Literature DB >> 18761484

Assessing the optical performance of multifocal (diffractive) intraocular lenses.

Timo Eppig1, Katja Scholz, Achim Langenbucher.   

Abstract

BACKGROUND AND
PURPOSE: While the optical performance of monofocal refractive lenses can be measured quite easily, more efforts are required to assess the performance of multifocal lenses, due to imaging to several foci. The purpose of this study was to develop a rugged test setup for evaluation of the optical properties of intraocular lenses (IOLs) in conformity to the International Standard ISO 11979.
METHODS: We built a test setup in order to measure the point spread function and to determine the modulation transfer function (MTF) of IOLs. We measured three multifocal IOLs with (1) a diffraction limited model cornea and (2) an individualized aspheric model cornea which shows a significant amount of spherical aberration.
RESULTS: All the three IOL samples tested showed a different impact of spherical aberration on the MTF. The bispheric lenses (Alcon and Dr Schmidt) showed a degradation of the MTF and Strehl ratio with model (2) compared to model (1). In contrast, the MTF of the Tecnis lens, which has a prolate aspheric front surface, improved dramatically in combination with model (2).
CONCLUSION: We demonstrated the functional capability of our measurement system by presentation of a set of working examples. Two different model corneas were used to study the influence of spherical aberration on the MTF of our working examples. An aspheric model cornea, which had already been shown to be suitable for testing monofocal aspherics, was shown to be well suited for testing aspheric multifocal lenses.

Mesh:

Year:  2008        PMID: 18761484     DOI: 10.1111/j.1475-1313.2008.00582.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  6 in total

Review 1.  [Vision with bifocal and multifocal intraocular lenses].

Authors:  K Kirschfeld; M F Land
Journal:  Ophthalmologe       Date:  2011-12       Impact factor: 1.059

2.  [Investigation of the theoretical image quality of aspheric intraocular lenses by decentration. Hoya AF-1 iMics1 und Zeiss ASPHINA(TM) (Invent ZO)].

Authors:  M Gillner; A Langenbucher; T Eppig
Journal:  Ophthalmologe       Date:  2012-03       Impact factor: 1.059

3.  Comparison of through-focus image quality across five presbyopia-correcting intraocular lenses (an American Ophthalmological Society thesis).

Authors:  Jay S Pepose; Daozhi Wang; Griffith E Altmann
Journal:  Trans Am Ophthalmol Soc       Date:  2011-12

Review 4.  Opto-Mechanical Eye Models, a Review on Human Vision Applications and Perspectives for Use in Industry.

Authors:  André Rino Amorim; Boris Bret; José M González-Méijome
Journal:  Sensors (Basel)       Date:  2022-10-10       Impact factor: 3.847

5.  Application of 3-dimensional printing technology to construct an eye model for fundus viewing study.

Authors:  Ping Xie; Zizhong Hu; Xiaojun Zhang; Xinhua Li; Zhishan Gao; Dongqing Yuan; Qinghuai Liu
Journal:  PLoS One       Date:  2014-11-13       Impact factor: 3.240

6.  Simulation of visual acuity by personalizable neuro-physiological model of the human eye.

Authors:  Csilla Fülep; Illés Kovács; Kinga Kránitz; Gábor Erdei
Journal:  Sci Rep       Date:  2019-05-24       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.