Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: comparison of both techniques in a clinical trial.

Jürgen Zix1, Stefan Hug, Gerda Kessler-Liechti, Regina Mericske-Stern.   

Abstract

PURPOSE: Two noninvasive methods to measure dental implant stability are damping capacity assessment (Periotest) and resonance frequency analysis (Osstell). The objective of the present study was to assess the correlation of these 2 techniques in clinical use.
MATERIALS AND METHODS: Implant stability of 213 clinically stable loaded and unloaded 1-stage implants in 65 patients was measured in triplicate by means of resonance frequency analysis and Periotest. Descriptive statistics as well as Pearson's, Spearman's, and intraclass correlation coefficients were calculated with SPSS 11.0.2.
RESULTS: The mean values were 57.66 +/- 8.19 implant stability quotient for the resonance frequency analysis and -5.08 +/- 2.02 for the Periotest. The correlation of both measuring techniques was -0.64 (Pearson) and -0.65 (Spearman). The single-measure intraclass correlation coefficients for the ISQ and Periotest values were 0.99 and 0.88, respectively (95% CI). No significant correlation of implant length with either resonance frequency analysis or Periotest could be found. However, a significant correlation of implant diameter with both techniques was found (P < .005). The correlation of both measuring systems is moderate to good. It seems that the Periotest is more susceptible to clinical measurement variables than the Osstell device. The intraclass correlation indicated lower measurement precision for the Periotest technique. Additionally, the Periotest values differed more from the normal (Gaussian) curve of distribution than the ISQs. Both measurement techniques show a significant correlation to the implant diameter.
CONCLUSION: Resonance frequency analysis appeared to be the more precise technique.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18700378

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  16 in total

Review 1.  Effect of photobiomodulation therapy on mini-implant stability: a systematic review and meta-analysis.

Authors:  Bo Zhang; Xinqi Huang; Sibei Huo; Chenghao Zhang; Xiao Cen; Zhihe Zhao
Journal:  Lasers Med Sci       Date:  2021-03-04       Impact factor: 3.161

2.  Assessment of damping capacity as an index of root proximity in self-drilling orthodontic mini-implants.

Authors:  Mitsuru Motoyoshi; Yasuki Uchida; Miho Matsuoka; Mizuki Inaba; Hiroaki Iwai; Yoko Karasawa; Noriyoshi Shimizu
Journal:  Clin Oral Investig       Date:  2013-03-22       Impact factor: 3.573

3.  Influence of geometric design characteristics on primary stability of orthodontic miniscrews.

Authors:  Eman Saad Radwan; Mona A Montasser; Ahmed Maher
Journal:  J Orofac Orthop       Date:  2018-04-10       Impact factor: 1.938

4.  Intra- and inter-operator concordance of the resonance frequency analysis. A cross-sectional and prospective clinical study.

Authors:  Raquel Antequera-Diaz; Maria-Pilar Quesada-García; Cristina Vallecillo; Marta Vallecillo-Rivas; Esther Muñoz-Soto; Maria-Victoria Olmedo-Gaya
Journal:  Clin Oral Investig       Date:  2022-07-08       Impact factor: 3.573

5.  Resonance frequency analysis-reliability in third generation instruments: Osstell mentor®.

Authors:  Mariano Herrero-Climent; Matteo Albertini; Jose-Vicente Rios-Santos; Pedro Lázaro-Calvo; Ana Fernández-Palacín; Pedro Bullon
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2012-09-01

6.  Effect of bone quality and implant surgical technique on implant stability quotient (ISQ) value.

Authors:  Hong-Gi Yoon; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim; Su-Young Lee
Journal:  J Adv Prosthodont       Date:  2011-03-31       Impact factor: 1.904

7.  Marginal bone response of implants with platform switching and non-platform switching abutments in posterior healed sites: a 1-year prospective study.

Authors:  Yun-Chi Wang; Joseph Y K Kan; Kitichai Rungcharassaeng; Phillip Roe; Jaime L Lozada
Journal:  Clin Oral Implants Res       Date:  2014-01-03       Impact factor: 5.977

8.  Assessment of Osstell ISQ's reliability for implant stability measurement: a cross-sectional clinical study.

Authors:  Mariano Herrero-Climent; Rocio Santos-García; Reyes Jaramillo-Santos; Manuel-Maria Romero-Ruiz; Ana Fernández-Palacin; Pedro Lázaro-Calvo; Pedro Bullón; Jose-Vicente Ríos-Santos
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2013-11-01

Review 9.  Does the Implant Surgical Technique Affect the Primary and/or Secondary Stability of Dental Implants? A Systematic Review.

Authors:  Rola Muhammed Shadid; Nasrin Rushdi Sadaqah; Sahar Abdo Othman
Journal:  Int J Dent       Date:  2014-07-07

10.  Customized SmartPeg for measurement of resonance frequency of mini dental implants.

Authors:  Jagjit Singh Dhaliwal; Rubens F Albuquerque; Ali Fakhry; Sukhbir Kaur; Jocelyne S Feine
Journal:  Int J Implant Dent       Date:  2017-02-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.