  1995-09       Impact factor: 4.668

7.  Time and temperature dependence of MR parameters during thermal coagulation of ex vivo rabbit muscle.

Authors:  S J Graham; M J Bronskill; R M Henkelman
Journal:  Magn Reson Med       Date:  1998-02       Impact factor: 4.668

8.  Simultaneous magnetic resonance phase and magnitude temperature maps in muscle.

Authors:  H E Cline; K Hynynen; E Schneider; C J Hardy; S E Maier; R D Watkins; F A Jolesz
Journal:  Magn Reson Med       Date:  1996-03       Impact factor: 4.668

9.  MR imaging of laser-tissue interactions.

Authors:  F A Jolesz; A R Bleier; P Jakab; P W Ruenzel; K Huttl; G J Jako
Journal:  Radiology       Date:  1988-07       Impact factor: 11.105

10.  Extracorporeal high intensity focused ultrasound ablation in the treatment of patients with large hepatocellular carcinoma.

Authors:  Feng Wu; Zhi-Biao Wang; Wen-Zhi Chen; Hui Zhu; Jin Bai; Jian-Zhong Zou; Ke-Quan Li; Cheng-Bing Jin; Fang-Lin Xie; Hai-Bing Su
Journal:  Ann Surg Oncol       Date:  2004-11-15       Impact factor: 5.344

View more
  2 in total

1.  Correction of breathing-induced errors in magnetic resonance thermometry of hyperthermia using multiecho field fitting techniques.

Authors:  Cory R Wyatt; Brian J Soher; James R MacFall
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

2.  An experimental model to investigate the targeting accuracy of MR-guided focused ultrasound ablation in liver.

Authors:  Lorena Petrusca; Magalie Viallon; Romain Breguet; Sylvain Terraz; Gibran Manasseh; Vincent Auboiroux; Thomas Goget; Loredana Baboi; Patrick Gross; K Michael Sekins; Christoph D Becker; Rares Salomir
Journal:  J Transl Med       Date:  2014-01-16       Impact factor: 5.531

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.